Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K231809

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-10-20

    (122 days)

    Product Code
    Regulation Number
    888.3027
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Momentum® Posterior Spinal Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracolumbar and sacroiliac spine. When used as a posterior system, Momentum is intended for the following indications: degenerative disc disease, spinal stenosis, spondylolisthesis, spinal deformities (i.e., scoliosis, and/ or lordosis), trauma (i.e., fracture or dislocation), tumor, pseudoarthrosis and failed previous fusion.

    When used in conjunction with G21 V-Steady Bone Cement, the Momentum Posterior Spinal Fixation System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. Momentum Posterior Spinal Fixation System Screws augmented with G21 V-Steady Bone Cement are limited for use at spinal levels where the structural integrity of the spine is not severely compromised. Iliac screws are not intended to be used with bone cement.

    In order to achieve additional levels of fixation the Momentum Posterior System can also be connected to the neon38 universal OCT spinal stabilization system via transition rods or connectors. Please refer to the neon3 Instructions for Use for a list of indications for use.

    The Momentum® MIS Posterior System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracolumbar and sacroiliac spine. When used as a posterior spine system, Momentum MS is intended for the following indications: degenerative disc disease, spinal stenosis, spondylolisthesis, spinal deformities (i.e., scoliosis, and or lordosis), trauma (i.e., fracture or dislocation), tumor, pseudoarthrosis and failed previous fusion.

    When used in conjunction with G21 V-Steady Bone Cement, the Momentum MIS Posterior System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. Momentum MIS Posterior System Screws augmented with G21 V-Steady Bone Cement are limited for use at spinal levels where the structural integrity of the spine is not severely compromised. Iliac screws are not intended to be used with bone cement.

    Device Description

    The Momentum® Posterior Spinal Fixation System, and Momentum® MIS Posterior Spinal Fixation System are standard pedicle screw and rod systems designed for fixation in the thoracolumbar and sacroiliac spine. There have been no major changes to the device since K191932, and K223274, respectively. The subject devices are identical to the previously cleared versions with the addition of the option to use bone cement (G21 V-Steady) with the fenestrated screws.

    G21 V-Steady Bone Cement consists of a liquid and powder component. The powder component is constituted of PMMA beads shaped particles containing the initiator benzoy1 peroxide required for starting to initiate the cement curing. The radiopacifier agent, zirconium dioxide, is necessary for the cement visibility under radiographs but it does not take part of the curing process (radical polymerization). The liquid component comprises the monomer, methylmemethylmethacrylate (MMA); dimethyl-para-toluidine (DMPT) as polymerization accelerator and hydroquinone (HQ) as stabilizer to prevent polymerization of the liquid during storage. V-STEADY bone cement has an immediate development of viscosity and thus it is a high viscosity cement that maintains its properties throughout the useful working time.

    AI/ML Overview

    While the provided text describes the regulatory clearance of a medical device (Momentum® Posterior Spinal Fixation System with G21 V-STEADY Bone Cement), it does not contain any information regarding acceptance criteria, performance studies (other than "bone cement usability testing and screw removal testing"), or detailed clinical data typically associated with AI/ML device evaluations.

    The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device based on indications for use, technological characteristics, and non-clinical performance (material composition, manufacturing, basic mechanical testing). It is not a clinical study report for an AI/ML diagnostic or prognostic tool.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them, specifically regarding:

    1. A table of acceptance criteria and the reported device performance: This information is not present. The document states "bone cement usability testing and screw removal testing was conducted to validate the use," but no specific metrics or acceptance criteria are given.
    2. Sample sized used for the test set and the data provenance: Not applicable to the type of testing described (usability and screw removal, likely biomechanical or benchtop).
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is a physical medical device, not an AI/ML-driven diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not clearly defined for the "usability" and "screw removal" tests, but would likely refer to engineering specifications or validated test methods.
    8. The sample size for the training set: Not applicable (no AI/ML training set).
    9. How the ground truth for the training set was established: Not applicable (no AI/ML training set).

    In summary, the provided document is a regulatory submission for a physical medical device (spinal fixation system with bone cement) and does not contain the characteristics of an AI/ML device study that your questions refer to.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1