    K Number
    K230510

    Validate with FDA (Live)

    Device Name
    MyEllevate
    Manufacturer
    Date Cleared
    2023-07-06

    (132 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MyEllevate is indicated for use in soft tissue approximation of sub dermis and underlying muscle

    Device Description

    The MyEllevate device consists of a Surgical Suture System consisting of synthetic non-absorbable, braided, polyester surgical suture, and a set of instruments designed to efficiently and accurately place the suture for the purpose of performing soft tissue approximation. MyEllevate contains a Skin Marking Tape, a Lancet, a Tunneling Rod, two stainless steel Suturods, and a Light Handle which contains an LED light source. Each Suturod™ has an internal fiber optic "light pipe" which, when connected to the Ligh Handle, illuminates the distal tip of the Suturod™ and permits visual verification of the location and depth of the Suturod™ as it is passed through the subcutaneous space. All components of MyEllevate are provided sterile and are intended for single use only. While the nonabsorbable suture is a Class II device as describes by 21CFR, §878.5000, the remaining instruments are Class I devices exempt from 510(k) requirements. The Lancet, Clearing Device, and Tunneling Rod are manual surgical devices described by 21 CFR §878.4800. The Skin Marking Pen is described in 21 CFR §878.4660, and the fiber optic "light pipe" and Light Handle to illuminate surgical sites, is described in 21 CFR, §876.4530.

    AI/ML Overview

    The provided document is a 510(k) summary for the MyEllevate device, which is a surgical suture system. This document is a premarket notification for a medical device and therefore does not contain information about an AI/ML-based device.

    The document explicitly states:

    • "807.92(b)(2) Clinical tests submitted – N/A – No clinical tests submitted" (Page 6), indicating no human performance or clinical studies were conducted for this submission.
    • The device being cleared is a "Nonabsorbable Polyester Suture" and associated manual surgical instruments (Suturods, Lancet, Tunneling Rod, Light Handle, Skin Marking Tape). There is no mention of any AI or machine learning component.

    Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving an AI/ML device meets these criteria because the submitted text is a 510(k) for a physical medical device (suture system), not an AI/ML device.