510(k) Data Aggregation

    K Number
    K081092

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-12-11

    (603 days)

    Product Code
    Regulation Number
    866.6040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MammaPrint® is a gualitative in vitro diagnostic test service, performed in a single laboratory, using the gene expression profile of fresh breast cancer tissue samples to assess a patients' risk for distant metastasis (up to 10 years for patients less than 61 years old, up to 5 years for patients ≥ 61 years).

    The test is performed for breast cancer patients with Stage I or Stage II disease, with a tumor size of ≤ 5.0 cm and lymph node negative. The MammaPrint result is indicated for use by physicians as a prognostic marker only, along with other clinicopathological factors.

    Device Description

    The MammaPrint service is a microarray based gene expression analysis of a tumor. The analysis is based on several processes: isolation of RNA from frozen tumor tissue sections, DNA'se treatment of isolated RNA, linear amplification and labeling of DNA'se treated RNA, cRNA purification, hybridization of the cRNA to the MammaPrint microarray, scanning the MammaPrint microarray and data acquisition (feature extraction), index calculation and determination of the risk of distant recurrence in breast cancer patients.

    The MammaPrint analysis is designed to determine the gene activity of specific genes in a tissue sample compared to a reference standard. The result is an expression profile, or fingerprint, of the sample.

    The correlation of the sample expression profile to a template (the mean expression profile of 44 tumors with a known good clinical outcome) is calculated and the molecular profile index of the sample is determined (Low Risk, High Risk).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the MammaPrint® device, based on the provided text:

    MammaPrint® Acceptance Criteria and Study Information

    1. Acceptance Criteria and Reported Device Performance

    The provided document primarily focuses on analytical performance as internally validated and clinical performance as demonstrated through peer-reviewed studies. It doesn't explicitly state "acceptance criteria" in a typical pass/fail numerical sense for clinical performance, but rather presents the performance achieved by the device in various clinical settings.

    Acceptance Criteria CategorySpecific Metric/DescriptionReported Device Performance/Finding
    Analytical PerformanceAccuracy of measurement (based on repeated experiments of control samples)98.5% Analytical Accuracy of measurement
    Accuracy of classifying a sample as High Risk or Low RiskAt least 98.9% (i.e., 1.1% false negative classification)
    Percentage of "Borderline Samples"Less than 5% of analyzed samples are considered "Borderline Samples"
    Classification accuracy for "Borderline Samples"Approximately 90% (i.e., 10% chance of false classification)
    Clinical PerformanceAssessment of risk for distant metastasis (up to 10 years for patients < 61 years) (Nature Paper)Within 5 year metastasis risk by profile multivariate OR 18
    Metastasis-free survival by profile at 10 years (low risk vs. high risk) (NEJM Paper)Low risk profile 87%, high risk profile 44% (at 5 yrs: 93% and 56% respectively)
    Highly reproducible MammaPrint as diagnostic tool (MammaPrint Paper)Highly reproducible (demonstrated by comparison to Nature and NEJM studies)
    Metastasis-free survival by profile at 10 years (low risk vs. high risk) (Transbig Paper - European validation)Low risk profile 88%, high risk profile 71% (at 5 yrs: 96% and 83% respectively)
    Metastasis-free survival by profile at 5 years (low risk vs. high risk) in older age patients (55-87 yrs) ("Older Age" NKI/AVL series)Low risk profile 94%, high risk profile 75%
    Overall Conclusion"Equivalent performance in the age extended patient group up to 87 years old at 5 years metastasis free survival."Achieved: "MammaPrint is a clinically and analytically accurate prognostic marker for providing a risk assessment of distant metastasis of breast cancer."
    "MammaPrint analytical methodology is identical to the FDA cleared MammaPrint device with 510k number K070675."Substantial equivalence shown for analytical performance; therefore, no need to show substantial equivalence for analytical performance for this submission (K081092) beyond confirming identity with the predicate device's methods. Validation studies listed above demonstrate clinical performance in specific populations and with extended age range compared to the predicate.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document describes several clinical studies, which served as the basis for clinical performance validation. These studies effectively act as the "test set" for demonstrating clinical utility.

