Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K130436

    Validate with FDA (Live)

    Date Cleared
    2013-08-20

    (180 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Extraoral, permanent cementation of ceramic structures made of lithium disilicate glass ceramic (LS2) or zirconium oxide on titanium/titanium alloy or zirconium oxide bases.

    Device Description

    Multilink® Hybrid Abutment is a self-curing luting composite for the extraoral, permanent cementation of ceramic structures made of lithium disilicate glass-ceramic (LS2) or zirconium oxide on titanium alloy or zirconium oxide bases (e.g. abutment or adhesive basis) in the fabrication of hybrid abutments or hybrid abutment crowns.

    AI/ML Overview

    This is a 510(k) summary for a dental cement, not an AI/ML powered device. Therefore, many of the requested criteria regarding AI device studies are not applicable. I will address the applicable criteria based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Measured Performance)Reported Device Performance (Multilink Hybrid Abutment Cement)Predicate Device Performance (Multilink Implant - K090704)Conclusion on Equivalence
    Water Absorption (ISO 4049:2000)Tested and results demonstrate substantial equivalence to predicate.Not explicitly stated, but assumed to meet ISO 4049:2000.Substantially Equivalent
    Water Solubility (ISO 4049:2000)Tested and results demonstrate substantial equivalence to predicate.Not explicitly stated, but assumed to meet ISO 4049:2000.Substantially Equivalent
    Radiopacity (ISO 4049:2000)Tested and results demonstrate substantial equivalence to predicate.Not explicitly stated, but assumed to meet ISO 4049:2000.Substantially Equivalent
    Flexural Strength (ISO 4049:2000)Tested and results demonstrate substantial equivalence to predicate.Not explicitly stated, but assumed to meet ISO 4049:2000.Substantially Equivalent
    Modulus of Elasticity (ISO 4049:2000)Tested and results demonstrate substantial equivalence to predicate.Not explicitly stated, but assumed to meet ISO 4049:2000.Substantially Equivalent
    Compressive Strength (ISO 4049:2000)Tested and results demonstrate substantial equivalence to predicate.Not explicitly stated, but assumed to meet ISO 4049:2000.Substantially Equivalent
    Shear Bond Strength (ISO 4049:2000)Tested and results demonstrate substantial equivalence to predicate.Not explicitly stated, but assumed to meet ISO 4049:2000.Substantially Equivalent
    Biocompatibility (ISO 10993)Testing and evaluation carried out according to ISO 10993.Not explicitly stated, but assumed to meet ISO 10993.Substantially Equivalent
    Intended UseExtraoral, permanent cementation of ceramic structures made of lithium disilicate glass ceramic (LS2) or zirconium oxide on titanium alloy or zirconium oxide bases.Intraoral permanent cementation of restorations (implied from product name and comparison).Differentiated by intraoral/extraoral use, but overall deemed substantially equivalent in context of cement properties.
    Chemical CompositionModified from the predicate.Not explicitly stated, but implied to be different from the subject device.Substantially Equivalent (despite modification, performance is equivalent)

    2. Sample size used for the test set and the data provenance

    • Sample size: Not specified.
    • Data provenance: Not specified, but likely from in-house laboratory testing based on the standards cited (ISO 4049:2000, ISO 10993). This would typically be considered prospective testing for the purpose of demonstrating device performance against a standard. The country of origin is not explicitly stated beyond "Ivoclar Vivadent, Inc. 175 Pineview Drive, Amherst, NY 14228".

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a material, not a diagnostic or AI device that requires expert ground truth for its performance evaluation in the context of the reported tests. Its performance is measured against physical and chemical standards.

    4. Adjudication method for the test set

    Not applicable. The tests are objective measurements of physical and chemical properties based on ISO standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML powered device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML powered device.

