Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K233798

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-02

    (247 days)

    Product Code
    Regulation Number
    876.5860
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Moda-flx Hemodialysis System™ is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute, post-acute, or chronic care facility. Treatments must be administered under a physician's prescription, by a trained person who is considered competent in the use of the device.

    Treatment types available include: Intermittent Hemodialysis (IHD), Sustained Low Efficiency Dialysis (SLED/ SLEDD), and Prolonged Intermittent Renal Replacement Therapy (PIRRT).

    The Moda-flx Hemodialysis System™ Cartridge is a single use, disposable arterial and venous bloodline set, with dialysate tubing, intended to provide extracorporeal access during hemodialysis. The Moda-flx Hemodialysis System™ Cartridge is compatible only with the Moda-flx Hemodialysis System TM.

    Device Description

    The Moda-flx Hemodialysis System™ is a transportable hemodialysis system that consists of three parts: Hemodialysis Delivery (HD) Device. Dialysate Generator (DG) Device, and a single use Moda-flx Hemodialysis System™ Cartridge that provides extracorporeal access during hemodialysis. The Moda-flx Hemodialysis System™ Cartridge is only compatible with the Modaflx Hemodialysis System™. The system creates purified water from tap water; from this purified water and chemical concentrates, the system creates dialysate in real time. The chemicals used to create the dialysate are standard 45X concentrates.

    Ultrafiltration is achieved by controlling the dialysate going in and coming out of the dialyzer independently, using pumps on the HD Device to control fluid flow from both the inlet and outlet of the dialyzer. An external bag of saline is used to provide solution infusions to combat hypotensive episodes. Saline is also used to prime the Moda-flx Hemodialysis System™ Cartridge prior to use.

    Prescribed blood flow rate is generated by a peristaltic pump, while heparin is administered to prevent blood coagulation. All fluid pathways are disposable and easily installed and replaced. Once installed, the device will check the connections of the fluid pathways and the system will automatically prepare itself for treatment.

    AI/ML Overview

    This document is a 510(k) summary for the Moda-flx Hemodialysis System and Cartridge. It outlines the device's indications for use, its description, and a comparison to predicate devices, but does not provide the acceptance criteria or results of a study designed to prove the device meets specific acceptance criteria related to its clinical performance.

    The "Performance Data" section lists various tests conducted, such as biocompatibility, electrical safety, software verification, and bench performance (including clearance, transport, and dialyzer integrity testing). It explicitly states that "no clinical studies were required to support this submission." This indicates that the FDA's substantial equivalence determination for this device was based on non-clinical performance data and comparison to predicate devices, rather than a clinical study demonstrating specific efficacy or clinical performance acceptance criteria.

    Therefore, I cannot provide the requested information about acceptance criteria and a study proving the device meets them, as such a study (specifically a clinical one with performance metrics) is stated as not being required and thus not provided in this document.

    However, based on the information provided, here's what can be inferred about the types of performance criteria and tests performed, even if specific numerical acceptance values aren't given:

    Inferred Information from the Document:

    While specific clinical acceptance criteria are not detailed (as no clinical study was required for this submission), the document does list various performance data provided in support of substantial equivalence. These imply underlying performance criteria for safety and technical functionality.

    1. Table of Acceptance Criteria and Reported Device Performance: This information is not provided in the document for clinical performance. The document only lists categories of tests performed.

    2. Sample size used for the test set and the data provenance: This information is not provided in the document for any of the listed performance tests. The provenance (e.g., country of origin, retrospective/prospective) is also not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not applicable/provided, as no clinical study with "ground truth" established by experts is mentioned. For the non-clinical tests (e.g., bench testing, software V&V), the "ground truth" would be the engineering specifications and regulatory standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: This information is not applicable/provided, as no clinical study requiring adjudication is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This information is not applicable/provided. The device is a hemodialysis system, not an AI-assisted diagnostic tool involving human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: This information is not applicable/provided. The device is a medical system, not an AI algorithm in the typical sense of "standalone performance" for diagnostic or predictive tasks. Performance was likely evaluated for the integrated system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the non-clinical tests mentioned (biocompatibility, electrical safety, software V&V, bench performance, etc.), the "ground truth" would be established by relevant engineering standards, regulatory requirements, and internal specifications for the device's functional performance. No clinical ground truth (like pathology or outcomes data) was used or required for this submission according to the document.

    8. The sample size for the training set: This information is not applicable/provided. As no clinical study requiring "training data" for an AI model is described, this question is not relevant to the provided text.

    9. How the ground truth for the training set was established: This information is not applicable/provided for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1