510(k) Data Aggregation

    K Number
    K052246

    Validate with FDA (Live)

    Date Cleared
    2005-08-22

    (5 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    0 - 75
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The en-bloc Biopsy System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities.

    The en-bloc Biopsy System is intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality.

    The en-bloc Biopsy System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

    The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpable or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

    In instances when a patient presents with a palpable abnormality that has been classified as benign by clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the en-bloc Biopsy System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histologic evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

    Device Description

    The en-bloc is a percutaneous high frequency, automated, vacuum-assisted electrosurgical device used to remove tissue by automated electrosurgical cutting and simultaneous capture of an incised tissue volume. The Neothermia en-bloc ™ consists of a hand-held biopsy handle, upon which the single-use en-bloc Biopsy Probe is attached, with an integral cable to connect the handle to the control unit. The Probe™ contains two sets of active electrodes at its distal end - a precursor electrode and cutting/capture electrodes. The shaft of the Probe™ is encased in a stainless steel cannula. An outer plastic sleeve surrounds this stainless steel cannula and an annular gap between the sleeve and the cannula provides a conduit for vacuum-assisted removal of the gaseous products of electrosurgical cutting and any liquids (e.g., blood) that may accumulate at the distal end of the Probe during the biopsy procedure. The vacuum also helps maintain the required cutting arc during automated tissue capture.

    AI/ML Overview

    The provided text is a 510(k) summary for the Neothermia Corporation's en-bloc Biopsy System™. It discusses the device's intended use and technological characteristics, as well as its substantial equivalence to predicate devices. However, the document does NOT contain information about acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, or specific performance metrics that would be typically found in a clinical study report.

    Therefore, I cannot provide the requested table and study details. The document is primarily a regulatory submission establishing substantial equivalence to existing devices, not a detailed technical report of a new study to prove device performance against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041347

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-06-04

    (15 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The en-bloc Biopsy System is intended for diagnostic sampling of breast tissue during a breast biopsy procedure.

    Device Description

    The en-bloc is a percutaneous high frequency, automated, vacuum-assisted electrosurgical device used to remove tissue by automated electrosurgical cutting and simultaneous capture of an incised tissue volume. The Neothermia en-bloc ™ consists of a hand-held biopsy handle, upon which the single-use en-bloc Biopsy Probe is attached, with an integral cable to connect the handle to the control unit. The Probe™ contains two sets of active electrodes at its distal end -- a precursor electrode and cutting/capture electrodes. The shaft of the Probe™ is encased in a stainless steel cannula. An outer plastic sleeve surrounds this stainless steel cannula and an annular gap between the sleeve and the cannula provides a conduit for vacuum-assisted removal of the gaseous products of electrosurgical cutting and any liquids (e.g., blood) that may accumulate at the distal end of the Probe during the biopsy procedure. The vacuum also helps maintain the required cutting arc during automated tissue capture.

    AI/ML Overview

    This 510(k) summary provides information about the Neothermia Corporation's en-bloc Biopsy System™, a device for diagnostic sampling of breast tissue. However, it does not contain the specific acceptance criteria or a study proving the device meets acceptance criteria in the way typically expected for an AI/ML medical device submission.

    This document describes a traditional medical device (electrosurgical generator and biopsy instrument), not an AI/ML powered device. The "Substantial Equivalence" section focuses on comparing its "intended use, principles of operation, and technological characteristics" to previously cleared predicate devices, specifically "Neothermia's cleared 10mm, 15mm and 20mm en-bloc probes," with an emphasis on a "material change" in the new probe.

    Therefore, many of the requested categories for AI/ML device studies (like sample size for test sets, expert qualifications, MRMC studies, standalone performance, training set details) are not applicable or not provided in this type of 510(k) submission.

    I will address the questions based on the information available and indicate when information is not present or not relevant to this type of device submission.


    1. A table of acceptance criteria and the reported device performance

    This document does not provide a table of acceptance criteria or specific reported device performance metrics in the way an AI/ML study would (e.g., sensitivity, specificity, AUC). Instead, the criterion for clearance is substantial equivalence to predicate devices.