    StudySample Size (Test Set)Data Provenance
    Nature Paper (1)78 patientsRetrospective (implied by "Development of breast cancer prognosis 70-gene profile")
    NEJM Paper (2)151 patientsRetrospective ("consecutive series of breast cancer patients")
    MammaPrint Paper (3)Not specifiedFocused on reproducibility of prior studies on MammaPrint.
    Transbig Paper (4)302 patientsRetrospective ("Independent European validation"). Implies multiple European countries by "European validation."
    "Older Age" NKI/AVL series (5)177 patientsRetrospective ("consecutive series of breast cancer patients"). NKI/AVL is a Dutch cancer institute, implying Dutch data.
    Analytical Performance ValidationMore than 190 independent analyses (for accuracy)Agendia (The Netherlands), and a three-way inter-laboratory comparison study between three independent laboratories in three different countries (Dutch, French and U.S.).
    Analytical Performance Validation2500 samples / 5000 hybridizations (for QC cut-off)Performed at Agendia (The Netherlands).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The MammaPrint test determines a "risk of distant recurrence." The "ground truth" in these studies is the actual clinical outcome, specifically distant metastasis-free survival. This type of ground truth is established through long-term follow-up of patients, typically overseen by clinical oncologists and other medical professionals, rather than a panel of experts reviewing images or other diagnostic outputs to establish a consensus. The studies are retrospective analyses of patient cohorts with known clinical outcomes.

    Therefore, the concept of "number of experts used to establish ground truth" with specific qualifications in the traditional sense (e.g., radiologists for imaging) does not directly apply here. The "ground truth" is the observed patient outcome over time, documented by treating physicians and medical records.

    4. Adjudication Method for the Test Set

    Given that the ground truth is patient outcomes (distant metastasis-free survival) over many years, an adjudication method for a "test set" like 2+1 or 3+1 typically used for medical image review is not applicable. Clinical outcomes are factual events recorded over time.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study is mentioned in the provided text. MammaPrint is a gene expression profiling test, not an imaging device or a tool directly interpreted by human readers in the same way an MRI or X-ray would be. Its output is a risk classification (Low Risk, High Risk), which is directly provided to the physician for consideration alongside other clinical factors. The improvement in human reader performance with AI assistance is not a relevant metric for this type of device.

    6. Standalone (Algorithm Only) Performance

    Yes, the studies presented demonstrate the standalone performance of the MammaPrint algorithm. The clinical studies (Nature, NEJM, Transbig, Older Age NKI/AVL) evaluate the prognostic power of the MammaPrint profile (generated by the algorithm) in predicting distant metastasis-free survival. The reported metastasis-free survival rates for "low risk profile" and "high risk profile" directly reflect the algorithm's ability to stratify patients based solely on its gene expression analysis.

    7. Type of Ground Truth Used

    The type of ground truth used for the clinical studies is patient outcomes data, specifically distant metastasis-free survival. This is a robust and objective measure of a patient's disease progression.

    8. Sample Size for the Training Set

    The document states that the Nature Paper (1) describes the "Development of breast cancer prognosis 70-gene profile (LNO, <55)" using 78 patients. This study likely represents the initial cohort used to develop or train the 70-gene signature that underlies MammaPrint. While not explicitly called a "training set" in the modern machine learning sense, this cohort was instrumental in defining the profile.

    The MammaPrint algorithm itself correlates a sample's expression profile to a template (the mean expression profile of 44 tumors with a known good clinical outcome). These 44 tumors could be considered part of the "training" or reference data for the algorithm's core functionality.

    9. How the Ground Truth for the Training Set Was Established

    For the "Nature Paper" (which describes the development of the 70-gene profile):

    • The ground truth would have been established through long-term clinical follow-up and patient outcomes data, similar to the test sets. Patients in this development cohort would have had their distant metastasis status recorded over a period of years (e.g., 5-year metastasis risk).
    • For the 44 tumors used to create the "template" for the MammaPrint Index calculation, their "known good clinical outcome" would have been established based on observed excellent patient prognosis/long-term metastasis-free survival from clinical records.
    Ask a Question

    Ask a specific question about this device

    K Number
    K080252

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-07-21

    (172 days)

    Product Code
    Regulation Number
    866.6040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MammaPrint is a qualitative in vitro diagnostic test service, performed in a single laboratory, using the gene expression profile of fresh breast cancer tissue samples to assess a patient's risk for distant metastasis.

