510(k) Data Aggregation

    K Number
    K170381

    Validate with FDA (Live)

    Date Cleared
    2017-09-07

    (212 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aquamantys™ Minimally Invasive Sealer (MIS) FLEX and Aquamantys™ Minimally Invasive Sealer (MIS) Flex Mini are sterile, single-use bipolar electrosurgical devices intended to be used in conjunction with a qualified Pump Generator for delivery of Radio-frequency (RF) energy and saline for hemostatic sealing and coagulation of soft tissue and bone at the operative site. They are intended for, but not limited to orthopaedic, neurosurgical, spine, endoscopic procedures, abdominal and thoracic surgery, and epidural vein sealing during surgery. These devices are not intended for contraceptive tubal coagulation (permanent female sterilization).

    Device Description

    MIS FLEX and MIS FLEX Mini are single-use, disposable, bipolar devices. The MIS FLEX and MIS FLEX Mini employ radio-frequency (RF) energy and saline irrigation (termed Transcollation Technology™) for hemostatic sealing and coagulation. The devices are equipped with dual electrode tips while saline and electrical lines exit the opposite end of the handpiece from the electrodes. The handpiece is equipped with an on-off button that simultaneously activates both RF and saline flow. A saline fluid delivery line is provided with the device and includes a section of pump tubing and drip chamber or spike. The three-pin electrical connector is designed to be plugged into either of the FDA cleared AEX (K143175) or Aquamantys (K052859) Pump Generators. The MIS FLEX devices feature a fully malleable shaft, altered saline delivery and dual ports for saline delivery (two for each electrode). The distal end of the device features small, rounded electrode geometry. The MIS FLEX and MIS FLEX Mini differ by the length of the shaft. The MIS FLEX is designed for transsphenoidal and MIS Spine procedures while the MIS FLEX Mini is designed for craniotomies with application in posterial cervical and other spinal procedures.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a study proving the device meets specific acceptance criteria in the manner typically seen for performance claims like accuracy, sensitivity, or specificity. This document is a 510(k) summary for a premarket notification, which focuses on demonstrating substantial equivalence to a predicate device rather than proving novel performance against acceptance criteria.

    However, based on the information provided, I can infer the "acceptance criteria" were primarily related to safety, electrical performance, biocompatibility, and functional access. The "study" refers to the non-clinical testing performed to demonstrate these aspects.

    Here's an interpretation based on the document's content:

    1. Table of Acceptance Criteria (Inferred) and Reported Device Performance

    Acceptance Criteria Category (Inferred)Specific Standard (Inferred)Reported Device Performance
    BiocompatibilityISO-10993-1, -4, -5, -10, -11Met requirements (Biocompatibility evaluation conducted, testing included in-vitro cytotoxicity, acute system toxicity, irritation/intracutaneous reactivity, sensitization, pyrogenicity. Considered externally communicating, limited exposure < 24 hours).
    SterilizationISO 11737-1, ISO 11135:2014Met requirements (Testing conducted).
    Shelf LifeN/A (Standard not explicitly listed)Met requirements (Testing conducted).
    Electrical SafetyIEC 60601-1, IEC 60601-2-2In compliance (Electrical safety and EMC testing were conducted and found to be in compliance with associated standards).
    Electromagnetic Compatibility (EMC)IEC 60601-1, IEC 60601-2-2In compliance (Electrical safety and EMC testing were conducted and found to be in compliance with associated standards).
    Mechanical TestingN/A (Standard not explicitly listed)Met requirements (Testing conducted).
    Performance Testing - Anatomical AccessCustomer needs for neurosurgical and spine proceduresMet requirements (Access study conducted to ensure ability to access key anatomy).
    Performance Testing - Thermal Effect"Premarket Notification (510(k) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff"Substantially equivalent to the predicate device (Comparative performance testing assessed thermal effect in in-vivo porcine model and ex-vivo tissue).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for most tests.

    • Cadaver Study (Anatomical Access): "The access study was conducted to ensure that the MIS FELX will be able to access key anatomy..." No sample size (number of cadavers) is given. Data provenance is not specified beyond being an "access study."
    • In-vivo Porcine Model (Comparative Performance): "The second study was a comparative performance test, which was conducted in an in-vivo porcine model." No specific number of animals is given. Data provenance is "in-vivo porcine model."
    • Ex-vivo Tissue Testing (Comparative Performance): "Ex-vivo tissue testing was also completed." No sample size (number of tissue samples) is given. Data provenance is "ex-vivo tissue."
    • Biocompatibility, Sterilization, Shelf Life, Electrical Safety, EMC, Mechanical Testing: These are generally performed on device units. The specific number of units tested is not provided, which is typical for a 510(k) summary. Data provenance would be laboratory testing.

    All testing described appears to be prospective, laboratory-based, or animal-based, as part of the device development and validation process. The country of origin of the data is not specified but is implicitly from the manufacturer's testing facilities or contract labs.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    • Cadaver Study (Anatomical Access): The document mentions ensuring access to "key anatomy and meet customer needs for neurosurgical and spine procedures." This implies involvement of medical experts (e.g., neurosurgeons, spine surgeons). However, the number of experts, their qualifications, or how they established "ground truth" for access (e.g., by subjective evaluation, objective measurement of reach) is not specified.
    • For other tests (biocompatibility, electrical safety, thermal effect), "ground truth" is typically defined by adherence to recognized standards and objective measurements, rather than expert consensus on a test set.

    4. Adjudication Method for the Test Set

    Not applicable in the conventional sense for a device of this type, especially for performance studies described. The "acceptance" was based on meeting pre-defined requirements or demonstrating substantial equivalence to a predicate, not on a human-expert adjudication of results. For the access study, evaluation criteria for "meeting needs" would have been set by the manufacturer, possibly in consultation with clinicians, but an adjudication method is not described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. The document describes a "comparative performance test" for thermal effect, but this was a technical comparison of the device against a predicate in animal and ex-vivo models, not a study evaluating human readers' performance with and without AI assistance. This device is an electrosurgical tool, not an AI-assisted diagnostic or interpretative system that would typically warrant an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • This question is not applicable as the device is not an algorithm or AI system. It is a physical electrosurgical device. The "standalone" performance refers to its direct functional capabilities (e.g., thermal effect, access).

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • Biocompatibility: Adherence to ISO standards, measured biological responses (e.g., cytotoxicity, irritation, sensitization).
    • Electrical Safety & EMC: Adherence to IEC standards, objective electrical measurements.
    • Performance (Thermal Effect): Objective measurements of thermal effect (e.g., temperature, tissue charring, coagulation depth) in animal and ex-vivo models, compared against the predicate and potentially against defined specifications derived from predicate performance. The "ground truth" is primarily the measured physical effect relative to established benchmarks and predicate performance.
    • Performance (Anatomical Access): Likely subjective or objective assessment of the device's ability to reach target anatomical locations in cadavers, possibly based on expert opinion/feedback from surgeons.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a physical electrosurgical instrument, not an AI or machine learning algorithm. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as there is no training set for this type of device.