510(k) Data Aggregation

    K Number
    K042969

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-11-08

    (11 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    0 - 0.083
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMART™ Control™ Nitinol Stent Transhepatic Biliary System is intended for use in the palliation of malignant neoplasms in the biliary tree.

    Device Description

    The device description of the proposed SMART™ Control™ Nitinol Stent Transhepatic Biliary System is as follows.
    • 6 French stent delivery system profile;
    • Stent material – Nickel Titanium alloy and tantalum micromarkers;
    • Expanded stent diameters 9 and 10 mm;
    • Stent lengths: 80 mm;
    • Stent delivery system usable length 80 and 120 cm;
    • Guidewire lumen 0.035";

    AI/ML Overview

    The provided text describes a medical device, the SMART™ Control™ Nitinol Stent Transhepatic Biliary System, and its substantial equivalence to predicate devices, but it does not contain information about acceptance criteria or a study proving the device meets specific acceptance criteria.

    The document is a 510(k) summary and the FDA's response letter, which focuses on regulatory approval based on demonstrating substantial equivalence to existing legally marketed devices. This regulatory pathway typically relies on comparison to predicate devices, and often involves pre-clinical testing, rather than extensive clinical studies with specific performance acceptance criteria as would be seen for new clinical efficacy claims or performance validation.

    Here's a breakdown of why this information is missing based on the input provided:

    • No specific acceptance criteria are listed. The document states "Performance standards have not been established by the FDA under section 514 of the Food, Drug and Cosmetic Act."
    • No detailed study is described that measures performance against acceptance criteria. The document states, "The equivalence was confirmed through pre-clinical testing," but does not detail the methodology, sample sizes, ground truth establishment, or specific results of this testing.

    Therefore, I cannot fulfill your request for a table of acceptance criteria, device performance, study details (sample sizes, provenance, expert involvement, adjudication, MRMC, standalone performance, ground truth types), or training set information as this information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032457

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-09-04

    (24 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMART™ Control™ Nitinol Stent Transhepatic Biliary System is intended for use in the palliation of malignant neoplasms in the biliary tree.

    Device Description

    The device description of the proposed SMART™ Control™ Nitinol Stent Transhepatic Biliary System is as follows. 6 French stent delivery system profile; Stent material – Nickel Titanium alloy and tantalum micromarkers; Expanded stent diameters 6, 7, 8, 9, and10 mm; Stent lengths: 20, 30, 40, 60, 80, and 100 mm; Stent delivery system usable length 80 and 120 cm; and Guidewire lumen 0.035". Proximal Deployment Handle

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (SMART™ Control™ Nitinol Stent Transhepatic Biliary System) and a letter from the FDA. It does not include information about acceptance criteria or a study proving device performance in the context of the requested details about AI/algorithm performance.

    The document discusses:

    • Device Name: SMART™ Control™ Nitinol Stent Transhepatic Biliary System
    • Predicate Devices: SMART™ Control Nitinol Stent Transhepatic Biliary System (K021898 and K031777)
    • Indications for Use: Palliation of malignant neoplasms in the biliary tree.
    • Device Description: Stent material (Nickel Titanium alloy and tantalum micromarkers), stent dimensions (diameters, lengths), delivery system (French profile, usable length, guidewire lumen), and proximal deployment handle.
    • Biocompatibility: Stated as biocompatible.
    • Substantial Equivalence: Confirmed through pre-clinical testing, comparing it to predicate devices.
    • FDA Communication: A letter affirming substantial equivalence but requiring specific warnings regarding use in the vascular system and prominent display of the biliary indication.

    The document does not contain any information related to:

    • Acceptance criteria in the context of an algorithm or AI.
    • Reported device performance as would be detailed for an AI.
    • Sample sizes for test sets, training sets, or data provenance for an AI study.
    • Number of experts, their qualifications, or adjudication methods for ground truth establishment for an AI.
    • MRMC comparative effectiveness studies or human reader improvement with AI.
    • Standalone algorithm performance.
    • Type of ground truth (e.g., pathology, outcomes data) for an AI.