    The test is performed for breast cancer patients who are less than 61 years old, with Stage I or Stage II disease, with tumor size <= 5.0 cm and who are lymph node negative. The MammaPrint result is indicated for use by physicians as a prognostic marker only, along with other clinicopathological factors.

    Device Description

    The MammaPrint service is a microarray based gene expression analysis of a tumor. The analysis is based on several processes: isolation of RNA from frozen turnor tissue sections, DNA'se treatment of isolated RNA, linear amplification and labeling of DNA'se treated RNA, cRNA purification, hybridization of the cRNA to the MammaPrint microarray, scanning the MammaPrint microarray and data acquisition (feature extraction), calculation and determination of the risk of recurrence in breast cancer patients.

    The MammaPrint analysis is designed to determine the gene activity of specific genes in a tissue sample compared to a reference standard. The result is an expression profile, or fingerprint, of the sample.

    The correlation of the sample expression profile to a template (the mean expression profile of 44 tumors with a known good clinical outcome) is calculated and the molecular profile of the sample is determined (Low Risk, High Risk).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the MammaPrint® device, based on the provided text:

    MammaPrint® Device Performance Study Analysis

    1. Acceptance Criteria and Reported Device Performance

    The provided document primarily focuses on demonstrating substantial equivalence to a predicate device (Agendia BV's MammaPrint (K070675)) rather than setting explicit, quantifiable clinical acceptance criteria for de novo approval. However, it does present analytical performance targets relative to the predicate and clinical outcomes from supporting studies.

    Here's a table summarizing the implicit acceptance criteria (based on predicate performance) and the reported device performance from the analytical studies:

    Acceptance Criteria / Performance MetricTarget (derived from predicate or equivalency)Reported Device Performance (New MammaPrint HD)
    Analytical Performance
    Repeatability and Precision (MammaPrint index Standard Deviation)At least as good as predicate (≤ 0.030)At least as good as predicate (0.030)
    Repeatability and Precision (MammaPrint index Variance)At least as good as predicate (≤ 0.001)At least as good as predicate (0.001)
    Concordance in MammaPrint outcome (HD vs. LD microarray)Within 97.7% technical accuracy of predicate98.9%
    MammaPrint Index difference between scanners (mean, median, standard deviation)Within predicate's accepted variance (1.96 * 0.030)Falls within accepted variance
    Accuracy of classifying as High Risk or Low RiskBetter than predicate (97.7% accuracy, 1.1% false negative)98.9% (0.5% false negative classification)
    Percentage of "Borderline Samples"Not explicitly stated as a target, but reported for contextLess than 5%
    "Borderline Samples" Classification AccuracyNot explicitly stated as a target, but reported for contextApproximately 90% (10% chance of false classification)

    2. Sample Size and Data Provenance for Test Set (Analytical Performance)

    • Precision Evaluation (Method Validation):
      • Sample Size: 3 samples with different outcome levels (high, low, borderline) were run repeatedly over 20 days. Two replicates per sample level per day. Also included High Risk Control (HRC) and Low Risk Control (LRC). Total 120 (3 samples x 2 replicates x 20 days), plus controls.
      • Data Provenance: Not explicitly stated, but likely internal laboratory data from Agendia BV. Retrospective, as these were established samples.
    • LD and HD Microarray Comparison:
      • Sample Size: 98 historical MammaPrint service samples.
      • Data Provenance: Samples from the period 2004 through 2007. Origin country is not explicitly stated in this section, but the company is based in the Netherlands. This is retrospective data.
    • Scanner Performance Comparison:
      • Sample Size: 26 newly hybridized slides (representing 104 samples). This included three samples with varying risk profiles, repeated, and LRC/HRC controls.
      • Data Provenance: Likely internal laboratory data from Agendia BV. Retrospective.