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial Equivalence to legally marketed predicate devices- Same intended use
    - Same principles of operation
    - Similar technological characteristics (despite material change)
    - Performs diagnostic sampling of breast tissue

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable or not provided. This 510(k) submission for a non-AI/ML device typically does not include a test set sample size or data provenance in the context of an accuracy study. Substantial equivalence is often demonstrated through engineering testing, biocompatibility, sterilization validation, and comparison of design features and intended use.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth, in the context of an AI/ML algorithm's performance on clinical data, is not established for this type of device submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are not described for this type of medical device submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. An MRMC study is not relevant for this traditional electrosurgical biopsy system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm; it is a physical electrosurgical system requiring human operation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. The concept of "ground truth" for evaluating an algorithm's diagnostic performance is not applied in this type of 510(k) submission for a physical device. Performance is generally assessed through functional testing, safety testing, and comparative analysis against predicate device specifications.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no training set.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI/ML algorithm, ground truth establishment for a training set is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033427

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-11-10

    (13 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The en-bloc Biopsy System is intended for diagnostic sampling of breast tissue during a breast biopsy procedure.

    Device Description

    The en-bloc is a percutaneous high frequency, automated, vacuum-assisted electrosurgical device used to remove tissue by automated electrosurgical cutting and simultaneous capture of an incised tissue volume. The Neothermia en-bloc ™ consists of a hand-held biopsy handle, upon which the single-use en-bloc Biopsy Probe is attached, with an integral cable to connect the handle to the control unit. The Probe™ contains two sets of active electrodes at its distal end - a precursor electrode and cutting/capture electrodes. The shaft of the Probe™ is encased in a stainless steel cannula. An outer plastic sleeve surrounds this stainless steel cannula and an annular gap between the sleeve and the cannula provides a conduit for vacuum-assisted removal of the gaseous products of electrosurgical cutting and any liquids (e.g., blood) that may accumulate at the distal end of the Probe during the biopsy procedure. The vacuum also helps maintain the required cutting arc during automated tissue capture.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for the Neothermia Corporation's en-bloc Biopsy System™. It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria for a new device or algorithm.

    Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" are not directly applicable or available in this type of submission.

    Here's an analysis based on the provided text, highlighting what is and isn't present:

    Key Takeaway: This 510(k) submission is for a modified version of an already cleared device ("modified cradle") and aims to show it has the "same intended use, principles of operation, and technological characteristics as the previously cleared predicate devices." It explicitly states, "The change does not raise new questions of safety or efficacy." This type of submission does not typically include new performance studies with specific acceptance criteria in the way a novel device or AI algorithm would.


    1. A table of acceptance criteria and the reported device performance

    • Not Applicable. This submission is based on substantial equivalence to predicate devices, not on proving new performance against defined acceptance criteria with reported metrics. The document asserts that the modified device has the "same intended use, principles of operation, and technological characteristics" as existing cleared predicate devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. No specific test set data is described for performance evaluation in this 510(k) summary. Given it's a "modified cradle" and not a fundamentally new device, performance testing (if any) would likely be limited to verification and validation of the modification's impact, not a full-scale clinical or diagnostic accuracy study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. No test set requiring expert-established ground truth is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No test set is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is an electrosurgical biopsy system, not an AI-based diagnostic tool. Therefore, an MRMC study related to AI assistance would not be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a hardware device (electrosurgical biopsy system). There is no "algorithm only" component to test for standalone performance in the context of diagnostic accuracy.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not Applicable. No specific ground truth establishment for a performance study is mentioned in this submission. The device itself is intended for diagnostic sampling of breast tissue, implying that the tissue samples it yields would then be subject to pathology for definitive diagnosis.

    8. The sample size for the training set

    • Not Applicable. This is an electrosurgical device, not a machine learning algorithm. Therefore, there is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. No training set is involved.

    Summary regarding acceptance criteria and studies:

    The provided text from the 510(k) summary emphasizes substantial equivalence to predicate devices. The core argument is that the "modified cradle" of the en-bloc Biopsy System™ retains the "same intended use, principles of operation, and technological characteristics" as previously cleared devices, and that the modification "does not raise new questions of safety or efficacy."

    Therefore, this document does not contain information about:

    • Specific quantitative acceptance criteria for device performance (e.g., sensitivity, specificity, accuracy).
    • Clinical studies or diagnostic accuracy studies with sample sizes, data provenance, ground truth establishment, or reader performance metrics.
    • Any AI-related performance metrics or studies.

    The "study" implicitly referenced is the comparison to predicate devices, demonstrating that the modified device is similar enough not to warrant new, extensive performance studies typically required for novel devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022296

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-07-30

    (14 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The en-bloc Biopsy System is intended for diagnostic sampling of breast tissue during a breast biopsy procedure.

