Found 2 results

510(k) Data Aggregation

    K Number
    K170781

    Validate with FDA (Live)

    Device Name
    MX50N(MX50YQS)
    Manufacturer
    Date Cleared
    2017-04-04

    (20 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MX50N(MX50YQS) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. It is specifically designed for digital mammography applications and digital breast tomosynthesis applications.

    Device Description

    MX50N(MX50YQS) is a flat panel hi-resolution LCD monitor system for displaying digital medical images. The system consists of a state-of-the-art LCD monitor and a high-resolution graphic control board that connects to a PACS workstation for grayscale image display. The WIDE controller board is installed into the PACS workstation computer or other computer system to display PACS medical images.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the MX50N(MX50YQS) LCD Monitor System, which is intended for displaying and viewing digital medical images, specifically for digital mammography and digital breast tomosynthesis applications. The submission aims to demonstrate substantial equivalence to a predicate device (K160326).

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the comparison to the predicate device and the compliance with international standards. The performance metrics are a direct comparison between the subject device and the predicate device.

    ParameterAcceptance Criteria (Predicate Device Performance)Reported Device Performance (MX50N(MX50YQS))
    Intended UseSame as predicateSame as predicate
    Technological CharacteristicsSame as predicateSame as predicate
    Response Time (Typical)25ms (On/Off)25ms (On/Off)
    LCD Panel Size21.3"21.3"
    Resolution2560 x 20482560 x 2048
    Pixel Pitch0.165mm x 0.165mm0.165mm x 0.165mm
    Brightness1200cd/m²1200cd/m²
    Contrast Ratio1200 : 11000 : 1
    Input SignalDVI-IDVI-I, DisplayPort
    Power Supply100~240 VAC, 50/60Hz100~240 VAC, 50/60Hz
    Color/MonochromeMonochromeMonochrome

    Note regarding "Acceptance Criteria": For the parameters where the subject device’s performance is directly aligned with the predicate (e.g., Response Time, LCD Panel Size), the predicate's performance acts as the de facto acceptance criterion for demonstrating substantial equivalence. For parameters with slight differences (like Contrast Ratio), the acceptance is based on the argument that the difference is not critical.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not describe a test set in the traditional sense of a dataset for evaluating an AI algorithm. This device is a display monitor, not an AI diagnostic tool that processes clinical data.

    Therefore, this information is Not Applicable in the context of this device and submission. There is no mention of clinical data or patient samples being used for testing. The testing performed is non-clinical, related to the display's technical specifications and safety.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    As there is no "test set" of clinical data to establish ground truth for this display monitor, this information is Not Applicable.

    4. Adjudication Method for the Test Set

    As there is no "test set" of clinical data, this information is Not Applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was conducted. This device is a display monitor, not an AI algorithm intended to assist human readers directly or to be evaluated for its impact on reader performance.

    Therefore, this information is Not Applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This device is not an algorithm, but a hardware display monitor. Thus, a standalone algorithm performance study is Not Applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Ground truth in the clinical sense (expert consensus, pathology) is Not Applicable, as this is a display monitor and not an AI diagnostic algorithm. The "ground truth" for the device's technical performance is established by the specifications and measurements according to recognized engineering and medical device standards.

    8. The Sample Size for the Training Set

    No training set is mentioned or applicable. This device is a display monitor, not a machine learning model.

    Therefore, this information is Not Applicable.

    9. How the Ground Truth for the Training Set was Established

    As no training set is applicable, this information is Not Applicable.


    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The study that proves the device meets the acceptance criteria is a non-clinical performance evaluation and compliance testing against recognized international standards and a comparison to a legally marketed predicate device.

    • Type of Study: Non-clinical (bench testing, technical specification validation, standard compliance). No clinical studies were considered necessary or performed.
    • Methodology:
      • Substantial Equivalence Comparison: The MX50N(MX50YQS) was directly compared, parameter by parameter, with the predicate device (5MP Monochrome LCD Monitor MS55i2, K160326). This comparison covered intended use, technological characteristics, and various technical specifications (response time, LCD panel size, resolution, pixel pitch, brightness, contrast ratio, input signal, power supply, color/monochrome). The conclusion was that the devices were substantially equivalent, with any differences (e.g., contrast ratio) not adversely affecting the device's use.
      • Compliance with Consensus Standards: The device underwent testing to comply with the following international and FDA-recognized consensus standards:
        • IEC 60601-1: Medical Electrical Equipment -- Part 1: General Requirements For Basic Safety And Essential Performance.
        • IEC 60601-1-2: Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
    • Performance Metrics Evaluated: Tests were conducted for resolution, luminance, contrast, and noise.
    • Conclusion: The results of these tests demonstrated that the MX50N(MX50YQS) met the acceptance criteria specified in the standards and was adequate for its intended use. The device was found to be as safe and effective as the predicate device.

