Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K050073

    Validate with FDA (Live)

    Date Cleared
    2005-02-11

    (30 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AMPHIRION™ DEEP 0.014'' OTW PTA Balloon Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

    Device Description

    The AMPHIRION DEEP catheter is a coaxial lumen device with a semi-compliant balloon. The lumen marked "WIRE" is the central lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire with a maximum outer diameter of 0.014 inch. The lumen marked "BALLOON" is the balloon inflation lumen, which is used to inflate and deflate the dilatation balloon. The catheter tapers beneath the balloon segment to achieve the lowest possible deflated profile. Two radiopaque marker bands are placed under the balloon segment of the catheter shaft to provide visual reference points for balloon positioning within the vessel. The distal catheter shaft is hydrophilic coated.

    AI/ML Overview

    The provided text describes a Premarket Notification (510(k)) Summary for the AMPHIRION™ DEEP 0.014" OTW PTA Balloon Catheter. This type of submission is for demonstrating substantial equivalence to a predicate device, rather than proving novel effectiveness through clinical studies with acceptance criteria in the traditional sense.

    Therefore, many of the requested categories for acceptance criteria and study details are not applicable in this context. The "study" mentioned is primarily bench testing to support substantial equivalence for newly added balloon sizes, not a clinical trial evaluating diagnostic or treatment efficacy against specific performance metrics.

    Here's an analysis based on the provided text, addressing the applicable points and explaining why others are not relevant:


    Acceptance Criteria and Device Performance for AMPHIRION™ DEEP 0.014" OTW PTA Balloon Catheter

    1. A table of acceptance criteria and the reported device performance

    In the context of a 510(k) for a device like a PTA balloon catheter, "acceptance criteria" are not typically expressed as performance metrics like sensitivity/specificity for a diagnostic device or success rates for a novel therapeutic device against predefined thresholds. Instead, the acceptance criteria revolve around demonstrating substantial equivalence to a predicate device, meaning it has the same intended use, similar technological characteristics, and performs as safely and effectively.

    For this specific submission, the "performance" referenced primarily pertains to bench testing to support the substantial equivalence claim for the newly added balloon sizes. The document explicitly states: "Performance testing (bench) further supports a substantial equivalence claim. The collective evidence therefore provides assurance that the AMPHIRIONTM DEEP 0.014" OTW PTA Balloon Catheter meets the requirements that are considered acceptable for the intended use."

    The document does not provide specific numerical acceptance criteria (e.g., burst pressure tolerance, deflation time) or reported performance values from these bench tests. It simply states that the tests were performed and support the claim.

    Acceptance Criteria Category (Implied by 510(k))Reported Device Performance (Summary)
    Intended Use EquivalenceSame as predicate device (dilate stenoses in specified arteries and A-V fistulae).
    Technological Characteristics EquivalenceSimilar to predicate device (coaxial lumen, semi-compliant balloon, radiopaque markers, hydrophilic coating). Added balloon diameters (2.5mm, 3.0mm, 3.5mm, 4.0mm).
    Material EquivalenceSame as predicate device.
    Performance (Bench Testing)Bench testing supports substantial equivalence for added balloon sizes. No specific numerical metrics reported in this summary.
    Safety and EffectivenessDeemed safe and effective based on substantial equivalence claim and supporting bench data.

    2. Sample size used for the test set and the data provenance

    The document refers to "Performance testing (bench)."

    • Sample Size for Test Set: Not specified. Bench tests often involve a certain number of units per size/configuration to ensure statistical validity, but the exact count is not disclosed in this summary.
    • Data Provenance: The tests are "bench" tests, meaning in-vitro laboratory testing, not human or animal studies. Therefore, concepts like country of origin for patient data are not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. For bench testing of a device like a balloon catheter, "ground truth" is typically defined by engineering specifications, material properties, and physical measurements, not by expert clinical consensus. The evaluation is against engineering standards and requirements, not expert interpretations of clinical data.

    4. Adjudication method for the test set

    Not applicable. As described above, the "test set" involves bench testing against engineering specifications, not clinical cases requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical instrument (PTA balloon catheter), not a diagnostic imaging system or an AI-powered tool. Therefore, MRMC studies involving human readers or AI assistance are entirely irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used

    For the bench testing, the "ground truth" would be the engineering specifications and validated testing methodologies against which the device's physical properties and performance characteristics (e.g., burst pressure, inflation/deflation times, guidewire compatibility, shaft flexibility) are measured. These are objective measurements compared to predefined limits and the predicate device's performance.

    8. The sample size for the training set

    Not applicable. This is a medical device, not an AI or machine learning model that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1