Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K214081

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-04-05

    (99 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MRCAT imaging is intended to provide the operator with information of tissue properties for radiation attenuation estimation purposes in photon external beam radiotherapy treatment planning.

    Indication for use:

    MRCAT Head and Neck is indicated for radiotherapy treatment planning for patients with soft tissue tumors in the Head and Neck region.

    Device Description

    MRCAT Head & Neck is a software application to Ingenia, Ingenia Ambition, and Ingenia Elition MR systems. MRCAT Head & Neck is available to the customer as an option to Ingenia MR-RT package, which is a set of accessories for Ingenia systems.

    Automated generation of MRCAT images takes place at the MR console of Ingenia. The embedded image post-processing runs in the background parallel to image acquisition. MRCAT algorithm enables automatic tissue characterization: Bones are segmented from mDixon in-phase and water images using machine learning based segmentation. Body outline is segmented using in-phase and water images. Tissues are then assigned a continuum of HU values depending on the fat and water intensities of the voxels. The HU assignment provides MRCAT images with CT-like density information.

    AI/ML Overview

    The Philips Oy MRCAT Head & Neck device, a software application for MR systems intended to provide tissue property information for radiation attenuation estimation in photon external beam radiotherapy treatment planning for soft tissue tumors in the Head and Neck region, underwent testing to demonstrate substantial equivalence to its predicate device, MRCAT Brain.

    The primary study focused on validating the accuracy of simulated dose calculations and geometric accuracy when using MRCAT Head & Neck images compared to CT-based plans.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (MRCATHN)Reported Device Performance
    Dose Accuracy: - Simulated dose based on MRCAT Head & Neck images shall not differ in 95% of the indicated patients (gamma analysis criterion 2%/2mm realized in 98% of voxels within the PTV or exceeding 75% of the maximum dose) when compared with CT-based plan. - The average simulated dose based on MRCAT Head & Neck shall not deviate more than 5% or 1 Gy, whichever is greater, in 99% of the indicated patients in the volume of sensitive organs when compared with a CT-based plan.- PTV dose differences obtained when using MRCAT in place of CT were well below 1% with a very small bias, indicating clinical insignificance. - Results for artificial PTVs (automatically placed around the head and neck region) agreed well with clinical plan results, strengthening the conclusion of accurate dose calculations. - The device met the dose accuracy criteria.
    Geometric Accuracy: - ± 1 mm accuracy: 200 mm diameter sphere - ± 2 mm accuracy: 400 mm diameter sphere (limited in the bore direction by +/- 160 mm from the z=0 mm plane) - ± 5 mm accuracy: 500 mm diameter sphere (limited in the bore direction by +/- 160 mm from the z=0 mm plane)- The document explicitly states: "MRCAT accuracy: - ± 1 mm accuracy: 200 mm diameter sphere - ± 2 mm accuracy: 400 mm diameter sphere (limited in the bore direction by +/- 160 mm from the z=0 mm plane) - ± 5 mm accuracy: 500 mm diameter sphere (limited in the bore direction by +/- 160 mm from the z=0 mm plane)" While the exact "reported performance" against these metrics is not detailed in the provided summary, the statement "MRCAT Head and Neck met the acceptance criteria and is adequate for this intended use" implies these geometric accuracy criteria were also met. The visibility of bone structures described as "equivalent for both products" (MRCAT Brain and MRCAT Head & Neck) also supports similar geometric performance.
    Safety and Effectiveness: - Compliance with relevant international and FDA-recognized consensus standards (ANSI/AAMI ES60601-1, IEC 60601-1-6, IEC 60601-2-33, IEC 62304, IEC 62366-1, ISO 14971).- The MRCAT Head and Neck complies with the aforementioned international and FDA-recognized consensus standards. Verified through non-clinical verification and validation tests. The device met the safety and effectiveness criteria.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: The summary mentions "PTV dose differences obtained when using MRCAT in place of CT" and "artificial PTV analysis". It also refers to "95% of the indicated patients" and "99% of the indicated patients," suggesting a study involving patient data or simulated patient cases. However, the exact number of patients or cases in the test set is not explicitly stated in the provided document.
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • The document does not explicitly state the number of experts used to establish ground truth for the test set or their specific qualifications.
    • However, it does mention that "The HU values for the MRCAT Head and Neck are calibrated using registered CT images from several sites." This implies that the CT images, which serve as the reference for HU values, were acquired and potentially interpreted by medical professionals, though their role in "ground truth establishment" for the specific test set is not detailed.

