    K Number
    K172587

    Validate with FDA (Live)

    Date Cleared
    2017-09-18

    (21 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mecta-C intervertebral body fusion device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients. The device systems are designed for use with autogenous bone graft to facilitate fusion. One device may be used per intervertebral space. The implants are intended to be used with supplemental spinal fixation.

    The Mecta-C device is intended for use at one level in the cervical spine, from C2-T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies). The cervical device is to be used in patients who have had six weeks of non-operative treatment prior to treatment with the device.

    Device Description

    The purpose of this submission is to add the Mecta-C with titanium markers and the Mecta-C TiPEEK with titanium markers to Medacta's Mecta-C product offering. The Mecta-C with titanium markers and the Mecta-C TiPEEK with titanium markers are intervertebral body fusion devices intended for stabilization and to promote bone fusion during the normal healing process following surgical correction of disorders of the cervical spine. The Mecta-C and the Mecta-C TiPEEK intervertebral body fusion devices are indicated for the treatment of degenerative diseases of the cervical disc and can be used for cervical fusion from C2-T1.

    The Mecta-C with titanium markers are intervertebral body fusion devices manufactured with an implant grade polyetheretherketone (PEEK) body and titanium markers. The Mecta-C TiPEEK with titanium markers are intervertebral body fusion devices manufactured with a PEEK body, commercially pure titanium (CPTi) coating, and titanium markers' material is changing from tantalum to titanium to reduce artefact during imaging scans.

    The titanium marker's spikes are also being modified to reduce the conical edge from 90° to 50°. The purpose of the conical edge change is to create a sharper tip design which allows for better penetration of the titanium pin and a more uniform contact between the PEEK cage and vertebral body.

    The markers are placed in the implant on each end of the TiPEEK cages to allow easier radiological assessment of the position and orientation of the radiolucent TiPEEK cages.

    The cages are offered in various widths, heights, footprint geometries, and lordosis which can be inserted between two cervical vertebra bodies to give support and correction during cervical interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft.

    AI/ML Overview

    This document describes the 510(k) submission for the Mecta-C with titanium markers and Mecta-C TiPEEK with titanium markers intervertebral body fusion devices. This submission aims to extend the existing Mecta-C product offering by changing the marker material from tantalum to titanium and modifying the conical edge of the marker spikes.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes the performance testing of the predicate devices and states that engineering rationales determined that the proposed implants (with titanium markers) did not represent a new worst-case for mechanical testing. This implies that the acceptance criteria for the new devices are met if their performance is equivalent to or better than the predicate devices, especially considering the minor changes. Explicit quantitative acceptance criteria are not detailed for the new device variant within this summary; instead, reliance is placed on the existing predicate device testing.

    Acceptance Criteria CategoryReported Device Performance (Predicate & Subject Devices)
    Mechanical PerformanceDemonstrated through existing tests on predicate devices (K112862, K142744), including: shear testing, tension testing, shear and bonding fatigue, subsidence testing, mechanical testing. Engineering rationales state the new implants do not represent a new worst-case.
    Material PropertiesAdherence to ASTM F2026-16 (PEEK), ISO 5832-3:1996 (Titanium), and ASTM F1580-12 (CPTi coating). Materials have a long history of use in implantable devices.
    BiocompatibilityDeemed unnecessary for additional testing due to extensive history of use of materials and identical manufacturing processes as predicate devices.
    Coating CharacteristicsDemonstrated through coating characterization and abrasion resistance tests on predicate devices.
    Imaging Properties"Implant imaging properties artefacts reduction" was performed on the predicate devices. The change from tantalum to titanium markers is intended to further reduce imaging artifacts.
    PyrogenicityTested using Bacterial Endotoxin Test (LAL) and Pyrogen Test (USP <151>) on predicate devices. Medacta does not intend to label the devices as non-pyrogenic.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state specific sample sizes for the performance tests conducted on the predicate devices. It refers to "verification activities, as identified through risk analysis, were previously conducted on the worst-case implants to written protocols with pre-defined acceptance criteria."

    The data provenance is retrospective, as the performance data cited was generated for the predicate devices (K112862 and K142744) in previous submissions. The country of origin for the data is not specified, but Medacta International SA is based in Switzerland.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable to this submission. The device is a physical intervertebral body fusion device, not an AI/software device requiring ground truth established by medical experts for diagnostic or prognostic purposes. The "ground truth" here relies on established engineering and material science standards and mechanical testing results.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reasons as point 3.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable as this is not an AI/software device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable as this is not an AI/software device.

    7. The Type of Ground Truth Used:

    For this type of medical device, the "ground truth" is established through:

    • Engineering Standards: Adherence to recognized national and international standards (e.g., ASTM F2026-16, ISO 5832-3:1996, ASTM F1580-12) for materials.
    • Mechanical Testing Results: Quantitative measurements obtained from shear, tension, fatigue, and subsidence tests showing mechanical integrity and performance under simulated physiological conditions.
    • Biocompatibility Standards: Compliance with established biocompatibility requirements for implantable devices.

    8. The Sample Size for the Training Set:

    This information is not applicable as this is not an AI/software device and thus does not have a "training set."

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable as this is not an AI/software device.

    In summary: This 510(k) submission relies on demonstrating substantial equivalence to its predicate devices by providing an engineering rationale that minor changes (marker material and conical edge) do not negatively impact device performance, combined with existing performance data from the predicate devices. The "study" proving the device meets acceptance criteria refers to the previous performance tests conducted on the predicate devices, the adherence to material standards, and the engineering justification for the minor modifications.