Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K193548

    Validate with FDA (Live)

    Date Cleared
    2020-01-18

    (29 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hydrophilic Guidewire is indicated for general intravascular use to aid in the selective placement of catheters in the peripheral, visceral and cerebral vasculature during diagnostic and/or therapeutic procedures.

    Device Description

    The Hydrophilic Guidewire is a stainless-steel guidewire with a radiopaque, platinum distal coil. The guidewire has a hydrophilic coating that spans the distal 170 cm. Included within the sterile pouch is a pin vise to assist in guidewire manipulation and an introducer needle to ease the introduction of the guidewire into the catheter hub and/or hemostasis valve.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Mirage™ Hydrophilic Guidewire and X-pedion™ Hydrophilic Guidewire. This document primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical (bench and biocompatibility) testing, rather than a study on an AI/ML-based medical device.

    Therefore, the information required to answer most of your questions about acceptance criteria for an AI/ML device, sample sizes for AI test sets, expert ground truthing, MRMC studies, standalone performance, training sets, and ground truth establishment for AI/ML models is not present in this document.

    However, I can extract information related to the acceptance criteria and performance data for the physical medical device as described:


    Acceptance Criteria and Device Performance (Based on Non-Clinical Bench Testing):

    The document details numerous non-clinical bench tests performed to demonstrate the safety and performance of the guidewires. The results consistently state that "The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for [Test Name]."

    Table of Acceptance Criteria and Reported Device Performance (Non-Clinical):

    Test CategorySpecific Test / ParameterAcceptance Criteria (Implied by positive outcome)Reported Device Performance
    Biocompatibility
    Chemical Characterization (Extractables/Leachables)No risk to the patient from Chemicals of Potential Concern (COPC).The extractables/leachables found are acceptable.
    CytotoxicityNo evidence of cell lysis or toxicity; grade < 2 (Mild).Considered non-cytotoxic.
    SensitizationNo delayed dermal contact sensitization.Does not elicit a sensitization response.
    IrritationDifference between test and control mean score = 0.0.Considered a non-irritant.
    Acute Systemic ToxicityNo mortality or systemic toxicity from extracts.Do not indicate signs of toxicity.
    Hemocompatibility (Hemolytic Index)Hemolytic index for direct contact < threshold (e.g., 0.5% / 0.6%).Hemolytic index was 0.5% (direct contact) and 0.6% (extract).
    Hemocompatibility (Complement System)Not considered to be potential activators of the complement system.Control and test articles are not considered to be potential activators.
    Hemocompatibility (Thrombogenic Potential)Acceptable thrombogenic potential.Demonstrates moderate thrombus formation (less than control article).
    PyrogenicityNo temperature rise of 0.5°C or more above baseline.No animal showed a temperature rise of 0.5°C or more.
    Microbial
    Ethylene Oxide (EO) ResidualMet criteria per ISO 10993-7.Met the acceptance criteria for ethylene oxide residual.
    Ethylene Chlorohydrin (ECH) ResidualMet criteria per ISO 10993-7.Met the acceptance criteria for ethylene chlorohydrin residual.
    Bioburden RecoveryMet criteria per ISO 11737-1.Met the acceptance criteria for bioburden recovery.
    BioburdenMet criteria per ISO 11737-1.Met the acceptance criteria for bioburden.
    Bacterial EndotoxinMet criteria per ANSI/AAMI ST72, USP <161>, and USP <85>.Met the acceptance criteria for bacterial endotoxin.
    Packaging
    Terminally Sterilized Medical DevicesMet criteria per ISO 11607.Met the acceptance criteria for packaging terminally sterilized medical devices.
    Performance (Bench)
    Device Compatibility/Distal AccessCompatible with ancillary devices, navigates tortuous vessels.Met the acceptance criteria for device compatibility/distal access.
    Distal FlexibilityNavigates through tortuous vessels.Met the acceptance criteria for distal flexibility.
    Visual FractureMet criteria per ISO 10555-1 and ISO 11070.Met the acceptance criteria for visual fracture.
    RadiopacityClearly visible during use.Met the acceptance criteria for radiopacity.
    Tip BucklingWithstands forces typical of clinical use.Met the acceptance criteria for tip buckling.
    Tip RetentionSatisfactory tip retention.Met the acceptance criteria for tip retention.
    Tip ShapeabilitySatisfactory tip shapeability.Met the acceptance criteria for tip shapeability.
    Torque ResponseDistal tip responds to proximal manipulations.Met the acceptance criteria for torque response.
    Turns to FailureWithstands torsional forces typical of clinical use.Met the acceptance criteria for turns to failure.
    Visual Inspection and Dimensional VerificationMeets specified dimensions.Met the acceptance criteria for visual inspection and dimensional verification.
    ParticulateMeets criteria per USP <788> for particulate counts and sizes.Met the acceptance criteria for particulate counts and sizes.
    Friction ForceSatisfactory friction force.Met the acceptance criteria for friction force.
    FlexingMet criteria per ISO 11070.Met the acceptance criteria for flexing.
    Corrosion ResistanceMet criteria per ISO 11070.Met the acceptance criteria for corrosion resistance.
    Tensile StrengthMet criteria per ISO 11070.Met the acceptance criteria for tensile strength.

    Since this is a 510(k) submission for a non-AI/ML medical device, the following information is not applicable or not provided in the document:

    1. Sample sizes used for the test set and the data provenance: Not an AI device, so no "test set" in the AI sense. Bench tests follow specific standards, implying sample sizes per those standards, but not explicitly stated here.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable; ground truth for a physical device is established via standardized physical and chemical tests, not expert human interpretation of data for AI.
    3. Adjudication method for the test set: Not applicable for non-AI bench testing.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used: For a physical guidewire, "ground truth" is typically defined by adherence to engineering specifications, material properties, and performance under simulated clinical conditions, verified through empirical testing following established ISO and USP standards.
    7. The sample size for the training set: Not applicable; this is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable; this is not an AI/ML device.

    In summary, this document is a regulatory submission for a physical medical device demonstrating substantial equivalence, not an AI/ML device requiring clinical validation for diagnostic or prognostic purposes.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1