    7. The type of ground truth used

    For the performance tests (Water Absorption, Solubility, Radiopacity, Flexural Strength, Modulus of Elasticity, Compressive Strength, Shear Bond Strength), the "ground truth" or reference is the standards set forth by ISO 4049:2000. The biocompatibility "ground truth" is defined by ISO 10993. The device's performance is compared against these established material performance criteria and against the performance of its predicate device to claim substantial equivalence.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML powered device. There is no "training set" in the context of material science evaluation for a 510(k) submission like this.

    9. How the ground truth for the training set was established

    Not applicable. As this is not an AI/ML powered device, there is no training set or associated ground truth establishment process in the AI sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123397

    Validate with FDA (Live)

    Date Cleared
    2013-01-31

    (87 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Multilink Automix is used for the permanent cementation of indirect restorations where a strong bond is desired:

    • Inlays, onlays, crowns, bridges and root posts made of: –
      • o Metal and metal ceramics
      • o All-ceramics, in particular opaque zirconium oxide ceramics
      • o Composites and fibre-reinforced composites
    Device Description

    Multilink Automix is a self-curing and self-etching luting composite system for the adhesive luting of indirect restorations made of metalceramic all-ceramic and composite. Multilink Automix is applied together with the selfetching and self-Ocuring Multilink Primer. Monobond Plus is recommended as a coupling agent to achieve a strong bond to precious and non-precious alloys, as well as to all ceramics made of zirconium and aluminum oxide and silicate ceramics.

    AI/ML Overview

    The provided text describes the submission for MULTILINK AUTOMIX to the FDA for 510(k) clearance. The focus of the study is to demonstrate substantial equivalence to a predicate device, not necessarily to set new, absolute acceptance criteria for the new device as a standalone product based on clinical outcomes.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the performance of the predicate device, Multilink (K032470), as the new device, Multilink Automix, aims to show substantial equivalence. The "acceptance criteria" for the Multilink Automix are therefore that its performance falls within an expected range relative to the predicate, demonstrating it is "substantially equivalent" and "no new issues of safety and effectiveness" persist.

    PropertyUOMPredicate Device (Multilink K032470)Multilink Automix (Self Curing) Reported PerformanceMultilink Automix (Dual Curing) Reported PerformanceAcceptance Criteria (Implied: Substantially Equivalent to Predicate)
    Working time (37C)Min3-43-4--Within range of 3-4 min
    Setting timeMin7-96-7--Within a comparable range to 7-9 min
    Film Thicknessµm<20<50<50<50 µm (Note: Higher than predicate, but still acceptable for this class of device)
    Water Absorptionµg/mm³<25.0<40<40<40 µg/mm³ (Note: Higher than predicate, but acceptable)
    Water solubilityµg/mm³<3.0<7.5<7.5<7.5 µg/mm³ (Note: Higher than predicate, but acceptable)
    Radiopacity%Al350350350350 %Al (Matches predicate)
    Flexural StrengthMPa90±1070±20110±10Comparable to 90±10 MPa (Self-curing slightly lower, dual-curing higher, but both within a reasonable range for equivalence)
    Modulus of ElasticityMPa5000±10003250±4006000±400Comparable to 5000±1000 MPa (Ranges overlap sufficiently)
    Compressive StrengthMPa250±20240±20280±20Comparable to 250±20 MPa (Ranges overlap sufficiently)
    TransparencyBase (trans and Cat)Yellow and CatOpaque and Cat%12±1.510±1.52±0.512±1.510±1.52±0.512±1.510±1.52±0.5Matches predicate values
    Vickers HardnessMPa370--440±30Comparable to 370 MPa (Dual-curing is higher, which is generally positive)
    Shear Bond StrengthDentin / 24hEnamel/ 24hMPa21±223+417±518+321±223+4Comparable to 21±2 and 23+4 MPa (Dual-curing matches, self-curing slightly lower but potentially acceptable)

    Note: The acceptance criteria are "implied" because the document focuses on demonstrating substantial equivalence to the predicate device, meaning the new device's performance should be within an acceptable range of the predicate's performance.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample sizes for each specific test (e.g., number of specimens for flexural strength tests). It only indicates that "The device was tested in accordance with ISO 4049:2000" and other standards. These standards typically specify minimum sample sizes for material property tests.