    For Clinical Performance, the document refers to four published studies. These studies serve as the clinical evidence for the device's utility, rather than a single "test set" in the context of analytical validation. The details provided are for each study:

    • Nature Paper (1): 78 patients
    • NEJM Paper (2): 151 patients
    • MammaPrint Paper (3): Not a clinical study for patient outcomes, focuses on reproducibility of (1) and (2).
    • Transbig Paper (4): 302 patients

    The provenance of these clinical studies would generally be international (e.g., European for Transbig). All are retrospective analyses of patient cohorts.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    For the analytical performance studies, "ground truth" refers to the known characteristics of the samples (e.g., whether they were designed to be high, low, or borderline risk, or previously classified by the predicate device). No human experts are used to establish ground truth in these analytical comparisons; instead, it relies on the pre-determined characteristics of the control samples or the results from the predicate device.

    For the extensive clinical studies referenced (Nature, NEJM, Transbig), the ground truth for patient outcomes (e.g., metastasis-free survival) would have been established through long-term clinical follow-up by medical professionals, including oncologists and pathologists. The number and specific qualifications of these experts are not detailed in this 510(k) summary, as it refers to published peer-reviewed literature. Typically, these studies involve panels of clinicians (e.g., pathologists, oncologists) determining initial diagnosis and patient outcomes.

    4. Adjudication Method for the Test Set

    • Analytical Performance: No adjudication method as human interpretation is not involved in comparing array results or scanner outputs. The comparisons are quantitative.
    • Clinical Performance (Referenced Studies): The 510(k) summary does not detail the specific adjudication methods used in the referenced clinical studies (e.g., for outcome endpoints like metastasis). Such details would typically be found within the full publications.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study is mentioned or implied. MammaPrint is a gene expression profiling test, which is an automated, quantitative assay. Human readers do not directly interpret the "cases" (gene expression profiles) in the same way they would an image. The device provides a quantitative risk score (Low Risk, High Risk) that physicians use to make decisions, but the device itself does not involve human interpretation for its output. Therefore, a study demonstrating how human readers improve with AI vs. without AI assistance is not applicable here.

    6. Standalone Performance Study

    Yes, the entire submission describes the standalone performance of the MammaPrint device (the algorithm/system itself) in classifying tissue samples as High Risk or Low Risk based on gene expression profiles. The analytical performance metrics (precision, reproducibility, classification accuracy) are all measures of the algorithm's performance in isolation from human interpretation.

    7. Type of Ground Truth Used

    • Analytical Performance:
      • Known outcome levels for control samples (high, low, borderline).
      • Results from the predicate device (FDA cleared Low Density microarray).
    • Clinical Performance (Referenced Studies):
      • Long-term clinical follow-up data: Metastasis-free survival.
      • Clinicopathological factors (e.g., tumor size, lymph node status, age).

    8. Sample Size for the Training Set

    The document mentions: "The correlation of the sample expression profile to a template (the mean expression profile of 44 tumors with a known good clinical outcome) is calculated and the molecular profile of the sample is determined (Low Risk, High Risk)." This "template" of 44 tumors appears to be a core component of the algorithm's development.

    The broader clinical studies referenced (e.g., Nature Paper (1) with 78 patients) were likely instrumental in the initial development and refinement (training) of the 70-gene profile itself. The MammaPrint Paper (3) describes the "Development of MammaPrint" and "Converting a breast cancer microarray signature into a high-throughput diagnostic test," implying further training and optimization steps using data from earlier studies.

    Therefore, the initial "training set" for defining the gene profile involved patient cohorts (e.g., the 78 patients in the Nature paper, or the 44 specific "good clinical outcome" tumors).

    9. How the Ground Truth for the Training Set Was Established

    For the "44 tumors with a known good clinical outcome" that form the template, the "known good clinical outcome" would have been established through:

    • Long-term clinical follow-up: Patients whose tumors formed this template would have had a prolonged period without recurrence or distant metastasis.
    • Clinicopathological data: Standard clinical assessments (e.g., pathology reports, surgical findings) would have confirmed the characteristics of these tumors and the patients' disease stage.

    For the broader training of the 70-gene profile, as implied by the referenced papers, the ground truth was established by:

    • Clinical outcome data: Patients were followed for several years (e.g., 5-10 years) to determine their metastasis-free survival or overall survival.
    • Pathology: Original biopsy/surgical pathology diagnoses, including tumor type, grade, and lymph node status, were used.
    • Expert Consensus: While not explicitly stated for individual training cases in this summary, the clinical outcomes and pathological assessments that defined "good clinical outcome" and "poor clinical outcome" would have been made by qualified medical professionals (pathologists, oncologists).