    In essence, the "study" is a comprehensive engineering and electrical safety validation against established industry benchmarks and a direct side-by-side technical comparison with a cleared device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160353

    Validate with FDA (Live)

    Device Name
    MX50N(MX50YQS)
    Manufacturer
    Date Cleared
    2016-04-22

    (74 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MX50N(MX50YQS) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. It is specifically designed for digital mammography applications.

    Device Description

    MX50N(MX50YQS) is a flat panel hi-resolution LCD monitor system for displaying digital medical images. The system consists of a state-of-the-art LCD monitor and a high-resolution graphic control board that connects to a PACS workstation for grayscale image display. The WIDE controller board is installed into the PACS workstation computer or other computer system to display PACS medical images.

    AI/ML Overview

    The document provided is a 510(k) premarket notification for a medical LCD monitor system (MX50N(MX50YQS)). This type of submission is focused on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving the safety and effectiveness of a novel device through extensive clinical trials. Therefore, the information typically found in studies proving a device meets acceptance criteria for an AI/algorithm-based diagnostic device will not be present.

    Based on the provided document, here's an analysis of the acceptance criteria and supporting study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (from predicate/standards)Reported Device Performance (MX50N(MX50YQS))
    Non-Clinical Performance Data (compared to predicate K052312)
    LCD Panel Size21.3" (Predicate)21.3"
    Resolution2560 × 2048 (Predicate)2560 × 2048
    Pixel pitch0.165mm x 0.165mm (Predicate)0.165mm x 0.165mm
    Brightness1000cd/m² (Predicate)1200cd/m²
    Contrast Ratio700 : 1 (Predicate)1000 : 1
    Input SignalDVI-I (Predicate)DVI-I, DisplayPort
    Power Supply100~240 VAC, 50/60Hz (Predicate)100~240 VAC, 50/60Hz
    Color/MonochromeMonochrome (Predicate)Monochrome
    Compliance to Standards
    IEC 60601-1General Requirements For Basic Safety And Essential PerformanceComplies
    IEC 60601-1-2Electromagnetic Compatibility - Requirements And TestsComplies
    Specific Performance Tests
    ResolutionAcceptance criteria specified in standardsMet acceptance criteria
    LuminanceAcceptance criteria specified in standardsMet acceptance criteria
    ContrastAcceptance criteria specified in standardsMet acceptance criteria
    NoiseAcceptance criteria specified in standardsMet acceptance criteria

    2. Sample Size for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable. This submission is for a medical monitor, not an AI/algorithm-based diagnostic device that processes image data for diagnosis directly. The "tests" mentioned are technical performance verifications of the display hardware (resolution, luminance, contrast, noise) against engineering specifications and international standards, not a diagnostic accuracy study on a dataset of patient cases.
    • Data Provenance: Not applicable. The tests are technical assessments of the device itself.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable. Ground truth in the context of diagnostic interpretation by experts is not relevant for technical performance testing of a monitor. The "ground truth" here is the specified technical performance metrics and compliance with industry standards.

    4. Adjudication Method for the Test Set

    • Not applicable. There is no diagnostic interpretation or consensus needed for the technical performance tests of a monitor.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. The device (a medical monitor) is not an AI algorithm that improves human reader performance. It is a display device for medical images. The submission focuses on its technical specifications and compliance with safety standards.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    • Not applicable. The MX50N(MX50YQS) is a display device, not an algorithm. Its "standalone performance" refers to its technical specifications (brightness, resolution, contrast, etc.) and compliance with electrical and safety standards.

    7. Type of Ground Truth Used

    • For the technical performance tests (resolution, luminance, contrast, noise), the "ground truth" or reference is established by specified acceptance criteria in international and consensus standards (e.g., IEC 60601-1, IEC 60601-1-2 and likely other display performance standards such as DICOM Part 14, though not explicitly listed as a test item in the summary).

    8. Sample Size for the Training Set

    • Not applicable. The MX50N(MX50YQS) is a hardware device (LCD monitor system) and not an AI or machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for this device.