    4. Adjudication Method for the Test Set:

    • The document does not describe an adjudication method (e.g., 2+1, 3+1). The "ground truth" seems to be primarily based on CT images and their associated dose calculations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • A MRMC comparative effectiveness study was not discussed in the provided text. The evaluation focuses on the device's performance in dose and geometric accuracy against a CT-based plan, rather than comparing human reader performance with and without AI assistance.

    6. Standalone (Algorithm Only) Performance:

    • Yes, the study appears to evaluate the standalone performance of the MRCAT Head & Neck algorithm. The dose accuracy and geometric accuracy criteria are stated as direct evaluations of the "simulated dose based on MRCAT Head & Neck images" and "MRCAT accuracy," respectively, implying no human intervention in the generation or initial interpretation of the MRCAT images for these assessments. The device provides "information of tissue properties for radiation attenuation estimation purposes," which is then used in treatment planning.

    7. Type of Ground Truth Used:

    • The ground truth primarily used for comparing dose accuracy and HU values is based on CT-based plans and registered CT images. CT is considered the gold standard for electron density information in radiotherapy planning.

    8. Sample Size for the Training Set:

    • The document states that the "CNN is trained using matched pairs of CT and MRCAT source images" and "The HU values for the MRCAT Head and Neck are calibrated using registered CT images from several sites." However, the exact sample size of the training set (number of images/patients) is not specified.

    9. How the Ground Truth for the Training Set Was Established:

    • The ground truth for the training set was established using "matched pairs of CT and MRCAT source images." This means that for each MR image used for training the CNN, a corresponding CT image was available. These CT images, presumably acquired as part of standard clinical practice, provided the reference Hounsfield Unit (HU) values for tissue characterization and density information that the MRCAT algorithm aims to replicate. The calibration of HU values also used "registered CT images from several sites," reinforcing the reliance on CT as the ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193109

    Validate with FDA (Live)

    Device Name
    MRCAT brain
    Date Cleared
    2020-01-24

    (77 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MRCAT imaging is intended to provide the operator with information of tissue properties for radiation attenuation estimation purposes in photon external beam radiotherapy treatment planning.

    Indications for use:

    MRCAT Brain is indicated for radiotherapy treatment planning for primary and metastatic brain tumor patients.

    Device Description

    MRCAT brain is a software application to Ingenia, Ingenia Ambition, and Ingenia Elition MR systems. MRCAT brain is available to the customer as an option to Ingenia MR-RT package, which is a set of accessories for Ingenia systems.

    Automated generation of MRCAT images takes place at the MR console of Ingenia. The embedded image post-processing runs in the background parallel to image acquisition. MRCAT algorithm enables automatic tissue characterization: Bones are segmented from mDixon in-phase and water images using machine learning based segmentation. Body outline is segmented using in-phase and water images. Tissues are then assigned a continuum of HU values depending on the fat and water intensities of the voxels. The HU assignment provides MRCAT images with CT-like density information.

    AI/ML Overview

    The document provides information on the Philips Medical Systems MR Finland MRCAT Brain device, which is a software add-on for MR systems intended for radiotherapy treatment planning for primary and metastatic brain tumor patients.

    Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (MRC-Brain)Reported Device Performance
    Dose Accuracy: - Simulated dose based on MRCAT Brain images shall not differ in 95% of the indicated patients (gamma analysis criterion 2%/2mm realized in 98% of voxels within the PTV or exceeding 75% of the maximum dose) when compared with CT-based plan. - The average simulated dose based on MRCAT Brain shall not deviate more than 5% or 1 Gy, whichever is greater, in 99% of the indicated patients in the volume of sensitive organs when compared with CT based plan.The robustness of the MRCAT brain algorithm for producing equivalent dose plans to CT using gamma analysis with criterion of 1%/1mm is shown by post-processing MRCAT images from patients, and calculating dose using the MRCAT images. (Though the reported criterion is 1%/1mm, the acceptance criterion specifically mentioned 2%/2mm, implying the device met or exceeded this with the 1%/1mm demonstration). The document states that MRCAT brain met the acceptance criteria and is adequate for its intended use.
    Geometric Accuracy: - MRCAT accuracy: ± 1 mm accuracy for a 200 mm diameter sphere. - MRCAT accuracy: ± 5 mm accuracy for a 500 mm diameter sphere (limited in the bore direction by +/- 160 mm from the z=0 mm plane).The document states "No significant difference" in geometric accuracy when comparing MRCAT Brain to MRCAT Pelvis, which has the same geometric accuracy criteria. It also generally states that "Non Clinical verification and or validation test results demonstrate that the MRCAT brain... Meets the acceptance criteria and is adequate for its intended use."
    Compliance with Standards: - ANSI/AAMI ES60601-1: 2012 - IEC 60601-1-6:2010 - IEC 60601-2-33:2015 - IEC 62304:2016 - IEC 62366-1:2015 - ISO 14971:2007 (A comprehensive list of international and FDA-recognized consensus standards for medical electrical equipment, usability, safety of MR equipment, medical device software lifecycle processes, usability engineering, and risk management)."The MRCAT brain complies with the following international and FDA-recognized consensus standards." and "Non Clinical verification and or validation test results demonstrate that the MRCAT brain: Complies with the aforementioned international and FDA-recognized consensus standards."
    MRCAT image generation correctness: - Sanity checks to ensure imaging field of view is correctly positioned. - Sanity checks to ensure MRCAT body outline matches that of the MR."The generated MRCAT images are checked for correctness to ensure validity of the generated MRCAT for radiation treatment. The sanity checks ensure that the imaging field of view has been positioned correctly and that the MRCAT body outline matches that of the MR." The overall conclusion on non-clinical tests also states: "met the acceptance criteria and is adequate for this intended use."
    HU value calibration: - The HU values for the MRCAT Brain are calibrated using registered CT images from several sites."The HU values for the MRCAT brain are calibrated using registered CT images." and "The overall conclusion on non-clinical tests also states: "met the acceptance criteria and is adequate for this intended use."

    The Study Proving Device Meets Acceptance Criteria:

    The document describes a "Summary of Non-Clinical Performance Data" and a "Summary of Clinical Data" to support the device's substantial equivalence and adherence to acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions that the robustness of the MRCAT brain algorithm for producing equivalent dose plans to CT using gamma analysis was shown by post-processing MRCAT images from patients. However, it does not specify the exact sample size used for this patient data (test set) or the country of origin/provenance (retrospective or prospective) of this patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    The document does not provide information regarding the number of experts used to establish ground truth or their specific qualifications for the test set.

    4. Adjudication Method:

    The document does not specify any adjudication method (e.g., 2+1, 3+1, none) used for establishing ground truth in the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    The document makes no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being done. The focus is on the device's technical performance in generating comparable dose plans to CT, not on human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only) Performance:

    Yes, a standalone performance was done. The dose accuracy and geometric accuracy criteria, as well as the comparison to CT-based plans, are direct assessments of the algorithm's performance without human intervention in the primary image generation and initial dose calculation. The "robustness of the MRCAT brain algorithm" and its ability to produce "equivalent dose plans to CT" describe standalone performance.

    7. Type of Ground Truth Used:

    The primary ground truth used for assessing the device's performance, particularly dose accuracy and HU value calibration, is registered CT images and CT-based treatment plans. The comparisons are made against these CT data, which are considered the established standard for radiation attenuation estimation in radiotherapy planning.

    8. Sample Size for the Training Set:

    The document states that the Convolutional Neural Network (CNN) used in MRCAT image generation is "trained using matched pairs of CT and MRCAT source images." However, the exact sample size used for the training set is not specified.

    9. How the Ground Truth for the Training Set Was Established:

    The ground truth for the training set was established using "matched pairs of CT and MRCAT source images." This implies that CT images served as the reference or ground truth against which the MRCAT source images were processed and the CNN was trained to generate CT-like density information (HU values).

