Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K230075

    Validate with FDA (Live)

    Date Cleared
    2023-06-28

    (169 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Upper Arm Blood Pressure Monitor, Model BP3K V 1-5K is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm for a circumference range from 17 to 52cm.

    The device detects the appearance of irregular heartbeat during measurement and gives a warning signal with the reading once the irregular heartbeat is detected.

    The device can be used in connection with a smart phone via Bluetooth or with a personal computer (PC) via USB. The measurement data can be transferred to a smart phone running the Microlife Connected Health+ mobile software (App) or a PC running the Microlife BP Analyzer+ (BPA+) software.

    Device Description

    Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KV1-5K is designed to measure systolic and diastolic blood pressure, pulse rate of an individual with arm circumference sizes ranging from 17-52 cm by using a noninvasive technique in which one inflatable cuff is wrapped around the single upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but using a semiconductor sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".

    The device detects the appearance of irregular heartbeat during measurement, and the symbol " " is displayed with the reading. The device can be used in connection with smart mobile devices running the APP and via Bluetooth.

    The blood pressure monitor is a fully automatic digital blood pressure measuring device use by adults on the upper arm at home.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a blood pressure monitor, not an AI/ML powered medical device. Therefore, the information typically requested regarding acceptance criteria, study design for AI models, and clinician performance with/without AI assistance cannot be extracted from this document, as it is not applicable.

    The document discusses the substantial equivalence of the Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KV1-5K, to predicate devices. The acceptance criteria in this context refer to engineering and clinical performance standards for blood pressure monitors as defined by various international standards (e.g., AAMI/ANSI/IEC 80601-2-30 for non-invasive sphygmomanometers), rather than performance metrics for an AI algorithm.

    Below, I'll structure the answer to address the prompt's points based on the information available in the document for a traditional medical device, noting where AI-specific information is not present.


    Device: Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KV1-5K

    This device is a traditional non-invasive blood pressure monitor that uses an oscillometric technique. It does not employ artificial intelligence or machine learning for its core blood pressure measurement function. Therefore, many of the questions asked, which are relevant to AI/ML powered devices, are not applicable here.

    1. A table of acceptance criteria and the reported device performance

    The document states that the device was tested to meet relevant national and international standards for electrical, mechanical, environmental performance, and basic safety and essential performance of automated non-invasive sphygmomanometers. While specific numerical acceptance criteria (e.g., mean difference and standard deviation per ISO 81060-2) are not provided in this summary, the document asserts that the device met all relevant requirements.

    Table: General Acceptance Criteria and Stated Performance (Based on Document)

    Criteria CategoryApplicable Standards / RequirementsReported Device Performance
    Basic Safety & Essential PerformanceIEC 60601-1: General requirements for basic safety and essential performance (AAMI / ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R) 2012 And A2:2010/(R) 2012)"Met all relevant requirements of the aforementioned tests."
    Electromagnetic Compatibility (EMC)IEC 60601-1-2:2014: Collateral standard: Electromagnetic Disturbances - Requirements And Tests."Tested and confirmed according to IEC 60601-1-2 EMC Test Report."
    Risk ManagementISO 14971: 2007: Medical devices Application of risk management of medical devices.Implicitly met, as part of device development and testing.
    BiocompatibilityAAMI/ANSI/ISO 10993-1:2009/(R)2013: Biological evaluation of medical devices - Part 1: Evaluation And Testing Within A Risk Management Process. AAMI/ANSI/ISO 10993-5:2009/(R)2014: Tests for In Vitro Cytotoxicity. AAMI / ANSI / ISO 10993-10:2010/(R)2014: Tests for Irritation and Skin Sensitization.Implicitly met, as part of device development and testing.
    Specific Performance for Automated NIBPAAMI/ANSI/IEC 80601-2-30: Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. 2018. The document mentions clinical validation based on ANSI/AAMI/ISO81060-2 standard for predicate devices."Subject device and the predicate device both meet the relevant performance specification requirements and have equivalent essential performance" (despite sensor difference). Clinical validation of predicate devices (K172498, K153450) applicable, meaning the device's accuracy is expected to meet the requirements of ISO 81060-2 (though specific performance values are not given in this summary).
    Home Healthcare EnvironmentIEC 60601-1-11:2015: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment."Tested met all relevant requirements of the aforementioned tests."
    Software ValidationFDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Moderate level of concern)."Software validation was conducted in accordance with a moderate level of concern designation."

    2. Sample size used for the test set and the data provenance

    The document states that repeated clinical testing for the subject device (BP3KV1-5K) was not necessary. Instead, the clinical validation of the predicate devices (BP3MW1-4Y and BP3MS1-4A), performed according to ANSI/AAMI/ISO81060-2, was deemed applicable.

    Therefore, the specific sample size for a test set for the subject device's core BP measurement accuracy is not mentioned as a new study was not performed. The data provenance would be from the original clinical validation studies performed for the predicate devices. The document does not specify country of origin or whether these prior studies were retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this device, as it is not an AI/ML device requiring expert consensus for ground truth or interpretation. The "ground truth" for blood pressure measurement devices is typically a reference measurement method (e.g., auscultatory method by trained observers) used during clinical validation studies (like those following ISO 81060-2). The document does not provide details on the number or qualifications of experts involved in the predicate device's clinical validation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this refers to adjudication for expert labeling/ground truth in AI/ML model validation, which is not relevant for this traditional blood pressure monitor.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-powered diagnostic tool, so an MRMC study comparing human performance with and without AI assistance was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of an AI algorithm intrinsically. For this traditional blood pressure monitor, the "algorithm" is the oscillometric measurement technique. Its performance is evaluated through clinical validation against a reference method (human observers using auscultation), which is essentially a standalone performance evaluation of the device's measurement accuracy. The document states that the clinical validation for the predicate devices is applicable and meets the ANSI/AAMI/ISO81060-2 standard.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For traditional blood pressure monitors, the ground truth in clinical validation studies is typically established by simultaneous measurements using a recognized reference method, such as the auscultatory method performed by trained human observers using a mercury sphygmomanometer or validated electronic reference device. This is the standard approach for studies structured according to ISO 81060-2.

    8. The sample size for the training set

    Not applicable. This device does not use an AI/ML algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no AI/ML training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1