510(k) Data Aggregation
(258 days)
The general-purpose magnetic resonance imaging (MRI) device is designed to scan any targeted area of the body, to collect, display and analyse MR images and other real-time imaging procedures.
The indications for use are the following:
Imaging portions of calf, knee, ankle, foot, thigh, hand, wrist, forearm, elbow, arm, shoulder, hip, lumbar column, sacral column, cervical column, thoracic spine, pelvis, temporomandibular joint (included only for "Open" configuration), head (included only for "Open" configuration) and upper abdomen (included only for "Open" configuration under option).
Magnifico device is a Magnetic Resonance (MR) system that produces cross-section images of the internal structures of the body. Images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and chemical shift.
When interpreted by a medical expert trained in use of MR equipment, the images can provide diagnostically useful information.
The updated Magnifico is the evolution of its predicate device, the Magnifico device cleared via K241133. Compared to the predicate device, the following modifications have been introduced:
- A new software version (MRI EVOlution 25), including:
- The optional functionality "HyperClarity", an AI-based algorithm for MRI image enhancement (K230854), integrated without any modification. The algorithm is classified under a different regulation in its 510(K) and this is out-of-scope from the current submission of the subject device.
- Abdominal Imaging package for abdomen examination, which delivers:
- Patient's breathing management (including respiratory movement reduction techniques)
- ROTAR sequences
- 3D FSE sequences
- Introduction of a patient breathing detection device (Respiratory Gating System)
- Introduction of a support for spontaneous movements of the knee (flexion/extension)
- A modification of the maximum gradient intensity
- New receiving Coils: DPA Shoulder Coil, Flex Coil, 4 ch. Foot/Ankle coil
Magnifico is substantially equivalent to the predicate device with respect to the fundamental scientific technology, technological characteristics and the principle of operation. The core components remain unchanged and are identical to those of the cleared Magnifico device.
The provided FDA 510(k) clearance letter and summary for the Magnifico Open MRI System (K251901) describes an updated version of a previously cleared device. While it mentions the integration of an AI-based algorithm called "HyperClarity" for MRI image enhancement, the document explicitly states that this algorithm is "out-of-scope from the current submission of the subject device" as it was cleared under a separate 510(k) (K230854).
Therefore, the acceptance criteria and study details for the Magnifico Open MRI System itself primarily focus on demonstrating substantial equivalence to its predicate device (K241133) and the safety and effectiveness of its new features (abdominal imaging, breathing management, gradient intensity modification, new coils, etc.). The document indicates a "validation based on radiologists' expert opinions" for HyperClarity, but the specifics of that study are not included in this submission.
Based on the provided document, here's a breakdown of the requested information for the Magnifico Open MRI System (K251901) as a whole, with the caveat that the detailed study for HyperClarity is referenced but not fully described here:
Acceptance Criteria and Device Performance (Magnifico Open MRI System K251901)
The document does not explicitly present a table of quantitative acceptance criteria for the entire Magnifico Open MRI system in the same way one might find for a standalone algorithm. Instead, the acceptance is implied by meeting safety and performance standards and demonstrating diagnostic image quality.
| Acceptance Criteria Category | Device Performance (as reported) |
|---|---|
| Biocompatibility | Complied with applicable medical device safety standards. |
| Cleaning and Disinfection Effectiveness | Complied with applicable medical device safety standards. |
| Thermal, Electrical, Electromagnetic, Mechanical Safety | Complied with applicable medical device safety standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, IEC 62304, IEC 60601-1-6, IEC 62366-1, ISO 10993-1, ISO 14971, NEMA MS-1, NEMA MS-3, NEMA MS-14). |
| Diagnostic Image Quality | Sample clinical images acquired using the proposed updated device were reviewed by a U.S. board-certified radiologist and demonstrated acceptable diagnostic image quality in accordance with FDA Guidance "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices". |
| Respiratory Motion Artifact Reduction | A comparative evaluation of conventional images and images acquired using the Respiratory Gating System demonstrated the clinical benefit of the patient breathing detection device in reducing respiratory motion artifacts. |
| HyperClarity Algorithm Integration (Note: Out-of-scope for this 510(k)) | Performance testing and a validation based on radiologists' expert opinions were conducted to demonstrate that HyperClarity performs as intended with the Magnifico device and the diagnostic quality of the images is maintained at the 0.4 T field strength. (Details of this specific study are not provided in the K251901 submission, as HyperClarity was cleared under K230854). |
| Overall Safety and Effectiveness | No new questions of safety and effectiveness were raised during non-clinical testing. All necessary performance tests were performed and results show the updated Magnifico meets its intended use and does not raise new questions of safety or effectiveness compared to the predicate device. The device is considered as safe, as effective, and with performance that is substantially equivalent to the predicate device. |
Study Information (for Magnifico Open MRI System K251901)
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated as a numerical count of patients or images. The document mentions "Sample clinical images" and "A comparative evaluation."
- Data Provenance: Not explicitly stated in terms of country of origin. The document refers to "clinical images acquired using the proposed updated device." It does not specify if the studies were retrospective or prospective, though the nature of acquiring "sample clinical images" suggests either freshly acquired images or a selection from an existing dataset.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: "a U.S. board-certified radiologist" (singular, for image quality review).
- Qualifications of Experts: "U.S. board-certified radiologist." Specific years of experience are not mentioned.
- For the HyperClarity feature, it mentions "radiologists' expert opinions" (plural), but further details are not provided within this document.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not specified. For the general image quality review, a single radiologist is mentioned, suggesting "none" in the sense of consensus or arbitration.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: Not explicitly mentioned for the Magnifico Open MRI System itself. The document refers to "A comparative evaluation of conventional images and images acquired using the Respiratory Gating System" to demonstrate clinical benefit, which likely involved comparison but not necessarily a formal MRMC study with AI assistance.
