Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K173454

    Validate with FDA (Live)

    Date Cleared
    2018-08-16

    (283 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    C30 Multi-parameter Patient Monitor (hereinafter called Patient Monitor) is indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. It's used for monitoring, reviewing and storing of multiple physiological parameters as following: ECG, heart rate (HR), respiration (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP) of single patient. All parameters can be applied to adult, pediatric or neonate. The Patient Monitor is not intended to be used on the patients with pacemaker.

    The monitor is used for monitoring the clinical patients, so only the doctors and nurses who are qualified through training can use these monitors.

    Device Description

    C30 Multi-parameter Patient Monitor (hereinafter called Patient Monitor) is intended to be used for monitoring, reviewing and storing of multiple physiological parameters as following: ECG, heart rate (HR), respiration (Resp), temperature (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP) of single patient. Other parameters can be applied to adult, pediatric or neonate.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Shenzhen Comen Medical Instruments Co., Ltd. Multi-parameter Patient Monitor, Model C30. It describes the device's characteristics, intended use, and comparisons to legally marketed predicate devices.

    However, the document focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and the results of a specific clinical study for the device's performance in isolation. The testing mentioned is primarily related to electrical safety, electromagnetic compatibility, and adherence to performance standards, along with some clinical verification for NIBP and respiration rate accuracy.

    Therefore, much of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of an AI algorithm's performance (such as sensitivity, specificity, or reader improvement) is not directly available or applicable to this type of medical device submission. This document describes a traditional multi-parameter patient monitor, not an AI-powered diagnostic device.

    Based on the provided text, here's what can be extracted and what cannot be provided:

    1. A table of acceptance criteria and the reported device performance:

    The document lists various performance characteristics for each parameter (ECG, Respiration, Temperature, SpO2, Pulse Rate, NIBP, CO2) as part of the technological comparison to predicate devices, including measurement ranges and accuracies. These can be considered the de facto performance criteria. While the document states the device was "tested and found to be in compliance with recognized safety, performance and electromagnetic compatibility standards," it does not provide a specific table of acceptance criteria with corresponding measured performance for each parameter from a single comprehensive study report. Instead, it infers compliance through comparison to predicate device specifications and adherence to standards.

    ParameterAcceptance Criteria (from predicate comparison/standards)Reported Device Performance (Implied from substantial equivalence and compliance statements)
    ECG Heart Rate (HR)Neonate/Pediatric: 15 to 350 bpm; Adult: 15 to 300 bpm; Accuracy: ±1% or ±1 bpm (whichever is larger)The document states the device has "similar" characteristics and "complies" with performance standards, implying it meets these ranges and accuracies, but no direct measured performance values are provided.
    Respiration Rate (Resp)Measurement range: Adult: 0 to 120 rpm; Neonate: 0 to 150 rpm; Accuracy: 7 to 150 rpm: ±1 rpm (for subject device, compared to predicate's ±2 rpm or ±2%); 0 to 6 rpm: Not specified.Clinical study indicated to verify RR accuracy by comparing subject device RR to capnography. Implies accuracy within the stated/expected range, but no specific results are quantified in the summary.
    Temperature (Temp)Measurement range: 0 to 50°C (32 to 122°F); Accuracy: ±0.1°C or ±0.2°F (without probe)No specific measured performance values provided, but implied to be compliant with these specifications.
    SpO2Masimo SpO2: Measurement range: 1 to 100%; Accuracy: 70 to 100%: ±2% (adult/pediatric, no motion), ±3% (neonate, no motion, and with motion). Nellcor SpO2: Measurement range: 0 to 100%; Accuracy: 70 to 100%: ±2% (adult/pediatric), ±3% (neonate). 0 to 69%: Not specified.No specific measured performance values provided, but implied to be compliant with these specifications as "Same" as predicate device for Masimo/Nellcor module characteristics.
    Pulse Rate (PR)From Masimo SpO2: Range 25 to 240 bpm; Accuracy: ±3 bpm (no motion), ±5 bpm (with motion). From Nellcor SpO2: Range 20 to 300 bpm (20-250 bpm: ±3 bpm, 251-300 bpm: not specified).No specific measured performance values provided, but implied to be compliant with these specifications as "Same" as predicate device for Masimo/Nellcor module characteristics.
    Non-Invasive Blood Pressure (NIBP)Measurement range (Adult/Pediatric/Neonate systolic/mean/diastolic ranges vary as per table). Accuracy: Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg. (These are performance requirements based on ISO 81060-2:2013)."According to ISO 81060-2:2013, NIBP clinical accuracy was verified." This statement indicates testing was done according to the standard, implying it met the accuracy requirements specified by that standard, but explicit performance values are not provided.
    Carbon Dioxide (CO2)Respironics: Range 0-150 mmHg; Accuracy: 0-40 mmHg: ±2 mmHg, 41-70 mmHg: ±5%, 71-100 mmHg: ±8%, 101-150 mmHg: ±10%. awRR: 0-150 bpm, Accuracy: ±1 rpm. Masimo: Range 0-190 mmHg; Accuracy: 0-114 mmHg: ±1.52 mmHg + 2%. awRR: 0-150 bpm, Accuracy: ±1 rpm.No specific measured performance values provided, but implied to be compliant with these specifications.