    Ask a Question

    Ask a specific question about this device

    K Number
    K182888

    Validate with FDA (Live)

    Device Name
    MRCAT Pelvis
    Date Cleared
    2019-04-30

    (197 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MRCAT Pelvis is a software add-on for Ingenia 1.5T and 3.0T MR systems.

    Intended Use:

    MRCAT imaging is intended to provide the operator with information of tissue properties for radiation estimation purposes in photon external beam radiotherapy treatment planning.

    Indications for use:

    MRCAT Pelvis is indicated for radiotherapy treatment planning of soft tissue cancers in the pelvic region.

    Device Description

    MRCAT Pelvis is a software application to Ingenia 1.5T and 3T MR systems. MRCAT Pelvis is available to the customer as an option to Ingenia MR-RT package, which is a set of accessories for Ingenia systems.

    Automated generation of MRCAT images takes place at the MR console of Ingenia. The embedded image post-processing runs in the background parallel to image acquisition. MRCAT algorithms enable automatic tissue characterization of five tissue types; air, fat, waterrich tissue, spongy bone and compact bone. Subsequent density assignment provides MRCAT images with CT-like density information.

    AI/ML Overview

    The provided text is a 510(k) summary for the Philips MRCAT Pelvis device. While it describes the device's intended use and general testing, it does not contain a detailed study demonstrating how the device meets specific acceptance criteria with performance metrics, sample sizes, ground truth establishment, or expert qualifications.

    Therefore, many of the requested details cannot be extracted directly from this document.

    Here's a breakdown of what can be inferred or is explicitly stated, and what is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, accuracy, etc. It only states that the robustness of the algorithm was shown by "producing equivalent dose plans to CT using gamma analysis with criterion of 3%/3mm." This implies an acceptance criterion related to dosimetric accuracy, but no precise performance numbers (e.g., % of plans meeting 3%/3mm gamma) are given.

    Acceptance Criteria (Implied)Reported Device Performance
    Production of equivalent dose plans to CT using gamma analysis."Robustness... shown by post-processing MRCAT images from patients, and calculating dose using the MRCAT images." (Criterion: 3%/3mm gamma analysis)
    Spatial accuracy of radiation attenuation estimates."MRCAT Pelvis images are spatially accurate radiation attenuation estimates."
    Software compliance with voluntary standards (list provided)."All requirements are met for the MRCAT Pelvis application."
    Successful completion of all tests performed for the software."All the tests performed for MRCAT Pelvis software were successful."
    No safety defects or hazardous situations from analyzed defects."All defects have been analyzed and are confirmed that they are not safety defects and will not cause any hazardous situation on using this application."

    2. Sample size used for the test set and the data provenance

    • Sample size: Not specified. The document only mentions "post-processing MRCAT images from patients." It does not provide the number of patients or cases.
    • Data provenance: Not specified (e.g., country of origin). It also doesn't explicitly state whether the data was retrospective or prospective, though "post-processing MRCAT images from patients" suggests retrospective analysis of data acquired for other purposes or as part of a study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not specified. The primary ground truth reference appears to be CT scans for dosimetric comparison, rather than human expert interpretation of the MRCAT images themselves for diagnostic purposes.


    4. Adjudication method for the test set

    Not applicable/specified. The primary comparison is between MRCAT-derived dose plans and CT-derived dose plans using gamma analysis, not subjective interpretation requiring adjudication among experts.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study is described. The device is a "software add-on" that generates CT-like density information from MR images for radiation treatment planning. It's not designed for human readers to interpret MR images directly for diagnosis with AI assistance, but rather to provide input for a treatment planning system. Therefore, an MRMC study in the traditional sense of human reader performance with/without AI assistance is not relevant to this device's function as described.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone evaluation was performed. The robustness of the "MRCAT Pelvis algorithm" was evaluated by comparing its output (dose plans from MRCAT images) against a reference (dose plans from CT) using gamma analysis. This is an algorithm-only performance evaluation.


    7. The type of ground truth used

    The ground truth for evaluating dosimetric accuracy appears to be CT imaging and subsequent dose plan calculations based on CT data. The MRCAT device aims to provide "CT-like density information" and "equivalent dose plans to CT."


    8. The sample size for the training set

    Not specified. Information about the training set size or methodology for the MRCAT algorithm is not provided in this document.


