    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medline Foot Plates and Screws are intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of bones of the feet and toes. Specific examples include:

    • · Arthrodesis of the first metatarsalcuneiform joint (Lapidus Fusion)
    • · Arthrodesis of the first metatarsophalangeal joint (MTP) including:
    • Primary MTP Fusion due to hallux ridgidus and/or hallux valgus
    • Revision MTP Fusion
    • · Revision of failed first MTP Arthroplasty implant

    Flatfoot Osteotomies:

    • · Lateral Column Lengthening (Evans Osteotomy)
    • · Plantar Flexion Opening Wedge Osteotomy of the Medial Cuneiform (Cotton Osteotomy)
    • · Medial Displacement Calcaneal Osteotomy (MDCO)

    Midfoot / Hindfoot Fusions:

    • · LisFranc Arthrodesis and/or Stabilization
    • · 1st(Lapidus), 2nd, 3rd, 4th, and 5th Tarsometatarsal (TMT) Fusions
    • · Intercuneiform Fusions
    • · Navicular-Cuneiform (NC) Fusion
    • · Talo-Navicular (TN) Fusion
    • · Calcaneo-Cuboid (CC) Fusion
    • · Medial Column Fusions (NC and 1st TMT)

    The Medline Locking and Non-Locking Cortical Screws are indicated for use with the Medline Foot Plates of the same base material. The Non-Locking Cortical Screws are also indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.

    Device Description

    The Medline Foot Plates and Screws are manufactured from Titanium Alloy. The System includes plates offered in various styles, sizes and options; each contoured for specific anatomy and designed for specific procedures, and 2.7mm and 3.5mm diameter locking and non-locking cortical screws to be used with the polyaxial locking holes and compression slots included in the plates. The non-locking cortical screws can also be used for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture fixation, appropriate for the size of the device. The system also includes reusable instrumentation necessary to implant the plates and screws, e.g. drill guides, tissue protectors, and drill bits. The Medline Foot Plates and Screws are within the currently marketed sizes and indications of the identified predicate devices.

    AI/ML Overview

    This appears to be a 510(k) summary for a medical device (Medline Foot Plates and Screws), not a study demonstrating AI algorithm performance. Therefore, many of the requested categories related to AI performance, ground truth, and expert evaluation are not applicable.

    However, I can extract the information relevant to the device's mechanical performance testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance Criteria (Standard)Reported Device Performance
    Static 4-Point Bending (Plates)ASTM F382 (ensuring equivalence to predicate)Equivalent to Predicate
    Static Pullout (Screws)ASTM F543 (ensuring equivalence to predicate)Equivalent to Predicate
    Static Torque (Screws)ASTM F543 (ensuring equivalence to predicate)Equivalent to Predicate

    2. Sample Size Used for the Test Set and Data Provenance
    The document does not specify the exact sample sizes (number of plates or screws) used for each mechanical test. The data provenance is from in vitro laboratory testing of the physical devices, not from clinical or human data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
    This is not applicable as the "ground truth" here refers to the physical properties and performance of the device measured against established engineering standards (ASTM F382 and F543), not biological or clinical diagnoses requiring expert consensus.

    4. Adjudication Method for the Test Set
    Not applicable. Mechanical testing results are objective measurements from laboratory equipment, not subjective assessments requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
    Not applicable. This is not an AI-assisted diagnostic device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
    Not applicable. This is not an AI algorithm.

    7. Type of Ground Truth Used
    The “ground truth” for the performance assessment of these medical devices is defined by recognized international standards for metallic bone fixation implants: ASTM F382 for plates and ASTM F543 for screws. The performance of the Medline devices was compared to the performance of a predicate device (Wright Medical Ortholoc™ System) tested against the same standards.

    8. Sample Size for the Training Set
    Not applicable. There is no training set as this is not an AI algorithm.

    9. How the Ground Truth for the Training Set Was Established
    Not applicable. There is no training set.