The intended use of this device is to fold and deliver Alcon ACRYSOF® intraocular lenses into the eye for replacement of the human crystalline lens
The MONARCH IOL Delivery System consists of two parts: an autoclavable, reusable titanium handbiece and a sterile, single-use cartridge. It is a device used for folding and delivering ACRYSOF intraocular lenses into the eye for replacement of the human crystalline lens. The system provides a controlled means to reliably place the ACRYSOF intraocular lens into the capsular bag.
This document is a 510(k) premarket notification for the Alcon MONARCH IOL Delivery System, submitted in 1998. It describes a medical device and its intended use, rather than a study on an AI/Machine Learning device. Therefore, much of the requested information regarding acceptance criteria, study design for AI evaluation, ground truth, and expert involvement is not applicable.
However, I can extract the available information and explain why other sections are not present in this type of submission.
Applicable Information:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Maintain overall diameter of the IOL | The performance test demonstrated that the MONARCH IOL Delivery System can be used to deliver ACRYSOF IOLs without affecting the overall diameter. |
| Maintain haptic angle of the IOL | The performance test demonstrated that the MONARCH IOL Delivery System can be used to deliver ACRYSOF IOLs without affecting the haptic angle. |
| Maintain power of the IOL | The performance test demonstrated that the MONARCH IOL Delivery System can be used to deliver ACRYSOF IOLs without affecting the power. |
| Maintain resolution of the IOL | The performance test demonstrated that the MONARCH IOL Delivery System can be used to deliver ACRYSOF IOLs without affecting the resolution. |
| Maintain cosmetic attributes of the IOL | The performance test demonstrated that the MONARCH IOL Delivery System can be used to deliver ACRYSOF IOLs without affecting the cosmetic attributes. |
| Smooth delivery of the IOL | The test lenses were delivered smoothly. |
| Complete unfolding of the IOL in a controlled fashion | The test lenses unfolded completely in a controlled fashion. |
Reasoning for Other Missing Information (Specific to AI/ML devices):
The provided document is a 510(k) submission for a physical medical device (an intraocular lens delivery system), not an AI/Machine Learning diagnostic or assistive tool. Therefore, the concepts of "test set," "training set," "ground truth established by experts," "MRMC studies," "standalone performance," and "data provenance" in the context of AI evaluation are not relevant to this document.
The "performance data" mentioned in the document refers to physical testing of the device's ability to handle and deliver an IOL without damage, not the performance of an algorithm.
Summary of Device and its "Study" (from the document's perspective):
- Device Name: MONARCH IOL Delivery System
- Intended Use: To fold and deliver Alcon ACRYSOF® intraocular lenses into the eye for replacement of the human crystalline lens.
- "Study" (Performance Test): The document refers to a "performance test" that demonstrated the device's ability to deliver ACRYSOF IOLs without causing damage or affecting their critical characteristics. This is a functional and integrity test of the physical device.
- Ground Truth: For this type of physical device, the "ground truth" is the established specifications and integrity of the IOLs themselves. The device is expected to deliver them without altering these pre-established qualities.
- Sample Size for Test Set: Not explicitly stated, but implied to be a sufficient number of ACRYSOF IOLs to conduct the performance test.
- Data Provenance: Not applicable in the context of AI/ML. The "data" here refers to the outcomes of physical testing.
- Number/Qualifications of Experts: Not applicable in the context of AI/ML. The evaluation would have been conducted by engineers and quality control personnel.
- Adjudication Method: Not applicable.
- MRMC Comparative Effectiveness Study: Not applicable (this is not an AI assistive device).
- Standalone Performance: Not applicable (this is a physical delivery system, not an algorithm).
- Type of Ground Truth: Physical integrity and functional performance of the IOLs post-delivery.
- Sample Size for Training Set: Not applicable (no "training" in the ML sense).
- How Ground Truth for Training Set was established: Not applicable.
Ask a specific question about this device
(110 days)
The intended use of this device is to fold and deliver Alcon ACRYSOF® intraccular lenses into the eye for replacement of the human crystalline lens.
The MONARCH™ IOL Delivery System consists of two parts: an autoclavable, reusable titanium handpiece and a sterile, single-use cartridge. It is a device used for folding and delivering ACRYSOF intraocular lenses into the eye for replacement of the human crystalline lens. The system provides a controlled means to reliably place the ACRYSOF® intraocular lens into the capsular bag.
This document is a 510(k) summary for the MONARCH™ IOL Delivery System. It describes the device, its intended use, and claims substantial equivalence to predicate devices. However, it does not provide the level of detail requested in the prompt regarding acceptance criteria and performance study specifics that would typically be found for a modern AI/ML medical device submission.
Here's a breakdown based on the provided text, highlighting what's missing:
1. A table of acceptance criteria and the reported device performance
- Acceptance Criteria: Not explicitly stated in a tabular format with numerical targets. The document mentions the device should deliver IOLs "without affecting the overall length, haptic angle, power, resolution, or cosmetic attributes of the lenses." It also states lenses "unfolded completely within 30 seconds." These are qualitative and quantitative performance aspects, but not presented as formal acceptance criteria with limits.
- Reported Device Performance:
- "The performance test demonstrated that the MONARCH™ IOL Delivery System can be used to deliver ACRYSOF® IOLs without affecting the overall length, haptic angle, power, resolution, or cosmetic attributes of the lenses."
- "The test lenses were delivered smoothly and unfolded completely within 30 seconds in a controlled fashion."
Missing Information: Specific numerical values for "overall length, haptic angle, power, resolution, or cosmetic attributes" and what constitutes "without affecting" (e.g., tolerance limits).
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size: Not specified. The document only refers to "the test lenses."
- Data Provenance: Not specified. It does not mention country of origin or whether the study was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable. This device is a mechanical delivery system, not an AI/ML diagnostic or prognostic tool that relies on expert interpretation for ground truth. The "performance test" would likely involve laboratory measurements rather than human expert assessment of a test set.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. See point 3.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is a mechanical device, not an AI system. Therefore, an MRMC study is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This is a mechanical device, not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" for this device would be based on physical measurements and observations of the delivered IOLs (e.g., integrity, dimensions) and the delivery process itself (e.g., smoothness, unfolding time). It is not based on expert consensus, pathology, or outcomes data in the context of an AI/ML system.
8. The sample size for the training set
- Not applicable. This document describes a mechanical device. There is no concept of a "training set" in the context of AI/ML algorithm development for this type of product. The "test lenses" mentioned are for performance evaluation, not training.
9. How the ground truth for the training set was established
- Not applicable. See point 8.
In summary: The provided 510(k) summary is for a traditional medical device (an IOL delivery system). The prompt's questions are largely geared towards the evaluation and validation of Artificial Intelligence/Machine Learning (AI/ML) powered medical devices. Therefore, most of the specific details requested regarding ground truth, expert adjudication, training/test sets, and comparative effectiveness (MRMC) studies are not applicable to this document. The "performance data" provided relates to the physical and functional characteristics of the mechanical device.