Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K213490

    Validate with FDA (Live)

    Date Cleared
    2022-04-01

    (151 days)

    Regulation Number
    870.1025
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The monitor B105M, B125M, B155M, B105P and B125P are portable multi-parameter patient monitors intended to be used for monitoring, recording, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.

    The monitor B105M, B155M, B105P and B125P are intended for use under the direct supervision of a licensed health care practitioner.

    The monitor B105M, B155M, B105P and B125P are not intended for use during MRI.

    The monitor B105M, B155M, B105P and B125P can be stand-alone monitors or interfaced to other devices via network.

    The monitor B105M, B125M, B155M, B105P and B125P monitor and display: ECG (including ST segment, arrhythmia detection, ECG diagnostic analysis and measurement), invasive blood pressure, heart/pulse rate, oscillometric noninvasive blood pressure (systolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring( including monitoring conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/ Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agent identification and respiratory rate), Cardiac Output (C.O.), Entropy and neuromuscular transmission (NMT).

    The monitor B105M, B125M, B105P and B125P are able to detect and generate alarms for ECG arrhythmias: Asystole, Ventricular tachycardia, VT>2, Ventricular Bradycardia, Accelerated Ventricular Couplet, Bigeminy, Trigeminy, "R on T", Tachycardia, Pause, Atrial Fibrillation, Irregular, Multifocal PVCs, Missing Beat, SV Tachy, Premature Ventricular Contraction (PVC), Supra Ventricular Contraction (SVC) and Ventricular fibrillation.

    Device Description

    The proposed monitors B105M, B125M, B155M, B105P and B125P are multi-parameter patient monitors that were developed based on predicate Monitor B125/B105(K201941) to provide additional monitored parameter: neuromuscular transmission (NMT), by supporting additional optional modules previously cleared by FDA: E-NMT module (K051635) with existing interface rack and/or second frame (B1X5-F2).

    In addition to the added parameter, the proposed monitors B105M, B125M, B155M, B105P and B125P offer several software enhancements:

    • Support 12-lead ECG measurement mode;
    • Additional SPV (Systolic Pressure Variation) and PPV (Pulse Pressure Variation) values calculation;
    • Enabled the Impedance Respiration measurement from lead RL-LL:
    • Display Pulse Rate (PR) from NIBP when performing NIBP determination;
    • Display real-time GE SpO2 Perfusion Index (PI) value; .
    • Adoption of TruSignal V3 SpO2 algorithm;
    • Additional optimizing IBP waveform scale function;
    • Additional connectivity capabilities within GE CARESCAPE network (K032582) including Bed-to-Bed View and Automatic View on Alarm (AVOA);
    • Additional remote service function:
    • Additional cybersecurity enhancements.

    The proposed monitors B105M, B125M, B155M, B105P and B125P include improved Industrial Design (ID) to be more portable and more compact for clinicians than the primary predicate Monitor B125/B105(K201941) while maintaining the same primary function and operation.

    The five models (B105M, B125M, B155M, B105P and B125P) share the same hardware platform and software platform to support the data acquisition and algorithm modules. The differences between them are the LCD screen size and configuration options.

    As with the predicate Monitor B125/B105 (K201941), the proposed monitors B105M, B125M, B155M, B105P and B125P are multi-parameter patient monitors, utilizing an LCD display and pre-configuration basic parameters: ECG, RESP, NIBP, IBP, TEMP, SpO2, and optional parameters whic include CO2 and Gases parameters provided by the E-MiniC module (K052582), CARESCAPE Respiratory modules E-sCO and E-sCAiO (K171028), Airway Gas Option module N-CAiO (K151063), Entropy parameter provided by the E-Entropy module (K150298), Cardic Output parameter provided by the E-COP module (K052976), and thermal recorder B1X5-REC.

    In addition, the predicate Monitor B125/B105 (K201941) and the proposed monitors B105M, B125M, B155M, B105P and B125P consists of the same interface to a variety of existing central station systems via a cabled or wireless network interface which implemented with identical integrated WIFI module. (WIFI feature is disabled in B125P/B105P)

    Moreover, both the predicate Monitor B125/B105(K201941) and the proposed monitors B105M, B125M, B155M, B105P and B125P can be mounted in a variety of ways (e.g. shelf, countertop, table, wall, pole, or head/foot board) using existing mounting accessories.

    AI/ML Overview

    The provided document is an FDA 510(k) Premarket Notification for a patient monitor. It does not describe a study involving an AI-driven device or an equivalent. Therefore, I cannot extract the information required to answer your prompt, as the prompt is geared towards AI/ML-based medical devices with specific performance metrics and validation studies.

    The document focuses on demonstrating substantial equivalence of the new monitor models (B105M, B125M, B155M, B105P, B125P) to existing predicate devices (Monitor B125/B105, CARESCAPE B450, CARESCAPE VC150, CARESCAPE ONE). The changes primarily involve physical design, software enhancements (display modes, calculation additions, connectivity), and support for an additional, previously cleared, NMT measurement module. The document explicitly states:

    "The subject of this premarket submission, the proposed monitors B105M, B125M, B155M, B105P and B125P did not require clinical studies to support substantial equivalence." (Page 16)

    This indicates that the device's performance was evaluated through bench testing, cybersecurity assessments, and verification of software and hardware changes against established standards, rather than through comparative effectiveness studies with human readers or standalone algorithm performance studies typical for AI/ML devices. Therefore, a table of acceptance criteria and proven performance for an AI/ML component, sample sizes for test sets, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, and training set information are not applicable or present in this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1