Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K990177

    Validate with FDA (Live)

    Date Cleared
    1999-11-18

    (303 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Achieva Ventilator is intended to provide ventilatory support for pediatric (Patients should weigh no less than 11 lbs. (5 kg) and adult patients who require positive pressure mechanical ventilation. The ventilator is for use in home, institutional, and portable settings. The Achieva is contraindicated for use with anesthetic gases. This device is intended to be used on the order, and under the supervision, of a physician.

    The Achieva Report Generator is a management and information tool, to aid in the care of patients requiring ventilatory support. The information displayed by the Report Generator software is intended to be used by healthcare professionals to supplement data obtained using established clinical procedures. The Achieva Report Generator allows users to download, view, and print ventilator performance data thave been stored in the ventilator's internal memory. Achieva Report Generator users can also capture and display a snapshot of real-time waveform data (pressure, peak flow, and volume).

    Device Description

    The Achieva Ventilator is a piston driven device which can deliver a wide range of volumes, inspiratory times, breath rates, oxygen concentrations, and PEEP levels. The Achieva has three operating modes (Assist/Control, SIMV, and Spontaneous) and six ventilation modes.

    The Achieva Ventilator is available in four different models. The Achieva PSx and Achieva PS have pressure support capability. The Achieva PSx, Achieva PS, and Achieva X have an internal oxygen blender. The Achieva PSx and Achieva X have an internal modem.

    The Achieva Ventilator is compatible with previously cleared Nellcor Puritan Bennett patient circuits and accessories. When AC power is unavailable, the Achieva can operate from an internal 24 VDC battery for approximately 4 hours (normal load) or an optional 24 VDC external battery for approximately 20 hours (normal load).

    The Achieva Ventilator is also capable of downloading data to the Achieva Report Generator for purposes of displaying and archiving ventilator data with a computer.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the Achieva Ventilator and Report Generator, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state quantitative acceptance criteria or a specific table summarizing "reported device performance" in terms of clinical outcomes or diagnostic metrics. Instead, it focuses on general performance claims and compliance with standards.

    Here's a breakdown of what the document implies regarding acceptance and performance:

    Acceptance Criteria (Implied)Reported Device Performance (Summary from text)
    Functional Performance (Ventilator): Deliver a wide range of volumes, inspiratory times, breath rates, oxygen concentrations, and PEEP levels. Operate in Assist/Control, SIMV, and Spontaneous modes with six ventilation modes. Internal oxygen blender (for certain models). Internal modem (for certain models). Compatibility with existing circuits/accessories. Battery operation for specified durations."All functions as defined in the published specifications were completely validated." The Achieva is a piston-driven device. Achieva PSx and Achieva PS have pressure support. Achieva PSx, Achieva PS, and Achieva X have internal oxygen blender. Achieva PSx and Achieva X have an internal modem. Compatible with previously cleared Nellcor Puritan Bennett patient circuits and accessories. Operates from internal 24 VDC battery for approx. 4 hours or optional external battery for approx. 20 hours (normal load). The Achieva "performs as intended according to its performance specification."
    Functional Performance (Report Generator): Download, display, and archive ventilator data. Display and print stored data. Capture and display real-time waveform data (pressure, peak flow, volume)."capable of downloading data to the Achieva Report Generator for purposes of displaying and archiving ventilator data with a computer." Allows users to "download, view, and print ventilator performance data" and "capture and display a snapshot of real-time waveform data (pressure, peak flow, and volume)."
    Safety and Effectiveness: Substantial equivalence to predicate devices. Compliance with relevant medical device safety and performance standards (IEC 601-1, IEC 601-1-2, UL 2601-1, ASTM F1100-90, ASTM F1246-91). Demonstrated through non-clinical testing."The Achieva is substantially equivalent to the predicate devices." Complies with IEC 601-1, IEC 601-1-2, UL 2601-1, ASTM F1100-90 (partial), ASTM F1246-91 (partial). "Safety and efficacy were established through non-clinical testing."
    Intended Use: Provide ventilatory support for pediatric (>= 11 lbs./5 kg) and adult patients. For home, institutional, and non-emergency transport settings. Not for use with anesthetic gases. Report Generator supplements data obtained via established clinical procedures.The device meets its "Indications For Use" as stated.
    Nonclinical Testing: Comprehensive testing including software, electrical, mechanical, environmental, and EMC. All functions in published specifications fully validated."The performance of the Achieva and its interface with the Achieva Report Generator was comprehensively tested (including software, electrical, mechanical, environmental and EMC). All functions as defined in the published specifications were completely validated."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: Not applicable. The document explicitly states: "Clinical testing was not performed on the device."
    • Data Provenance: Not applicable. The "study" (non-clinical testing) was conducted internally by the manufacturer, Nellcor Puritan Bennett, Inc.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Not applicable, as no clinical testing with a "test set" and corresponding "ground truth" was performed. The evaluation relied on non-clinical engineering and performance testing against internal specifications and recognized standards.

    4. Adjudication Method for the Test Set:

    • Not applicable, as no clinical testing requiring adjudication was performed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device's safety and effectiveness were established through non-clinical testing and substantial equivalence to predicate devices.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • This question is slightly nuanced in the context of this device. The "Achieva Ventilator" is a standalone medical device in its primary function, delivering ventilation. Its performance was tested as a standalone system (or integrated with its Report Generator). The "Report Generator" software component, while processing data, is described as being used by "healthcare professionals to supplement data obtained using established clinical procedures," implying a human-in-the-loop context for interpretation of the data it generates. However, the performance of the software itself (e.g., its ability to accurately download, display, and archive data) would have been validated in a 'standalone' sense during the comprehensive non-clinical testing. There is no AI algorithm specifically detailed where 'standalone' performance in a diagnostic sense would be reported.

    7. The Type of Ground Truth Used:

    • The "ground truth" for the non-clinical testing was defined by the device's published performance specifications and the requirements of relevant industry standards (IEC, UL, ASTM). The device was tested to ensure it met these predetermined functional and safety criteria.

    8. The Sample Size for the Training Set:

    • Not applicable. This device is not an AI/ML algorithm that requires a "training set" in the conventional sense. Its design and functionality are based on established engineering principles for mechanical ventilation, and its performance was validated through non-clinical testing against specifications, not through training on a dataset.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there was no training set for an AI/ML algorithm.

    Summary of the Study:

    The primary "study" described in this 510(k) pertains to non-clinical performance testing. This comprehensive testing involved:

    • Software, electrical, mechanical, environmental, and EMC evaluations.
    • Validation of all functions as defined in the manufacturer's published specifications.
    • Demonstration of compliance with various international and national safety and performance standards, including IEC 601-1, IEC 601-1-2, UL 2601-1, ASTM F1100-90 (partial), and ASTM F1246-91 (partial).

    The conclusion was that "The Achieva performs as intended according to its performance specification" and "The Achieva is substantially equivalent to the predicate devices." No clinical studies or human reader performance evaluations were conducted or deemed necessary for this 510(k) submission, relying instead on substantial equivalence and exhaustive non-clinical verification.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1