- For the HyperClarity AI feature, the document mentions "validation based on radiologists' expert opinions," which could imply a comparative aspect (with and without the enhancement), but the details (including effect size or MRMC design) are not provided in this specific submission. Given HyperClarity is out-of-scope for this 510(k), that information would be in K230854.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The Magnifico Open MRI System is a hardware device (MRI scanner) with integrated software and features, not a standalone algorithm. Its performance is intrinsically tied to producing images for human interpretation.
- For the HyperClarity AI feature, its performance as an image enhancement algorithm would be considered in a standalone context (algorithm output vs. original image), but the details are not part of this 510(k).
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Expert Opinion/Clinical Review: For diagnostic image quality, the ground truth was established by the "U.S. board-certified radiologist's" review of the images.
- For respiratory motion artifact reduction, the clinical benefit was demonstrated through "comparative evaluation," implying expert assessment of artifact presence.
- For the HyperClarity feature, it refers to "radiologists' expert opinions."
-
The sample size for the training set:
- The document describes the Magnifico Open MRI System (K251901), an MRI scanner, not a machine learning algorithm that undergoes a "training set" in the conventional sense. The "training set" concept would apply to the HyperClarity AI algorithm (K230854), but that information is not part of this submission.
-
How the ground truth for the training set was established:
- This question is not applicable to the Magnifico Open MRI System itself, as it's a hardware device. For the HyperClarity AI algorithm, this information would be detailed in its own 510(k) submission (K230854).
Ask a specific question about this device
(63 days)
The general-purpose magnetic resonance imaging (MRI) device is designed to scan any targeted area of the body, to collect, display and analyze MR images and other real-time imaging procedures.
Imaging portions of calf, knee, ankle, foot, thigh, hand, wrist, forearm, elbow, arm, shoulder, hip, lumbar column, sacral column, cervical column, thoracic spine, pelvis, temporomandibular joint (included only for "Open" configuration), head (included only for "Open" configuration).
MRI provides better soft tissue contrast than CT and can differentiate better between fat, water, muscle, and other soft tissue than CT (CT is usually better at images provide information to physicians and can be useful in diagnosing a wide variety of diseases and conditions.
The new Magnifico EVO 23 employs the same fundamental scientific technology as its predicate device, the first version of Esaote Magnifico, cleared via K212419.
The modifications, do not affect the intended use or alter the fundamental scientific technology of the device.
The software, used on the proposed Magnifico system has been modified, from EVO 21 to include the
• LSDWI sequence for brain examination
-
· MR Angiography
· Maximum Intensity Projection (MIP) -
and to support:
· Additional multi-channel technology coil for body -
· Enhancement of PC board with new processor
Maqnifico keeps all the other technological characteristics of the first cleared, version, as the 0.4T permanent Magnet, based on NdFeB (neodymium), C-shape.
The provided text indicates that the device in question is the Magnifico Open (100009900) and Magnifico MSK (100009910) MRI systems, specifically focusing on the new EVO 23 software release.
Here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with numerical targets. Instead, it describes general categories of testing and qualitative conclusions. The reported device performance is described as meeting these criteria.
| Acceptance Criteria Category | Reported Device Performance (Conclusion from Study) |
|---|---|
| Software Verification and Validation | Activities were provided. |
| Bench Testing Outputs for New Features (LSDW, MRA, MIP) | Outputs were provided and deemed acceptable. |
| Image Quality Test | Tests were conducted and deemed acceptable. |
| Relevant Electromagnetic Compatibility Tests | Tests were conducted and deemed acceptable. |
| Surface Heating Testing | Tests were conducted and deemed acceptable. |
| Relevant Performance Tests for the Body Coil | Tests were conducted and deemed acceptable. |
| Diagnostic Quality of Sample Clinical Images (for new features) | Reviewed by an ACR registered radiologist and determined to be of good diagnostic quality. |
| Substantial Equivalence Determination | The Magnifico MRI system with EVO 23 Software release is substantially equivalent to the predicate device and conforms to applicable medical device safety and performance requirements. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a numerical sample size for the test set. It mentions "Sample clinical images for the new features were reviewed."
The data provenance (country of origin, retrospective/prospective) is not specified in the provided text.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts: One expert.
- Qualifications: An "ACR registered radiologist." The document does not specify their years of experience.
4. Adjudication Method for the Test Set
The adjudication method is not explicitly stated as a formal process like 2+1 or 3+1. Given that only one expert (an ACR registered radiologist) reviewed the sample clinical images for new features, it appears to be a single-reader review rather than a consensus or adjudicated process.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study comparing human readers with and without AI assistance was not mentioned in the provided text. The study focuses on the substantial equivalence of the device itself and the diagnostic quality of its output.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
The device is an MRI system, not an AI algorithm performing diagnostic tasks independently. The "software" modifications refer to new sequences (LSDWI, MR Angiography) and features (MIP), which are integrated into the MRI system to produce images, and these images are then assessed by a human expert. Therefore, a standalone algorithm-only performance study in the context of diagnostic AI is not applicable here. The software verification and validation, along with image quality tests, implicitly evaluate the system's standalone performance in generating medically relevant images.
7. The Type of Ground Truth Used
The ground truth for evaluating the "diagnostic quality" of the sample clinical images for new features was expert consensus/opinion. Specifically, it was the opinion of a single "ACR registered radiologist."
8. The Sample Size for the Training Set
The document does not provide any information about a training set since the device is an MRI system with new imaging sequences and features, not a machine learning model that requires a distinct training phase. The software changes involve implementation of established imaging techniques and display methods.
9. How the Ground Truth for the Training Set was Established
As no training set is mentioned or applicable in the context of the described device and its modifications, this information is not provided.