Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K213418

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-08-30

    (679 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MONTAGE Settable, Resorbable Hemostatic Bone Putty is indicated for the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade. The material may be used during surgical procedures and in treating traumatic injuries. MONTAGE is also indicated for use in the control of bleeding from bone surfaces in cardiothoracic surgery following sternotomy.

    Device Description

    MONTAGE Settable. Resorbable Hemostatic Bone Putty is a sterile, biocompatible, resorbable material for use in the control of bleeding from bone surfaces. The MONTAGE device comprises two separate components of putty-like consistency containing granular calcium phosphate, calcium stearate, vitamin E acetate, a triglyceride, a polyalcohol and a mixture of a lactide-diester and polyester-based polymers. When mixed together, the components of the MONTAGE device form a cohesive putty-like material that adheres to the bleeding bone surface and remains in place following application. The resulting hardened, resorbable material is primarily calcium phosphate. MONTAGE components must be mixed immediately prior to use. When applied to surgically cut or traumatically broken bone, MONTAGE Settable, Resorbable Hemostatic Bone Putty achieves local control of bleeding by acting as a mechanical barrier (tamponade).

    AI/ML Overview

    The provided text describes the Montagne Settable, Resorbable Hemostatic Bone Putty. The document is an FDA 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive de novo performance data.

    Here's an analysis of the provided information concerning acceptance criteria and studies:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not provide a table with specific, quantitative acceptance criteria for the device's hemostatic performance (e.g., time to hemostasis, success rate within a defined timeframe). Instead, it states more general performance outcomes.

    Acceptance Criterion (Implicit)Reported Device Performance
    Ability to achieve hemostasis"Studies have been conducted which demonstrate that Montage can achieve hemostasis..."
    Acceptable bone remodeling"...and allows for acceptable bone remodeling."
    Biocompatibility"Montage has been previously tested to demonstrate biocompatibility through an appropriate series of studies as required for compliance with ISO 10993."
    Acceptable handling characteristics"Montage has been evaluated for handling characteristics..."
    Acceptable shelf-life/stability"...and shelf-life / stability."
    Endotoxicity"Each lot is evaluated for endotoxicity."
    Sterility Assurance Level (SAL)"The device is provided sterile with a Sterility Assurance Level (SAL) of 10-6."
    Efficacy in sternotomy procedures"A human clinical evaluation was conducted that supports the use of Montage during sternotomy procedures." (No specific metrics or success rates from this study are detailed in this summary.)
    Maintenance of technological characteristics vs. predicate"The technological characteristics of the devices are unchanged. The only change is the addition of a statement regarding the use of the devices in sternotomy procedures." (This is a statement of design intention, implying no performance degradation due to the added indication).

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "A human clinical evaluation was conducted that supports the use of Montage during sternotomy procedures." However, it does not specify the sample size for this clinical evaluation, nor does it explicitly state the data provenance (e.g., country of origin, retrospective or prospective nature of the study).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not provide any information about the number or qualifications of experts used to establish ground truth for the test set of the clinical evaluation.

    4. Adjudication Method for the Test Set:

    The document does not describe an adjudication method for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    The document does not mention or describe a MRMC comparative effectiveness study, nor does it discuss human improvement with or without AI assistance. This is expected as the device is a medical device (putty), not an AI-powered diagnostic tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    The document does not describe a standalone performance study in the context of an algorithm. This is not applicable to a physical medical device like bone putty.

    7. Type of Ground Truth Used:

    For the "human clinical evaluation" mentioned, the ground truth would implicitly be clinical outcomes (e.g., successful hemostasis, absence of re-bleeding, safety events) as observed and documented by the operating surgeons and medical staff. However, the document does not explicitly state the specific ground truth used. For other performance tests (biocompatibility, sterility, etc.), the ground truth would be established through standardized testing protocols and laboratory measurements as per ISO 10993 and other relevant standards.

    8. Sample Size for the Training Set:

    The document does not specify a training set sample size. As a physical medical device (bone putty) undergoing a 510(k) submission, the concept of a "training set" in the context of machine learning algorithms is not directly applicable. Performance is typically established through a combination of bench testing, animal studies (if applicable), and clinical evaluation.

    9. How the Ground Truth for the Training Set Was Established:

    Since there is no "training set" in the context of an algorithm, this question is not applicable. The performance characteristics for the device were likely established through the aforementioned laboratory tests and clinical evaluation, with ground truth based on observed biological or physical responses to the device.

    In summary, the provided document is a 510(k) summary focused on substantial equivalence for a medical device. It briefly mentions performance testing and a clinical evaluation but lacks specific details typically found in summaries for AI/ML-powered devices, such as sample sizes, expert qualifications, or detailed performance metrics. The main purpose of this 510(k) was to add a new indication for use (sternotomy procedures) based on demonstrated equivalence to the previously cleared version of the same device and a predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152005

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-10-15

    (87 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MONTAGE Settable, Resorbable Hemostatic Bone Putty is indicated for the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.