    Therefore, I cannot provide the requested table and study details as they pertain to AI/algorithm performance based on the given input. The "study" mentioned is "pre-clinical testing" used to establish substantial equivalence for a physical medical device, not an AI evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031197

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-06-16

    (61 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMART™ Control™ Nitinol Stent Transhepatic Biliary System is intended for use in the palliation of malignant neoplasms in the biliary tree.

    Device Description

    The device description of the proposed SMART™ Control™ Nitinol Stent Transhepatic Biliary System is as follows:
    6 French stent delivery system profile;
    Stent material - Nickel Titanium alloy and tantalum micromarkers;
    Expanded stent diameters 9, and 10 mm
    Stent lengths: 100 mm;
    Stent delivery system usable length 80 and 120 cm; and
    Guidewire lumen 0.035".
    Proximal Deployment handle

    AI/ML Overview

    I am sorry, but the provided text describes a medical device (a biliary stent) and its regulatory clearance (510(k) submission). It does not contain information about studies or acceptance criteria that would allow me to populate the requested table and answer the questions about device performance against acceptance criteria, sample sizes, ground truth establishment, or human reader performance with AI assistance. The document is primarily a regulatory summary stating substantial equivalence to predicate devices based on pre-clinical testing, rather than a clinical study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023217

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-10-25

    (29 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMART™ Control™ Nitinol Stent Transhepatic Biliary System is intended for use in the palliation of malignant neoplasms in the biliary tree.

    Device Description

    The device description of the proposed SMART™ Control™ Nitinol Stent Transhepatic Biliary System is as follows.
    • 6 French stent delivery system profile;
    • Stent material - Nickel Titanium alloy and tantalum micromarkers;
    • Expanded stent diameters 9 and 10 mm;
    • Stent lengths: 80 mm;
    • Stent delivery system usable length 80 and 120 cm; Guidewire lumen 0.035";
    • Proximal Deployment Handle.

    AI/ML Overview

    The provided text is related to a 510(k) submission for a medical device (SMART™ Control™ Nitinol Stent Transhepatic Biliary System) and describes its substantial equivalence to predicate devices, rather than a study outlining acceptance criteria and detailed device performance metrics.

    Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth for an AI/ML device is not applicable or cannot be extracted from the given document.

    However, I can provide the following based on the available text:

    Acceptance Criteria and Device Performance (Not Applicable - This document is a 510(k) summary for a substantial equivalence determination for a medical stent, not a performance study with explicit acceptance criteria for a new device's efficacy as per your request's format.)

    The document focuses on establishing substantial equivalence to predicate devices through pre-clinical testing, rather than defining and proving specific performance metrics against pre-defined acceptance criteria for a novel AI/ML device.


    Here's an attempt to answer the questions based on the type of document provided:

    1. A table of acceptance criteria and the reported device performance

      • Not applicable. This document is a 510(k) summary for a medical stent, asserting "substantial equivalence" to predicate devices, not reporting performance against explicit, quantifiable acceptance criteria in the manner typically seen for AI/ML diagnostic devices. The equivalence was confirmed through "pre-clinical testing," but no specific metrics or targets are provided in this summary.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Not applicable. This document does not pertain to a study with a "test set" in the context of AI/ML or diagnostic performance. It refers to "pre-clinical testing" for a physical medical device. No sample size, data provenance, or study type (retrospective/prospective) is mentioned.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not applicable. This document does not describe a study involving expert-established ground truth for a test set.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable. This document does not describe a study involving adjudication for a test set.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not applicable. This document does not describe an MRMC study or AI assistance.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not applicable. This document is not about an algorithm/AI device.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • Not applicable. This document does not describe a study requiring ground truth in the context of an AI/ML device. The "pre-clinical testing" likely refers to bench testing, mechanical stress tests, biocompatibility assessments, and potentially animal studies, which establish performance against engineering specifications or biological responses, not a "ground truth" for diagnostic accuracy.
    8. The sample size for the training set

      • Not applicable. This document does not describe a training set as it is not about an AI/ML device.
    9. How the ground truth for the training set was established

      • Not applicable. This document does not describe a training set or its ground truth.