    9. How the ground truth for the training set was established

    Not specified. How the ground truth for any training data (if machine learning was used extensively to develop the underlying algorithm) was established is not mentioned in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151435

    Validate with FDA (Live)

    Device Name
    MRCAT
    Date Cleared
    2016-02-25

    (273 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MRCAT is used with Ingenia 1.5T and 3.0T MR systems.

    Intended Use:

    MRCAT imaging is intended to provide the operator with information of tissue properties for radiation attenuation estimation purposes in photon external beam radiotherapy treatment planning.

    Indications for use:

    MRCAT is indicated for radiotherapy treatment planning for prostate cancer patients.

    Device Description

    MRCAT is a software application to Ingenia 1.5T and 3T MR systems. MRCAT is available to the customer as an option to Ingenia MR-RT package, which is a set of accessories for Ingenia systems.

    Automated generation of MRCAT images takes place at the MR console of Ingenia. The embedded image post-processing runs in the background parallel to image acquisition. MRCAT algorithms enable automatic tissue characterization of five tissue types; air, fat, water-rich tissue, spongy bone and compact bone. Subsequent bulk-density assignment provides MRCAT images with CT-based density information.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria and the comprehensive study that proves the device meets them in the format requested. However, it offers some insights into the validation process and the claims made.

    Here's an attempt to extract and infer information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document mentions gamma analysis with criterion of 3%/3mm as a method to show the robustness of the MRCAT algorithm for producing equivalent dose plans to CT using MRCAT images. This implies that the acceptance criterion is that the dose plans generated using MRCAT images should be equivalent to those generated using CT, with a 3%/3mm gamma pass rate.

    Acceptance CriteriaReported Device Performance
    Robustness of MRCAT algorithm for producing equivalent dose plans to CT using gamma analysis with criterion of 3%/3mm (implicitly, a high pass rate would be required)"The robustness of the MRCAT algorithm for producing equivalent dose plans to CT using gamma analysis with criterion of 3%/3mm is shown by post-processing MRCAT images from prostate cancer patients, and calculating dose using the MRCAT images." (No specific pass rate percentage is provided in this summary.)

    Additional acceptance criteria related to software development and validation were mentioned in the "Summary of Non-Clinical Tests," but these are general quality assurance measures rather than specific performance metrics for the final output:

    • Risk Analysis
    • Testing on unit level (Subsystem verification)
    • Integration testing (System verification)
    • Final acceptance testing (Validation)
    • Performance testing (Verification)
    • Safety testing (Verification)

    The document states, "All the tests performed for MRCAT software were successful. All defects have been analyzed and are confirmed that they are not safety defects and will not cause any hazardous situation on using this application."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states, "The robustness of the MRCAT algorithm... is shown by post-processing MRCAT images from prostate cancer patients."

    • Sample Size: The exact number of prostate cancer patients used in the clinical test is not specified.
    • Data Provenance: The document does not specify the country of origin or whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The ground truth for dose plan comparison is implicitly assumed to be CT-based dose plans, but the expert involvement in establishing this ground truth or comparing the MRCAT-derived plans is not detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No MRMC comparative effectiveness study involving human readers and AI assistance is described. The device's primary function is to generate images for treatment planning, not to assist human interpretation directly in a diagnostic workflow. The "clinical test" described focuses on the comparison of dose plans generated by the algorithm (MRCAT) against CT, not human performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the clinical test described appears to be a standalone performance evaluation of the MRCAT algorithm. The "robustness of the MRCAT algorithm for producing equivalent dose plans to CT" was shown by "post-processing MRCAT images... and calculating dose using the MRCAT images." This implies an algorithm-to-algorithm comparison (MRCAT-generated dose plans vs. CT-generated dose plans).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for comparison in the clinical test was CT-based dose plans. The intent is for MRCAT images to provide "radiation attenuation estimation purposes in photon external beam radiotherapy treatment planning," implying that the standard for attenuation estimation is conventional CT. The gamma analysis compares the dose distributions calculated from MRCAT images to those calculated from CT images.

    8. The sample size for the training set

    This information is not provided in the document. The description focuses on the clinical evaluation of the algorithm's output rather than its development or training data.

    9. How the ground truth for the training set was established

    This information is not provided in the document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1