    The data provenance is from Ivoclar Vivadent, Inc., which is listed as being in Amherst, NY, USA. The data is from laboratory testing performed to demonstrate material properties, and thus is considered prospective in the sense that the tests were conducted specifically for this submission, although not on live patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information is not applicable to this submission. The "ground truth" here is objective physical and chemical properties of a dental material, measured using standardized laboratory methods (e.g., ISO 4049:2000). It does not involve expert interpretation or clinical assessment by medical professionals to establish a truth.

    4. Adjudication Method for the Test Set

    This is not applicable. As explained above, the "truth" is established by laboratory measurement, not by expert consensus or adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is not applicable. This submission is for a dental material/device (luting composite system), not an AI-powered diagnostic or imaging device that would involve human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. This is a dental material, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance data presented is derived from standardized laboratory testing and measurements of material properties (e.g., Water Absorption, Flexural Strength, Shear Bond Strength) as per ISO 4049:2000 and other relevant standards.

    8. The Sample Size for the Training Set

    This is not applicable. This submission is for a physical dental material, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090704

    Validate with FDA (Live)

    Date Cleared
    2009-06-17

    (92 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K061781

    Validate with FDA (Live)

    Device Name
    MULTILINK SPRINT
    Date Cleared
    2006-09-05

    (71 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Multilink Sprint is a self-adhesive, dual curing luting composite for the permanent cementation of direct restorations made of:

    • Metal and metal-ceramic (inlays, onlays, crowns, bridges and posts) .
    • High-strength all ceramic: zirconiumoxide-, Lithiumdisilicate and . Aluminum oxide ceramic (crowns, bridges, and posts)
    • Fiber-enforced composites (posts:) .
    Device Description

    Multilink Sprint is a self-adhesive, dual curing luting composite.

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a dental cement called "Multilink Sprint." It does not contain any information about acceptance criteria or a study proving the device meets acceptance criteria.

    The letter primarily:

    • Confirms that the device is substantially equivalent to legally marketed predicate devices.
    • Outlines general regulatory requirements the manufacturer must comply with.
    • Lists the "Indications for Use" for the Multilink Sprint, which specify the types of restorations it can permanently cement.

    Therefore, I cannot provide the requested information based on the provided text. The document is regulatory approval, not a scientific study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032470

    Validate with FDA (Live)

    Device Name
    MULTILINK
    Date Cleared
    2003-11-05

    (86 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Multilink® is a self-curing and self-etching luting composite system for the adhesive whiles is a coll cannons made of metal, metal-ceramic, all-ceramic and composite.

    It is Contraindicated if the patient has allergies to any ingredient; and no direct application on pulpa-near dentin or on the open pulp or irritations of the pulp may occur.

    Device Description

    Multilink® is a self-curing and self-etching luting composite system

    AI/ML Overview

    I am sorry, but the provided text from the FDA 510(k) summary for the Multilink device (K032470) does not contain the detailed information necessary to answer your request about acceptance criteria and a study proving device performance.

    The document is a clearance letter from the FDA to Ivoclar Vivadent, Inc., confirming that their Multilink dental cement is substantially equivalent to legally marketed predicate devices. It primarily focuses on regulatory approval, classification, and general controls provisions of the Federal Food, Drug, and Cosmetic Act.

    The document does not include:

    • A table of acceptance criteria and reported device performance.
    • Details about a study with sample sizes, data provenance, ground truth establishment, expert qualifications, or adjudication methods.
    • Information on Multi-Reader Multi-Case (MRMC) comparative effectiveness studies or standalone algorithm performance.
    • Training set sample sizes or how their ground truth was established.

    Therefore, I cannot provide the requested information based on the input text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1