    2. Sample size used for the test set and the data provenance:

    • NIBP Clinical Study: The document states "According to ISO 81060-2:2013, NIBP clinical accuracy was verified." This standard specifies requirements for Non-Invasive Sphygmomanometers, including clinical investigation procedures. However, the exact sample size for this specific study and the data provenance (country, retrospective/prospective) are not given in this summary.
    • Respiration Rate Clinical Study: "To verify respiration rate (RR) accuracy, Clinical study was conducted comparing the RR from the subject device with that from capnography." The exact sample size and data provenance are not given.
    • For other parameters, bench testing and compliance with standards are mentioned, but no specific clinical test sets or sample sizes are detailed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided as this is not an AI diagnostic device and does not inherently involve expert image interpretation or similar processes for ground truth establishment. Clinical studies mentioned (NIBP, RR) would likely use reference measurements from other validated medical devices as a "ground truth" or comparator, but details on expert involvement are absent.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable/provided. Adjudication methods are typically used in studies involving subjective assessments, especially common in AI diagnostic imaging where multiple readers reconcile disagreements. This device is a physiological monitor measuring objective parameters.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable/provided. This is a physiological monitoring device, not an AI-powered diagnostic tool intended to assist human readers in interpreting complex medical data or images. Therefore, an MRMC study is beyond the scope of this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This information is not applicable/provided. This device is a multi-parameter patient monitor that measures physiological signals directly rather than an AI algorithm performing a standalone diagnostic function. The device's performance is its direct measurement output, which is then used by healthcare professionals.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For the NIBP and Respiration Rate clinical studies, the "ground truth" would likely be established through reference measurements from other validated medical devices or established clinical methods. For example, NIBP accuracy usually compares the device's readings to those obtained through an invasive arterial line or another highly accurate, validated NIBP device. For respiration rate, capnography (a precise measurement of CO2 in exhaled breath) was explicitly mentioned as the comparator.

    8. The sample size for the training set:

    The device described is a traditional patient monitor developed through engineering and design principles, not a machine learning or AI model that requires a "training set." Therefore, this concept is not applicable to this product.

    9. How the ground truth for the training set was established:

    As above, the concept of a "training set" and "ground truth for the training set" is not applicable to this traditional medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100394

    Validate with FDA (Live)

    Date Cleared
    2010-04-15

    (58 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PMS8210A Multi Parameter Patient Monitor is a multi-parameters monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PMS8210A Patient Monitor has certain features and functions.

    The patient parameters that can be monitored by PMS8210A Patient Monitor are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Arterial Hemoglobin Oxygen Saturation(SpO2) and Temperature (TEMP). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.

    The PMS8210A Patient Monitor is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.

    The PMS8210A Patient Monitor is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.

    Device Description

    PMS8210A Patient Monitor is a battery or line-powered patient monitor. The Patient Monitor acquires the physiological signals such as ECG, respiration (RESP), Non-Invasive blood pressure (NIBP), Saturation of pulse oxygen (SPO2), Temperature (TEMP), Heart rate (HR) and Pulse Rate (PR), The signals are converted into digital data and processed, examines the data for alarm conditions and displays the data. The monitor also provides operating control for the user.

    AI/ML Overview

    The provided text does not contain detailed information about specific studies, acceptance criteria, or performance data for the PMS8210A (IRIS) Multi-parameter Patient Monitor. Instead, it focuses on comparing the device's technical specifications and adherence to various medical standards with a predicate device (PM-9000 EXPRESS PATIENT MONITOR).

    Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor details about sample sizes, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, or training set information.

    The document primarily acts as a 510(k) summary, which is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. This typically involves showing that the new device has the same intended use and the same technological characteristics, or different technological characteristics that do not raise different questions of safety and effectiveness and are as safe and effective as the predicate device.

