510(k) Data Aggregation

    K Number
    K062631

    Validate with FDA (Live)

    Device Name
    MYOMO E100
    Manufacturer
    Date Cleared
    2007-04-12

    (219 days)

    Product Code
    Regulation Number
    890.1375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Myomo e100 is indicated for use by stroke patients undergoing rehabilitation to facilitate the following:
    . Stroke rehabilitation by muscle re-education
    . Maintaining or increasing range of motion .

    Device Description

    The Myomo e100 is an electromechanically powered device intended to facilitate movement and increase range of motion for stroke patients. This device would be included as part of prescribed physical therapy to enable stroke patients to exercise that would otherwise be unable to independently do so. The Myomo e100 is an automatic strength amplifier using the detected electromyograph (EMG) as an indication of residual muscle strength. In the case of a patient who is unable to move his/her arm, the addition of strength amplification allows him/her to perform standard physical therapy exercises.
    The Myomo e100 is available in left and right arm configurations. Foam pad inserts of various shapes are provided for fitting the brace for individual patients. Placement of the EMG sensors is configured for the individual patient allowing placement on either the tricep or bicep depending on the prescribed exercise. Optional adjustments for bicep and tricep gain are set and stored as part of the customization.
    As pictured in Figure 1, adjustable straps are used to attach the brace to the arm. Once the brace is properly fitted to the arm the Myomo e100 is turned on using the "Control User Interface". The EMG signals are calibrated in the resting position and then the brace is ready for operation. The battery pack attaches to the "Control User Interface" and is not in contact with the patient.
    The prescribing health care provider incorporates use of the Myomo e100 as part of a patient's physical therapy regimen. Progress can be monitored through the prescribed therapy using a goniometer, Fugl-Meyer testing, or other conventional evaluation techniques.
    The Myomo e100 consists of three main elements:

    • Brace: provides support for the forearm during lift. 4.1
    • 4.2 EMG Sensors: B & L Electrodes
    • Control User Interface: provides the signal processing function and the motor 4.3 mechanism and controls for operation of the Myomo e100.
    AI/ML Overview

    The Myomo e100 510(k) submission (K062631) does not contain the type of acceptance criteria and performance study details typically associated with AI/ML-based medical devices or diagnostics. Instead, this submission focuses on demonstrating substantial equivalence to predicate devices based on indications for use and technical characteristics, and compliance with general safety and performance standards.

    The document explicitly states: "Neither bench, animal nor clinical testing were assessed." and "Performance Testing: Design verification and design validation, e.g., bench testing was performed according to FDA's Design Control Requirements, Title 21 Code of Federal Regulations, Part 820.30." This means a formal clinical performance study, as would be expected for AI/ML devices, was not required or submitted for this particular device.

    Therefore, many of the requested categories in your prompt cannot be answered from the provided text. I will provide the information that is available.


    Description of Acceptance Criteria and Reported Device Performance

    The submission for the Myomo e100 does not present a table of specific acceptance criteria for a diagnostic or AI/ML performance study, nor does it report device performance in terms of metrics like sensitivity, specificity, or accuracy.

    Instead, the "Performance Data" section primarily addresses:

    1. Compliance with recognized standards:

      • IEC 60601-1-1 (Thermal, Electrical, Mechanical Safety)
      • IEC 60601-1-2 (Electromagnetic Compatibility)
      • FDA's "Reviewer Guidance for the Content of Premarket Submission for Software Contained in Medical Devices", May 11, 2005
      • FDA's "Guidance for Off-The-Shelf Software Use in Medical Devices"
      • ISO 14971:2000 (Risk Analysis)
      • ISO 10993-1:2003 (Biocompatibility)
      • Title 21 CFR Part 898 Performance Standard for Electrode Lead (demonstrated by conformity to IEC 60601-1-1).
    2. Design Verification and Validation: Bench testing was performed according to FDA's Design Control Requirements (Title 21 Code of Federal Regulations, Part 820.30).

    The conclusion states that the Myomo e100 is "substantially equivalent to the afore-mentioned predicate devices with respect to indications for use/intended use, and technical characteristics." This is the primary "acceptance criterion" for a 510(k) premarket notification when formal clinical performance data is not deemed necessary by the FDA.


    Information Not Available (and why)

    The following information cannot be extracted from the provided text because the submission did not include a clinical performance study as typically seen for AI/ML devices:

    1. A table of acceptance criteria and the reported device performance: Not applicable in the context of this 510(k) submission.
    2. Sample size used for the test set and the data provenance: No test set (clinical data) was used or reported in this submission for device performance evaluation.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no test set with expert-established ground truth was reported.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI/ML diagnostic device and no human reader study was conducted.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable, as this is not an AI/ML device.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
    8. The sample size for the training set: Not applicable, as no AI model training was mentioned.
    9. How the ground truth for the training set was established: Not applicable.

    Summary of what is available from the text:

    • Device: Myomo e100, an electromechanically powered device for stroke patients undergoing rehabilitation to facilitate muscle re-education and maintain/increase range of motion. It uses detected electromyograph (EMG) as an indication of residual muscle strength to amplify movement.
    • Regulatory Pathway: 510(k) Premarket Notification (K062631).
    • Basis for Clearance: Substantial equivalence to predicate devices (Kinetec 6080 CPM for Elbow, Biomove 3000, Boston Digital Arm System) based on indications for use/intended use and technical characteristics.
    • Performance Data: Focuses on compliance with recognized safety, electrical, EMC, software, risk management, and biocompatibility standards (e.g., IEC 60601-1-1, IEC 60601-1-2, ISO 14971, ISO 10993). Bench testing for design verification and validation was performed as per FDA Design Control Requirements (21 CFR Part 820.30).
    • Absence of Clinical Studies: The submission explicitly states that "Neither bench, animal nor clinical testing were assessed" for the comparison of the Myomo e100 to predicate devices. This indicates that a separate, formal clinical performance study was not required by the FDA for this particular 510(k) clearance, as the device was deemed substantially equivalent through other means (design documentation, technical characteristics, and standards compliance).