Found 2 results

510(k) Data Aggregation

    K Number
    K232532

    Validate with FDA (Live)

    Date Cleared
    2023-11-02

    (73 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use: The Nd: Yag Laser Therapy Systems is indicated for the incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.
    532nm wavelength:
    -Tattoo removal: light ink (red, tan, purple, orange, skyblue, green)
    -Removal of Epidermal Benign Pigmented Lesions
    -Removal of Minor Benign Vascular Lesions including but not limited to telangiectasias
    -Treatment of Lentigines
    -Treatment of Cafe-Au-Lait
    -Treatment of Seborrheic Keratoses
    -Treatment of Post Inflammatory Hyper-Pigmentation
    -Treatment of Becker's Nevi, Freckles and Nevi Spilus
    1064nm wavelength:
    -Tattoo removal: dark ink (black, blue and brown)
    -Removal of Nevus of Ota
    -Removal or lightening of unwanted hair with or without adjuvant preparation.
    -Treatment of Common Nevi
    -Skin resurfacing procedures for the treatment of acne scars and wrinkle

    Device Description

    Nd:Yag Laser Therapy Systems is a kind of laser pigment treatment instrument. It uses the principle of "photo-blasting". The laser emits high energy instantaneously, causing the pigment particles to absorb the laser energy and rapidly expand and rupture. Part of the pigment is broken down into smaller particles and excreted from the body, and the other part is swallowed by human macrophages and excreted through the lymphatic system.
    The Nd: Yag laser therapy systems is mainly composed of a host, an operating hand tool and a foot switch. The operation hand is a treatment hand. The host is composed of a power module, a control module and a display module.

    AI/ML Overview

    I apologize, but I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria based on the provided text.

    The document you provided is an FDA 510(k) clearance letter and an associated 510(k) Summary for a Nd:Yag Laser Therapy System. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving specific performance criteria through clinical studies in the way you've described for, for example, an AI/ML software as a medical device.

    Here's why the requested information isn't present in this document:

    • Type of Device: This is a laser hardware device, not an AI/ML algorithm. The performance evaluation focuses on physical parameters (wavelength, pulse energy, etc.) and compliance with electrical, EMC, and laser safety standards.
    • 510(k) Process: The 510(k) pathway (Traditional, Abbreviated, or Special) primarily relies on demonstrating that the new device is "substantially equivalent" to a predicate device. This often involves comparing technological characteristics, intended use, and non-clinical performance data (like compliance with recognized standards).
    • Lack of Clinical Studies: The "10. Clinical Test Conclusion" section explicitly states: "No clinical study is included in this submission." This means there was no human clinical trial to assess the device's effectiveness or safety in patients, which would typically be required for an AI/ML device to demonstrate performance against acceptance criteria in a real-world scenario.
    • No AI/ML Specifics: There's no mention of AI/ML, image analysis, diagnostic performance metrics (like sensitivity, specificity, AUC), or human reader studies (MRMC, standalone algorithm performance) because this device does not perform those functions.

    Therefore, the document does not contain the information required to answer your specific questions regarding acceptance criteria, test sets, ground truth establishment, expert adjudication, or MRMC studies for an AI/ML device.

    The tables provided (Table 1: General Comparison, Table 2: Performance Comparison, Table 3: Safety Comparison) compare the characteristics of the proposed device to a predicate device and a reference device to demonstrate substantial equivalence, focusing on physical and safety parameters, not diagnostic or AI performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193464

    Validate with FDA (Live)

    Device Name
    ND:YAG Laser
    Date Cleared
    2020-04-13

    (119 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ND: YAG Laser is indicated for the treatment of: benign cutaneous lesions, such as Warts, Scars, Striae and Psoriasis; benign pigmented lesions, such as Lentigines, Nevus, and the removal of black or blue tattoos.

    Device Description

    The ND YAG Q-switch Laser Therapy Machine is laser system which delivers laser at a wavelength 1064nm or 532nm.

    The device is indicated for the treatment of: benign cutaneous lesions, such as Warts, Scars, Striae and Psoriasis; benign pigmented lesions, such as Lentigines, Nevus, and Birthmark; and the removal of black or blue tattoos. It is for prescription use only.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for a medical device called the "ND:YAG Laser" (Model SL-NY602) from Beijing Superlaser Technology Co., Ltd. It seeks to demonstrate substantial equivalence to a predicate device, the "ND YAG Q-switch Laser Therapy Machine" (K161926) from Beijing ADSS Development Co., Ltd.

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    Strictly speaking, this document does not present "acceptance criteria" in the traditional sense of specific performance metrics with defined thresholds for sensitivity, specificity, accuracy, etc. Instead, it aims to demonstrate substantial equivalence (SE) based on comparative tables for general, technical, and safety aspects. The "acceptance criteria" for this type of submission are that the proposed device is as safe and effective as the predicate device. The reported performance is implicitly the device meeting the described characteristics.

    Criterion/MetricPredicated Device K161926 Performance (Implied Acceptance)Proposed Device ND:YAG Laser PerformanceRemark
    General
    Product CodeGEXGEXSE
    Regulation No.21 CFR 878.481021 CFR 878.4810SE
    Class22SE
    Where usedhospitalhospitalSE
    Intended UseTreatment of benign cutaneous lesions (Warts, Scars, Striae, Psoriasis), benign pigmented lesions (Lentigines, Nevus, Birthmark), and removal of black or blue tattoos.Treatment of benign cutaneous lesions (Warts, Scars, Striae, Psoriasis), benign pigmented lesions (Lentigines, Nevus, Birthmark), and removal of black or blue tattoos.SE
    Technical
    Laser MediumNd:YAGNd:YAGSAME
    Wavelength1064 nm, 532 nm1064 nm, 532 nmSAME
    Output energy100-1000mJ (1064nm), 50-500mJ (532nm)100-1000mJ (1064nm), 50-500mJ (532nm)SAME
    Max. Energy Density31.8J/cm2, 15.9 J/cm231.8J/cm2, 15.9 J/cm2SAME
    Spot Size2-10mm2-10mmSAME
    Pulse Width5ns-8ns5ns-8nsDIFFERENT
    Frequency1-10 Hz1-10 HzSAME
    DisinfectionDisinfect handpiece by 75% medicinal alcoholWipe outer surface with 70% alcohol and non-abrasive medical grade anti-bacterialSIMILAR (both relate to cleaning/disinfection methods for safety)
    Laser ClassClass 4Class 4SAME
    Aiming BeamRed Laser, <6mWRed (650nm) laser, ≤5mWSAME
    Safety
    Electrical SafetyComply with IEC 60601-1, IEC 60601-2-22Comply with IEC 60601-1, IEC 60601-2-22SE
    EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2SE
    Laser SafetyComply with IEC 60601-2-22, IEC 60825Comply with IEC 60601-2-22, IEC 60825SE

    2. Sample Size used for the Test Set and the Data Provenance

    The submission does not rely on a test set of patient data for clinical performance evaluation. Instead, it relies on demonstrating equivalence through comparison to a legally marketed predicate device and non-clinical testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. No ground truth was established by experts for a clinical test set in this 510(k) submission.

    4. Adjudication Method for the Test Set

    Not applicable. There was no clinical test set requiring adjudication in this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a laser device, not an AI-assisted diagnostic or imaging system, so an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a laser device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. Clinical ground truth (e.g., pathology, outcomes data) was not used in this 510(k) submission. The "ground truth" for demonstrating substantial equivalence relied on established safety and performance standards for similar laser devices and the characteristics of the predicate device.

    8. The sample size for the training set

    Not applicable. This is a hardware medical device (laser system), not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.