    Device Description

    MONTAGE Settable, Resorbable Hemostatic Bone Putty is a sterile, biocompatible, resorbable material of putty-like consistency for use in the control of bleeding from bone surfaces. The single use MONTAGE device contains two separate components of putty-like consistency comprised of granular calcium phosphate, calcium stearate, vitamin E acetate, a triglyceride, a polyalcohol and a mixture of a lactide-diester and polyester-based polymers. When mixed together, the components of the MONTAGE device form a cohesive putty-like material that adheres to the bleeding bone surface and remains in place upon application. The resulting hardened, resorbable material is primarily calcium phosphate. MONTAGE must be mixed immediately prior to use.

    When applied to surgically cut or traumatically damaged bone, MONTAGE Settable, Resorbable Hemostatic Bone Putty achieves local control of bleeding by acting as a mechanical barrier (tamponade).

    AI/ML Overview

    The medical device in question is the MONTAGE™ Settable, Resorbable Hemostatic Bone Putty.

    This device is a bone hemostat intended for the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.

    1. A table of acceptance criteria and the reported device performance

    The provided document describes the MONTAGE device as substantially equivalent to a predicate device, Orthocon HBP4 Hardening, Resorbable Hemostatic Bone Putty (K141502). The acceptance criteria are therefore implicitly aligned with demonstrating this substantial equivalence. The document does not list specific numerical acceptance criteria with corresponding performance values for the MONTAGE device itself, but rather states that it is "exactly the same device" as the predicate, with the only difference being the packaging configuration. Therefore, the performance criteria and results are based on the predicate device.

    Acceptance Criteria CategoryPredicate Device Performance / Evidence for MONTAGE
    Intended UseMONTAGE has the same indicated intended use as the predicate: "control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade."
    Technological CharacteristicsThe document states MONTAGE is "exactly the same device" with the same materials (granular calcium phosphate, calcium stearate, vitamin E acetate, a triglyceride, a polyalcohol and a mixture of a lactide-diester and polyester-based polymers), radiopacity, resorbability, resorption time (>30 days), method of application, intended use (bone hemostasis), degradation process (dissolution, hydrolysis, cellular removal), mechanism of action (mechanical tamponade), sterility (gamma irradiation), and form of device (two-part settable putty).
    Bench TestingTesting performed on the predicate HBP4™ demonstrated acceptable handling properties (relative stiffness, spreadability, stickiness), temperature sensitivity, electrocautery compatibility, dissolution, and swelling. Since MONTAGE has the identical formulation, these results are considered relevant for MONTAGE.
    BiocompatibilityTesting performed on the predicate HBP4™ met ISO 10993 recommendations for irritation, sensitization, acute systemic toxicity, implantation, subacute systemic toxicity, chronic systemic toxicity, hemolysis, endotoxicity, and pyrogenicity. Since MONTAGE has the identical formulation and sterilization method (gamma irradiation), these results are considered relevant for MONTAGE.
    Animal Functionality TestingAnimal studies performed on the predicate HBP4™ demonstrated intraoperative in vivo hemostasis, resistance to irrigation, and characterized resorption time. Since MONTAGE has the identical formulation, these results are considered relevant for MONTAGE.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for the bench, biocompatibility, or animal studies conducted on the predicate device.

    • Bench Testing: No specific sample sizes mentioned.
    • Biocompatibility Testing: No specific sample sizes mentioned. Conducted in accordance with GLP requirements.
    • Animal Testing: No specific sample sizes mentioned.

    The data provenance is not specified regarding the country of origin. The studies appear to be prospective in nature, as they were conducted to evaluate the predicate device's safety and performance for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The studies mentioned (bench, biocompatibility, animal) typically involve laboratory testing and observation rather than expert ground truth establishment for a diagnostic output.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and therefore not provided in the document. The studies described are not clinical trials requiring adjudication of outcomes by multiple experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and therefore not provided. The device is a bone hemostat, not an AI-powered diagnostic tool, so MRMC studies and AI assistance metrics are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable and therefore not provided. The device is a physical medical product, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically applied to diagnostic algorithms is not directly applicable here. Instead, the studies focused on:

    • Bench Testing: Engineering measurements and physical property characterization.
    • Biocompatibility: Standardized biological assays and histological evaluations.
    • Animal Testing: In vivo observations of hemostasis, resistance to irrigation, and histological assessment of resorption.

    8. The sample size for the training set

    This information is not applicable and therefore not provided. The device is a physical product, not an algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable and therefore not provided. As stated above, the device does not involve a training set for an algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1