Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K970460

    Validate with FDA (Live)

    Date Cleared
    1998-04-30

    (448 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 840 Ventilator System is used to provide continuous ventilation to patient's requiring respiratory support. This device is used for a wide range of patients from infant to adult and for a wide variety of clinical conditions.

    Device Description

    The device is a critical care ventilator intended to provide continuous ventilation for infant, pediatric, and adult patients. The 840 Ventilator's graphic user interface (GUI) includes ventilator status keys and indicators, symbols and abbreviations with associated definitions, ventilator settings, and patient data. User input to the 840 Ventilator is provided by means of two monochrome or color touch-screen LCDs and a rotary knob. Feedback to the user is accomplished by the LCD displays and LED indicators. The alarms associated with the 840 Ventilator meet and exceed alarm standards for modern critical care ventilators and have been developed in compliance with ISO 9703-1, ISO 9703-2, ISO 9703-3, and EN 475. These alarms are within the classification of "smart" alarms as per the ISO 9703 series of standards. The 840 Ventilator provides both non-technical (patient related) and technical (ventilator related) alarms. These alarms are arranged in a hierarchical structure with high, medium, and low urgency categories. The 840 Ventilator includes two microprocessors: 1) the breath delivery unit (BDU) microprocessor which controls ventilation and 2) the GUI microprocessor which manages the user interface and monitors ventilator and patient data. Each microprocessor verifies that the other's instructions are being carried out properly. Using two independent microprocessors in this fashion prevents a single fault from causing a simultaneous failure of controlling and monitoring functions. Oxygen and air connect directly to the BDU. supplying gas to each of two proportional solenoid (PSOL) valves. The optional 804 Compressor provides pressurized air to the BDU and can be used in place of wall or bottled air. The optional 802 Backup Power Source provides DC power to the BDU power supply in the event that AC power is lost. The 840 Ventilator supplies mandatory or spontaneous breaths with a preset oxygen concentration. A mandatory (or assisted) breath can be pressure or volume controlled. Volume controlled breaths provide the patient with a preset tidal volume, peak flow, waveform, and oxygen concentration. Pressure controlled breaths provide preset pressure, inspiratory time and %O2. A spontaneous breath allows inspiratory flows of up to 200 L/min, with or without pressure support. The 840 Ventilator offers three modes of ventilation: Assist/control (A/C), Spontaneous (SPONT), Synchronous intermittent mandatory ventilation (SIMV). The 840 Ventilator offers four breathing types: Volume-Controlled (VC), Pressure-Controlled (PC), Pressure Support (PS), No Pressure Support (NONE).

    AI/ML Overview

    The Puritan-Bennett 840 Ventilator System is a continuous ventilator intended for critical care use with infant, pediatric, and adult patients requiring respiratory support. The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than presenting a detailed study with specific acceptance criteria and performance metrics for the device itself.

    Based on the provided text, a table of acceptance criteria and reported device performance for the 840 Ventilator System cannot be fully constructed as this type of information is generally not detailed in a 510(k) summary. The document emphasizes compliance with standards and successful internal testing to demonstrate safety and effectiveness, rather than quantitative performance against specific acceptance criteria.

    Here's a breakdown of the information that can be extracted, and where limitations exist due to the nature of the K970460 submission:

    1. A table of acceptance criteria and the reported device performance

    The 510(k) summary does not explicitly list quantitative acceptance criteria and corresponding reported device performance values in the way a clinical trial report would. Instead, it states that the device:

