Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K103233

    Validate with FDA (Live)

    Date Cleared
    2011-02-03

    (93 days)

    Product Code
    Regulation Number
    888.3353
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MDMIM liner and X30 Acetabular Inserts are a sterile, single-use device intended for use in primary and revision total hip arthroplasty to alleviate pain and restore function. These devices are intended to be used only with currently available Howmedica Osteonics 22.2 mm and 28 mm diameter femoral heads.

    The indications for use for total hip arthroplasty include:

    • Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
    • Rheumatoid arthritis;
    • Correction of functional deformity;
    • Revision procedures where other treatments or devices have failed;
    • Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
    • Dislocation risks

    The MDM™ liners are intended for cementless use only.

    Device Description

    The Modular Dual Mobility (MDM™) liner is a highly polished cobalt chrome liner that features a Trident® locking mechanism. This feature will allow for compatibility with other acetabular cups containing the same locking mechanism. A Duration® or X3® polyethylene insert will articulate within the MDM™ liner. The additional X3® Acetabular Inserts are intended to accommodate the size range of the MDM™ liner device.

    AI/ML Overview

    This 510(k) summary describes the Modular Dual Mobility (MDM™) Liner and X3® Acetabular Insert, which are components for artificial hip replacement. The document focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through clinical studies. Therefore, much of the requested information regarding acceptance criteria, study details, and expert involvement is not present in the provided text.

    Here is an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (from document)Reported Device Performance (from document)
    Mechanical PerformanceDisassembly force evaluation (ASTM F1820-03)Performed, demonstrating "equivalence of the subject device compared to its predicate devices."
    Push-out strength (ASTM F1820-03)Performed, demonstrating "equivalence of the subject device compared to its predicate devices."
    Lever-out force testingPerformed, demonstrating "equivalence of the subject device compared to its predicate devices."
    Pull-out force testingPerformed, demonstrating "equivalence of the subject device compared to its predicate devices."
    Biomechanical PerformanceRange of motionEngineering analysis performed.
    Jump distanceEngineering analysis performed.
    Fretting wearEngineering analysis performed.
    Material CompatibilityNot explicitly stated as a criterion, but inherent in the choice of materials (Cobalt Chrome, X3® polyethylene).Implied to be acceptable through substantial equivalence to predicate devices.
    BiocompatibilityNot explicitly stated as a criterion, but a standard requirement for implantable devices.Implied to be acceptable through substantial equivalence to predicate devices.

    Missing Information: The document states that testing was performed to "demonstrate equivalence" to predicate devices. However, it does not provide specific acceptance values (e.g., minimum push-out strength in Newtons) or the exact quantitative results achieved by the MDM™ Liner and X3® Acetabular Insert for the mechanical or biomechanical tests. The K103233 does not contain a formal "Clinical Study" section with detailed performance results against pre-defined acceptance criteria for the new device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not mentioned. The document primarily refers to "testing" and "engineering analysis" to demonstrate equivalence rather than a specific clinical or performance test set with a defined sample size. These tests are typically benchtop mechanical tests on a limited number of device samples.
    • Data Provenance: Not explicitly stated. For benchtop mechanical testing, the "provenance" would typically refer to the manufacturing batch or lot of the devices tested. It is not clinical data, so country of origin of patient data or retrospective/prospective status is not applicable here.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable. This information is relevant for studies involving human data and expert review (e.g., image interpretation, clinical outcome assessment). The provided document describes engineering and mechanical tests aimed at demonstrating equivalence to predicate devices, not clinical performance assessed by experts.

    4. Adjudication Method for the Test Set

    • Not Applicable. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among multiple experts when establishing ground truth in studies involving human assessment. This is not relevant to the engineering and mechanical tests described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    • Not Applicable. This device is a hip implant component. The provided documentation does not involve Artificial Intelligence (AI) or human readers interpreting data. Therefore, an MRMC study and AI assistance effect size are not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    • Not Applicable. This question pertains to AI algorithm performance. The device is a physical medical implant, not an algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for the mechanical and engineering tests described would be the established scientific and engineering principles, material properties, and the performance characteristics of the predicate devices as defined by their own testing and regulatory clearance. For example, for push-out strength, the "ground truth" is a measurable force value, and the new device's performance is compared against that of the predicate to demonstrate equivalence.

    8. The Sample Size for the Training Set

    • Not Applicable. This information is relevant for AI models, which require training data. This device is a physical implant.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. This information is relevant for AI models.

    In summary: The provided 510(k) summary for the Modular Dual Mobility (MDM™) Liner and X3® Acetabular Insert focuses on demonstrating substantial equivalence through benchtop engineering and mechanical testing against a set of predicate devices. It does not provide detailed quantitative acceptance criteria or specific performance data in the context of a clinical study or a study directly addressing human performance or AI. Therefore, much of the requested information, which is typically found in clinical trial reports or AI/CADe system submissions, is not contained within this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1