    The study described is a comparison of the PMS8210A Multi-parameter Patient Monitor's specifications and compliance with relevant medical standards against a predicate device, the PM-9000 EXPRESS PATIENT MONITOR (K053234). The intent is to demonstrate that the subject device is substantially equivalent to the predicate device.

    Here's what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists various technical specifications and accuracy claims for the PMS8210A device (Subject Device) and compares them to the predicate device. While these are performance characteristics, they are not explicitly labeled as "acceptance criteria" through a formal study results table. The implication is that meeting or being comparable to the predicate's performance, and adhering to referenced standards, constitutes meeting acceptance criteria for substantial equivalence.

    ParameterAcceptance Criteria (Implied by Predicate/Standard)PMS8210A Reported Performance
    ECG
    GainPredicate: ×0.125; ×0.25; ×0.5; ×1; ×2; auto.×0.5; ×1.
    Range of Heart Rate MonitoringAdult: 15300 bpm; Neonate/Pediatric: 15350 bpm (Predicate)Adult: 20300 bpm; Neonate/Pediatric: 20350 bpm
    Resolution (HR)1 bpm (Predicate)1 bpm
    Precision (HR)1 bpm or ±1% (Predicate)20200 bpm: 5% or ±5bpm; 201350 bpm: 10%
    CMRRDiagnostic mode: ≥90dB; Monitoring mode: ≥105dB; Surgical mode: ≥105dB (Predicate)≥89 dB
    Frequency Characteristic (Mode 1)0.05Hz-100Hz (Diagnostic), 0.5Hz-40Hz (Monitoring), 1Hz-20Hz (Surgical) (Predicate)Mode 1: 0.1Hz-40Hz; Mode 2: 0.67Hz-40Hz; Mode 3: 1Hz-25Hz
    Respiration
    Measuring RangeAdult: 0120 rpm; Neonate/Pediatric: 0150 rpm (Predicate)15 ~ 120 rpm
    Resolution1 rpm (Predicate)1 rpm
    Precision06 rpm: Unspecified; 7120rpm: ±2 rpm or ±2% (Predicate)The bigger one between ±2 rpm or ±2%
    NIBP
    BP AccuracyArithmetic mean values: ±5 mmHg; Standard deviation no greater than 8 mmHg (Predicate, also stated by ANSI/AAMI SP10:2002; EN1060-4)Arithmetic mean values: ±5 mmHg; Standard deviation no greater than 8 mmHg
    SpO2
    SpO2 Measurement Range0~100% (Predicate)0~100%
    SpO2 Measurement AccuracyAdult/Pediatric: 70100%: ±2%; 069%: Unspecified; Neonate: 70100%: ±3%; 069%: Unspecified (Predicate)Adult/Pediatric: 70100%: ±2%; 069%: Unspecified; Neonate: 70100%: ±3%; 069%: Unspecified
    Pulse Rate Measurement20 to 250 BPM (Predicate)20~250bpm
    Temperature
    Measurement Range0°C50°C (32122°F) (Predicate)0°C50°C (32122°F)
    Measurement Accuracy±0.1°C (±0.2°F) (Predicate, also stated by ASTM E1112:00)±0.1°C (±0.2°F)

    2. Sample size used for the test set and the data provenance:
    This information is not provided in the given text. The document refers to "performance testing in accordance with IEC standards" but does not detail the size or nature of the test set or its provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This information is not provided in the given text. The document focuses on technical specifications and standard compliance, not clinical ground truth establishment by experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    This information is not provided in the given text.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This information is not applicable/provided. The device is a multi-parameter patient monitor, not an AI-assisted diagnostic tool involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
    The device itself is a standalone monitor. Its performance is assessed against technical specifications and standards (e.g., IEC, AAMI, EN, ISO). The document states the device has "software of Moderate concern" and that "the appropriate level of Software evaluation was performed," but does not detail the standalone performance study specific to an algorithm. The overall product performance is standalone.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    The ground truth for evaluating the device's technical specifications and accuracy is largely based on measurements against established reference standards and calibrated equipment, as implied by adherence to standards like ANSI/AAMI SP10 for NIBP accuracy and ASTM E1112 for temperature accuracy. For SpO2, it leverages the Nellcor SpO2 module, implying its proven accuracy. There is no mention of expert consensus, pathology, or outcomes data as ground truth in this document.

    8. The sample size for the training set:
    This information is not provided in the given text. As a patient monitor rather than an AI/ML algorithm requiring a training set in the typical sense, this concept might not directly apply or is not detailed. The "software of Moderate concern" does not imply a classical AI/ML model.

    9. How the ground truth for the training set was established:
    This information is not provided in the given text, and likely not applicable in the context of this traditional medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1