    • Meets and exceeds alarm standards for modern critical care ventilators and has been developed in compliance with ISO 9703-1, ISO 9703-2, ISO 9703-3, and EN 475.
    • Provides "smart" alarms with high, medium, and low urgency categories.
    • The combined testing and analysis of results provides assurance that the device meets its specifications and is safe and effective for its intended use.
    • Biocompatibility analysis and laboratory testing demonstrate the product to be safe for its intended use.
    • Software design and development was conducted using FDA's Reviewers Guidance of Medical Device Software Submissions, 15 Dec. 1995 draft as a guidance and per internal company requirements.
    • Environmental and electrical testing was conducted using FDA's Reviewers Guidance for Premarket Notification Submissions, Nov. 1993 draft as a guideline.
    • Performance testing was conducted using FDA's Reviewer Guidance for Ventilators draft as a guidance and per internal company requirements.
    • Device design and testing are also compliant with various voluntary international standards including: EN60601-1:1990, EN 60601-1-2:1993, CAN/CSA C22.2 No. 601-1M90:1994, UL 2601-1:1994, prEN 794-1, ISO 10651-1, and 93/42/EEC MDD.

    Therefore, a table cannot be completed with specific numerical acceptance criteria and reported performance from the provided text. The acceptance criteria are implicitly defined by compliance with the referenced standards and internal company requirements, and the reported performance is a general statement of meeting those specifications.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size used for any test set (e.g., for performance testing, software validation, or biocompatibility). It only mentions "laboratory testing" and "performance testing" without detailing the number of units tested or whether patient data was involved. Given the nature of a ventilator, the testing would primarily involve bench testing, simulated patient scenarios, and potentially animal studies (though not mentioned here), rather than human clinical trials for a 510(k) submission.

    The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective studies. The testing appears to be internal to Puritan-Bennett Corp.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided in the document. Ventilator performance testing typically relies on engineering specifications, physiological models, and recognized international standards, rather than expert consensus on interpreted output from human or animal subjects. The "ground truth" for a ventilator's performance is its ability to deliver specified gas volumes, pressures, and concentrations accurately and reliably, which is evaluated through physical measurements and simulations, not expert interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided. Adjudication methods are typically used in clinical studies where human interpretation of medical images or patient outcomes requires multiple expert opinions to establish a ground truth. For a ventilator, performance is assessed objectively through measurement against engineering specifications and regulatory standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical images, often with AI assistance. The 840 Ventilator is a treatment device, not a diagnostic imaging device with human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable to a ventilator as a "standalone algorithm performance." The ventilator itself is a complex system involving mechanical, electrical, and software components. Its "performance" is inherently a standalone system performance in delivering ventilation, albeit always used with human oversight (clinician-in-the-loop). Verification and validation testing (including software testing) would have evaluated the device's functions independently and as a whole system, without human intervention in the actual operation during those tests, to ensure it performs according to specifications. The document mentions software design and development, including verification and validation testing, which encompasses evaluating the "algorithm only" aspects of the software's control logic.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the 840 Ventilator, as inferred from the document, is based on:

    • Engineering Specifications: The device's design specifications for delivering specific tidal volumes, pressures, flow rates, oxygen concentrations, and alarm parameters.
    • International Standards: Compliance with established international standards for medical electrical equipment (e.g., EN60601-1), ventilator-specific standards (e.g., ISO 9703 series, prEN 794-1, ISO 10651-1), and FDA guidance documents.
    • Predicate Device Performance: The demonstrated safety and effectiveness of the existing predicate devices to which the 840 Ventilator claims substantial equivalence.

    8. The sample size for the training set

    This information is not applicable and not provided. Training sets are relevant for machine learning algorithms. While the 840 Ventilator contains microprocessors and software, it's not described as using machine learning that would require a "training set" in the modern sense. Its software likely operates based on deterministic control logic and pre-programmed parameters, validated against known physical models and standards.

    9. How the ground truth for the training set was established

    As there is no mention of a training set for machine learning, this question is not applicable based on the provided information. If the device's software utilizes a form of adaptive control or predictive modeling, the "ground truth" for calibrating or validating such components would have been established through a combination of theoretical models, bench testing with known inputs, and potentially pre-clinical (animal) or simulated physiological data, but this is not detailed in the 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1