    Device Description

    The en-bloc is a percutaneous high frequency, automated, vacuum-assisted electrosurgical device used to remove tissue by automated electrosurgical cutting and simultaneous capture of an incised tissue volume. The Neothermia en-bloc ™ consists of a hand-held biopsy handle, upon which the single-use en-bloc Biopsy Probe is attached, with an integral cable to connect the handle to the control unit. The Probe™ contains two sets of active electrodes at its distal end - a precursor electrode and cutting/capture electrodes. The shaft of the Probe™ is encased in a stainless steel cannula. An outer plastic sleeve surrounds this stainless steel cannula and an annular gap between the sleeve and the cannula provides a conduit for vacuum-assisted removal of the gaseous products of electrosurgical cutting and any liquids (e.g., blood) that may accumulate at the distal end of the Probe during the biopsy procedure. The vacuum also helps maintain the required cutting arc during automated tissue capture.

    AI/ML Overview

    The provided text is a 510(k) summary for the Neothermia Corporation's en-bloc Biopsy System and an FDA clearance letter. It focuses on demonstrating substantial equivalence to pre-existing, legally marketed devices. It does not contain information about acceptance criteria, detailed study designs, or performance metrics in the way a clinical trial report or a comprehensive validation study would.

    Therefore, I cannot provide the requested information in the format you specified. The document explicitly states that "The dimensional change does not raise new questions of safety or efficacy. The modified probe is substantially equivalent to Neothermia's cleared en-bloc probes." This indicates that the clearance was based on equivalence to existing devices rather than new performance data against specific acceptance criteria.

    Here's why each point you requested is not available in the provided text:

    1. A table of acceptance criteria and the reported device performance: Not present. The submission focuses on substantial equivalence based on intended use, principles of operation, and technological characteristics, not on quantitative performance metrics or acceptance criteria.
    2. Sample size used for the test set and the data provenance: Not present. No test set or associated data is described.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not present. No ground truth establishment is described.
    4. Adjudication method for the test set: Not present. No adjudication is described.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not present. This device is an electrosurgical biopsy system, not an AI-assisted diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not present. This is a physical medical device, not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not present.
    8. The sample size for the training set: Not present. No training set is mentioned as this is not an AI/ML device.
    9. How the ground truth for the training set was established: Not present.

    In summary, the provided 510(k) summary and FDA clearance letter do not contain the type of detailed performance data and study information you requested because the clearance was granted based on substantial equivalence, not a de novo pathway requiring extensive new performance data against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021577

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-05-29

    (15 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The en-bloc Biopsy System is intended for diagnostic sampling of breast tissue during a breast biopsy procedure.

    Device Description

    The en-bloc is a percutaneous high frequency, automated, vacuum-assisted electrosurgical device used to remove tissue by automated electrosurgical cutting and simultaneous capture of an incised tissue volume. The Neothermia en-bloc ™ consists of a hand-held biopsy handle, upon which the single-use en-bloc Biopsy Probe is attached, with an integral cable to connect the handle to the control unit. The Probe™ contains two sets of active electrodes at its distal end - a precursor electrode and cutting/capture electrodes. The shaft of the Probe™ is encased in a stainless steel cannula. An outer plastic sleeve surrounds this stainless steel cannula and an annular gap between the sleeve and the cannula provides a conduit for vacuum-assisted removal of the gaseous products of electrosurgical cutting and any liquids (e.g., blood) that may accumulate at the distal end of the Probe during the biopsy procedure. The vacuum also helps maintain the required cutting arc during automated tissue capture.

    AI/ML Overview

    This submission (K021577) is a 510(k) premarket notification for the Neothermia Corporation's en-bloc Biopsy System™. It primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices rather than presenting a performance study with detailed acceptance criteria in the manner one might find for a novel AI/software component within a medical device.

    Therefore, the provided text does not contain the information requested regarding acceptance criteria related to a performance study, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, or detailed ground truth establishment.

    The document discusses:

    • The intended use of the device ("diagnostic sampling of breast tissue during a breast biopsy procedure").
    • Its technological characteristics (percutaneous high frequency, automated, vacuum-assisted electrosurgical device for tissue removal).
    • The components of the system (hand-held biopsy handle, single-use probe with electrodes, cannula, plastic sleeve, vacuum).
    • The basis of substantial equivalence: The 15mm en-bloc probe has the same intended use, principles of operation, and technological characteristics as the previously cleared predicate devices (Neothermia's 10mm and 20mm en-bloc probes).
    • A minor difference in technological characteristics (size of basket) is noted, but the submission claims this "does not raise new questions of safety or efficacy."

    Conclusion:

    Based on the provided text, it is not possible to extract the detailed information requested about acceptance criteria and a performance study. The submission relies on establishing substantial equivalence to existing devices, implying that the safety and effectiveness are supported by the predicate devices, not a new clinical performance study within this document.