Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K021052

    Validate with FDA (Live)

    Date Cleared
    2002-04-09

    (8 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated use- Measure of individual temperature
    Measurements- Oral, Axillary -under arm, Rectal
    Range of measurement- 90.0 degrees - 109.9 degrees F (32.0 degrees - 43 degrees C)
    Accuracy- +/-0.2 degrees F
    Targeted population- Individuals requiring temperature measurements
    Environment of use- Hospital and home
    Disposable / reusable- Reusable, clean with alcohol

    Device Description

    The intended product is an electronic digital thermometer for measuring patient temperature.

    AI/ML Overview
    {
      "acceptance_criteria_and_study": {
        "acceptance_criteria_table": {
          "headers": [
            "Parameter",
            "Acceptance Criteria",
            "Reported Device Performance"
          ],
          "rows": [
            [
              "Temperature range",
              "Not explicitly stated as an acceptance criterion for the subject device but compared to predicate device's 89.6-109.4° F",
              "90.0-109.9° F"
            ],
            [
              "Ambient temperature",
              "Not explicitly stated as an acceptance criterion for the subject device but compared to predicate device's 60.8-104° F",
              "60.8-104° F"
            ],
            [
              "Response time",
              "Not explicitly stated as an acceptance criterion for the subject device but compared to predicate device's ~1 minute",
              "About 20 seconds"
            ],
            [
              "Accuracy and Performance",
              "Meets ASTM E1112",
              "Meets ASTM E1112"
            ],
            [
              "Accuracy (for indications for use)",
              "+/- 0.2° F",
              "+/- 0.2° F (stated in Indications for Use, implies conformance)"
            ],
            [
              "Material safety",
              "Non-toxic and in compliance with EN ISO 10993",
              "All materials used are non-toxic and in compliance with EN ISO 10993"
            ]
          ]
        },
        "study_details": {
          "sample_size_test_set": "Not specified. The document states \"Performance testing,\" and notes that the device \"meets ASTM E1112,\" but does not detail the sample size for this testing.",
          "data_provenance": "Not specified. The testing was conducted to meet ASTM E1112 standards.",
          "number_of_experts_ground_truth": "Not applicable. The ground truth for performance and accuracy is based on compliance with a recognized standard (ASTM E1112) rather than expert interpretation of data.",
          "qualifications_experts": "Not applicable.",
          "adjudication_method": "Not applicable.",
          "mrmc_comparative_effectiveness_study": "No. This is a digital thermometer, not an AI-assisted diagnostic device, so an MRMC study comparing human readers with and without AI assistance is not relevant.",
          "standalone_performance_study": "Yes, a standalone performance study was conducted to demonstrate compliance with ASTM E1112. The data provenance and sample size are not explicitly detailed beyond the standard reference.",
          "type_of_ground_truth": "Standard compliance (ASTM E1112) for accuracy and performance, and safety standard compliance (EN ISO 10993) for materials.",
          "sample_size_training_set": "Not applicable. This is a physical device, not an AI algorithm requiring a training set.",
          "ground_truth_training_set_established": "Not applicable."
        }
      }
    }
    
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1