510(k) Data Aggregation

    K Number
    K231203

    Validate with FDA (Live)

    Device Name
    Mosie Baby Kit
    Manufacturer
    Date Cleared
    2023-11-29

    (216 days)

    Product Code
    Regulation Number
    884.6110
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mosie Baby Kit is indicated for over-the-counter home use, by individuals who have been unable to conceive through intercourse or choose not to conceive through intercourse, for semen collection and the delivery of semen or donor sperm to the vaginal canal. The Mosie Baby Kit should be used during the ovulatory phase of the menstrual cycle.

    Device Description

    The Mosie Baby Kit is an over-the-counter (OTC) device intended for intravaginal insemination (IVI) at home. The Mosie Baby Kit includes the following components:

    • Two Mosie syringes
    • Two Mosie collection cups
    • Instructions for Use.
      Each component is single-use only and supplied non-sterile. The collection cup is used to hold and contain the sperm and work in conjunction with the syringe when it is time to transfer the sperm into the syringe. The syringe is used to withdraw the semen from the collection cup and deploy the semen intravaginally. The syringe features a rounded front end with a slotted tip opening. The syringe has a rounded stopper that fits within the rounded end of the syringe to aid in sperm delivery from the syringe to the vagina. The syringe also contains flanges designed to aid in device handling during use.
    AI/ML Overview

    The provided text describes the regulatory clearance of the Mosie Baby Kit and includes summaries of non-clinical and clinical performance testing. However, it does not describe an AI/ML-driven medical device and therefore does not contain information related to multi-reader multi-case (MRMC) studies, ground truth establishment by experts, or the effect size of AI assistance.

    The document primarily focuses on the physical and functional characteristics of a medical device (a home insemination kit) and regulatory requirements for its approval for over-the-counter use.

    Based on the provided document, here's an attempt to answer the questions, highlighting where the information is not applicable to AI/ML devices:


    Acceptance Criteria and Device Performance (Based on Mosie Baby Kit Document)

    Since this is not an AI/ML device, the acceptance criteria are not in terms of AI model performance metrics like sensitivity, specificity, or AUC. Instead, they relate to the physical and functional properties of the device, as well as user comprehension and self-selection for an over-the-counter product.

    Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Target)Reported Device Performance
    Shelf Life/StabilityBioburden:Met bioburden specifications for vaginal use in accordance with USP <111> at baseline and end-of-shelf-life: - TAMC: <10^2 cfu/g- TYMC: <10^1 cfu/g- Absence of Pseudomonas aeruginosa, Staphylococcus aureus, Candida albicans
    Sperm Compatibility (HSSA): ≥80% of control motility at 24 hoursMet HSSA acceptance specification (≥80% of control motility at 24 hours) at all time points assessed (baseline and end-of-shelf-life).
    BiocompatibilityCytotoxicity: Non-cytotoxicDemonstrated to be non-cytotoxic.
    Sensitization: Non-sensitizingDemonstrated to be non-sensitizing.
    Irritation: Non-irritating (vaginal)Demonstrated to be non-irritating (vaginal).
    Functional PerformanceSyringe Functional Tests (ISO 7886-1:2017 based):All tested parameters (extraneous matter, barrel dimensions, barrel flanges, plunger stopper/assembly design, freedom from liquid leakage, force to operate piston, fit of plunger stopper/plunger) were assessed at baseline and end-of-shelf-life, implying they met specifications (though specific performance numbers are not given, only that tests were conducted).
    Collection Cup Functional Tests: Visual & DimensionalAssessed at baseline and end-of-shelf-life, implying they met specifications.
    User Comprehension/SafetySelf-Selection & Labeling Comprehension:All 32 participants (15 contraindicated, 17 non-contraindicated) were able to accurately identify: who the kit is indicated/intended for, whether it was acceptable for them to use the product and why, what difficulty they had, who should not use the kit, and what the product does not do.
    Human Factors Validation (Critical Tasks):All questions related to critical information in the Instructions for Use (IFU) were answered correctly by at least 90% of participants, and none reported difficulty understanding. The study validated that potential users could successfully use the kit with labeling information and that performance of critical tasks would not result in use errors or harm. The study covered when/how often to use, when to contact physician, semen sample collection/storage, syringe handling, post-insemination actions, and disposal.

    Detailed Study Information (Addressing AI/ML-specific questions where applicable, or stating N/A)

    1. A table of acceptance criteria and the reported device performance:

      • See the table above.
    2. Sample sizes used for the test set and the data provenance:

      • Shelf-Life, Bioburden, HSSA, Functional Testing: Test articles were subjected to accelerated aging conditions. Specific sample sizes for each test are not explicitly stated, but are implied to be sufficient for the tests mentioned (e.g., "samples" for HSSA, "test articles" for bioburden).
      • Biocompatibility Testing: Not specified as a "test set" in the context of an AI model, but rather as individual tests performed on device materials. Specific sample sizes for cytotoxicity, sensitization, and irritation tests are not provided.
      • Clinical Performance Testing (User Studies):
        • Self-Selection Study: 32 participants (15 contraindicated female participants and 17 non-contraindicated female participants).
        • Human Factors Validation Study: 30 participants (15 self-administered, 15 administered to a female recipient).
      • Data Provenance: The document states that participants in the clinical performance studies were "representative of the intended user in the United States." The studies appear to be prospective user studies conducted to evaluate the device's usability and label comprehension.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • N/A for this device. This device is not an AI/ML diagnostic tool that requires expert-established ground truth for image/data interpretation. Ground truth for the user studies was based on the defined indications for use, contraindications, and correct understanding/performance of tasks as defined by the manufacturer and assessed by the study design.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • N/A for this device. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in expert-based review of medical images/data, which is not relevant here. The user studies' "ground truth" was based on whether participants correctly understood instructions or performed tasks as intended, as assessed by study protocols.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC study was not done. This is not an AI/ML diagnostic device, so a study comparing human reader performance with and without AI assistance is not applicable.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • N/A for this device. This device is a physical product (insemination kit), not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the physical tests (bioburden, HSSA, biocompatibility, functional), the "ground truth" was defined by established laboratory testing standards (e.g., USP, ISO, ASTM) and manufacturer specifications.
      • For the user studies (self-selection, human factors), the "ground truth" was based on predefined correct answers for comprehension questions and successful completion of tasks as outlined in the Instructions for Use (IFU) and verified by study observers. Essentially, it's a task-based performance and comprehension ground truth as defined by the device's labeling and intended use.
    8. The sample size for the training set:

      • N/A. This is not an AI/ML device that requires a training set. The data presented are from verification and validation studies of the physical device and user interaction.
    9. How the ground truth for the training set was established:

      • N/A. As there is no training set for an AI/ML model, this question is not applicable.