510(k) Data Aggregation

    K Number
    K080143

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-02-14

    (23 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput -T3), the Stryker Spine OASYS® System is intended for:

    • Degenerative Disc Disease (as defined by neck and back pain of discogenic origin with . degeneration of the disc confirmed by history and radiographic studies)
    • Spondylolisthesis .
    • Spinal Stenosis .
    • Fracture/Dislocation .
    • Atlanto/axial fracture with instability .
    • Occipitocervical dislocation .
    • Revision of previous cervical spine surgery .
    • Tumors .

    When used with the occipital plate, the bone screws are limited to occipital fixation only. The bone screws are not intended to be used in the cervical spine.

    The use of the polyaxial screws is limited to placement in the upper thoracic spine (TI-T3) in treating thoracic conditions only. They are not intended to be placed in the cervical spine.

    The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1 -T3) spine.

    The Stryker Spine OASYS® System can also be linked to the Xia System, SR90D System and Xia 4.5 Spinal System via the rod-to-rod connectors.

    Device Description

    The Stryker Spine OASYS® System is comprised of rods, polyaxial screws, bone screws, hooks, connectors, and an occiput plate. The components are available in a variety of lengths in order to accommodate patient anatomy. The components are fabricated from Titanium alloy and CP Titanium and are provided non-sterile. The Stryker Spine OASYS® System can be linked to the Stryker Spine Xia® Spinal System, Xia 4.5 System and SR90D System via the rod-to-rod connectors.

    This Special 510(k) submission is intended to introduce a line extension to the predicate OASYS™ System, which consists of the addition of a cross connector plate and associated components (connector blocker and nut).

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a line extension to the Stryker Spine OASYS® System, introducing a cross connector plate and associated components. This document is not for a new medical device that utilizes AI or produces performance metrics in the way your request describes. Therefore, most of the requested information regarding acceptance criteria and studies proving the device meets them (especially related to AI performance, expert ground truth, adjudication, MRMC studies, or standalone algorithm performance) is not applicable to this type of submission.

    Here's a breakdown of what can be extracted from the document based on your request, and why other parts are not provided:


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implicit)Reported Device Performance (Implicit)
    Mechanical properties equivalent to predicate device."Testing has demonstrated that the additional cross connector components have equivalent mechanical properties to the predicate OASYS® System (K032394)." (Page 2)
    Intended use aligns with predicate device."Both the new components and the existing system components are intended to address the same indications for use." (Page 2) The detailed Indications for Use for the extended system are provided on page 4, which are consistent with the predicate's purpose of spinal fusion in the cervical and occipitocervico-thoracic junction for conditions like Degenerative Disc Disease, Spondylolisthesis, etc.
    Materials are equivalent to predicate device."Both the new components and the existing components are made from the same materials [Titanium alloy and CP Titanium]." (Page 2, and implied from predicate description on Page 1)
    (Implicit) No new safety or effectiveness concerns are introduced.The FDA’s determination of "substantial equivalence" (Page 2-3) indicates that the device meets the regulatory requirements and is as safe and effective as its predicate.

    Explanation: This 510(k) is for a line extension of an established spinal fixation system, not a device with a measurable performance output like an AI algorithm. The acceptance criteria are implicit: the new components must be mechanically equivalent to the predicate, made of the same materials, and address the same indications for use without raising new questions of safety or efficacy. The "study" proving this is a mechanical testing comparison.


    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable. This submission describes mechanical testing of components, not a clinical study on human subjects where a "test set" of data would be used. The "test set" would refer to the physical components tested in a lab, but the number is not specified.
    • Data Provenance: Not applicable. The "data" comes from mechanical testing of manufactured components, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. There is no "ground truth" to be established by experts in the context of mechanical testing of a spinal implant. The "ground truth" is determined by established engineering standards and tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are used for resolving disagreements in expert assessments of medical images or other clinical data, which is not relevant here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-powered device and no reader studies were performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth: Mechanical testing standards and engineering specifications. The "ground truth" for proving substantial equivalence lies in demonstrating that the mechanical properties of the new components match those of the predicate and that the materials are identical.

    8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device, so there is no training set.

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for an AI model, this question does not apply.

    Summary regarding the device:

    This 510(k) document is a regulatory submission for a mechanical medical device (spinal implant components), specifically a line extension. It focuses on demonstrating substantial equivalence to a predicate device by showing that the new components have the same intended use, materials, and equivalent mechanical properties. The regulatory pathway is based on established engineering principles and material science, not on clinical performance metrics, AI algorithms, or human reader studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062853

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-10-24

    (29 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput-T3), the Stryker Spine OASYSTM System is intended for:

    • Degenerative Disc Disease (as defined by neck or back pain of discogenic origin with . degeneration of the disc confirmed by history and radiographic studies)
    • Spondvlolisthesis .
    • Spinal Stenosis .
    • Fracture/Dislocation ●
    • Atlanto/axial fracture with instability .
    • Occipitocervical dislocation .
    • Revision of previous cervical spine surgery .
    • Tumors .

    When used with the occipital plate, the bone screws are limited to occipital fixation only. The bone screws are not intended to be used in the cervical spine.

    The use of the polyaxial screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. They are not intended to be placed in the cervical spine.

    The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.

    The Stryker Spine OASYSTM System can also be linked to the Xia System, SR90D System and Xia 4.5 Spinal System via the rod-to-rod connectors.

    Device Description

    This Special 510(k) submission is intended to introduce a line extension to the predicate OASYSTM System, which consists of two rod-to-rod connectors (axial and parallel versions) used to link the 3.5mm OASYS™ System rods with the Xia® 4.5 Spinal System rods (510(k) numbers #K050461 and #K060361). The Xia® 4.5 Spinal System rods will retain their original cleared name.

    AI/ML Overview

    This document is a Special 510(k) Premarket Notification for a line extension to the Stryker Spine OASYS™ System. It does not describe a study proving a device meets acceptance criteria in the traditional sense of a clinical or performance study with numerical targets. Instead, the submission focuses on demonstrating substantial equivalence to a predicate device based on material composition and validated mechanical properties.

    Therefore, the requested information elements related to clinical study design, ground truth, expert adjudication, sample sizes for test/training sets, and comparative effectiveness studies are not applicable in this context.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    The primary "acceptance criterion" for this 510(k) is demonstrating equivalent mechanical properties to the predicate device.

    Acceptance CriterionReported Device Performance
    Equivalent mechanical properties to the predicate OASYS™ System (K032394)"Testing has demonstrated that the additional rod-to-rod connector components have equivalent mechanical properties to the predicate OASYS™ System (K032394)."
    Same indications for use as the predicate device"Both the new components and the existing system components are intended to address the same indications for use."
    Made from the same materials as the predicate device"Both the new components and the existing components are made from the same materials."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. This document describes a Special 510(k) submission for a line extension, not a clinical study involving human or animal subjects that would require a "test set" and associated data provenance in this manner. The "testing" mentioned refers to mechanical property validation on the device components themselves.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. See point 2. The "ground truth" here is the established mechanical properties of the predicate device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. See point 2.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This submission is for spinal fixation components, not an AI or imaging device involving human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This submission is for spinal fixation components, not an algorithm or software device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" in this context is the established mechanical properties of the predicate device, as determined by engineering standards and testing, and the material composition of the predicate device.

    8. The sample size for the training set

    • Not applicable. There is no "training set" in the context of this type of device submission.

    9. How the ground truth for the training set was established

    • Not applicable. There is no "training set" in the context of this type of device submission. The "ground truth" for demonstrating substantial equivalence (as described in point 7) would have been established through engineering design, material specifications, and standardized mechanical testing processes.