510(k) Data Aggregation

    K Number
    K974341

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-04-16

    (149 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To hold a Semi Tendonosus ST (soft tissue) graft in place in the femoral tunnel during the healing period following ACL reconstrution surgery.

    Femoral fixation of autograft or allograft ACL soft tissue grafts

    Device Description

    The Cross Pin is 1.654 " long by 0.132" dia. with a conical tip by 0.180" long. The Cross Pin is molded from Purac PLA.

    AI/ML Overview

    The provided text describes a submission for a medical device (Mitek 3.3mm ST Cross Pin) seeking 510(k) clearance. The document focuses on comparing the new device to a predicate device (Acufex EndoButton) to establish substantial equivalence based on non-clinical tests.

    Here's an analysis of the acceptance criteria and study information, based only on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Fixation Strength (Bench Tests)Statistical equivalence (in both strength and stiffness) to the Acufex EndoButton.
    Bending (Bench Tests)Statistical equivalence to the Acufex EndoButton.
    Shear (Bench Tests)Statistical equivalence to the Acufex EndoButton.
    Effect of Cross Pin Orientation on Ultimate Strength of Hamstring GraftsIn-vitro evaluation tests were conducted to show this effect. (No specific performance metric or "acceptance" is stated for this; it's a descriptive test.)

    2. Sample size used for the test set and the data provenance

    The text mentions "Bench fixation strength tests, bending and shear" and "In-Vitro evaluation tests." However, it does not specify the sample size used for these tests. It also does not mention the country of origin of the data or whether it was retrospective or prospective, as these were bench and in-vitro tests, not human trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable and not provided in the document. The "tests" performed were non-clinical bench and in-vitro evaluations of mechanical properties, not diagnostic or clinical assessments requiring expert ground truth establishment in the traditional sense for medical imaging or clinical outcome studies.

    4. Adjudication method for the test set

    This information is not applicable and not provided. Adjudication methods are typically used in clinical studies or expert reviews where there's a need to resolve discrepancies in observations or diagnoses. Since these were bench tests, an adjudication method for a test set is not relevant.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. The document explicitly states: "No clinical tests were done." This device is a mechanical implant, and the study was focused on its mechanical properties compared to a predicate device, not on diagnostic accuracy or AI assistance for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable and not provided. The device is a physical medical implant (a cross pin), not an algorithm or AI system. Therefore, standalone performance in the context of an algorithm is irrelevant.

    7. The type of ground truth used

    For the bench fixation strength, bending, and shear tests, the "ground truth" was established by direct measurement of physical properties (strength, stiffness) of the devices under controlled experimental conditions. For the in-vitro evaluation tests, the "ground truth" was the directly measured ultimate strength of the hamstring grafts under various cross pin orientations. This is based on objective, quantifiable physical properties rather than expert consensus or pathology in a biological sense.

    8. The sample size for the training set

    This information is not applicable and not provided. The assessment was a direct comparison of physical properties of two medical devices. There was no "training set" in the context of machine learning or predictive modeling.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided for the same reason as point 8. There was no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K974291

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-03-05

    (111 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To hold a bone tendon bone graft in place by cross pinning in the femoral tunnel during a period of time following surgery allowing the natural healing process(es) to occur.
    Femoral fixation of autograft or allograft ACL Bone Tendon Bone Grafts

    Device Description

    The Cross Pin is 1.654″ long by 0.105″ dia with a conical tip by .180 long. The Cross Pin is molded from Purac(PLA). To hold a bone tendon bone graft in place by cross pinning in the femoral tunnel during a period of time following surgery allowing the natural healing process(es) to occur.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies:

    This document describes a medical device submission for a Staple, Fixation, Bone, & soft tissue device, specifically the Mitek BTB Absorbable (PLA) Cross Pin. The submission aims to demonstrate substantial equivalence to a predicate device, the Linvatec PLA Bioscrew.

    It's important to note that the provided text focuses on non-clinical (benchtop) tests to demonstrate substantial equivalence. No clinical tests were performed or required for this submission. Therefore, questions related to human reader performance, AI assistance, or clinical ground truth are not applicable in this context.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Mitek BTB Absorbable (PLA) Cross Pin were implicitly established by demonstrating comparable performance to the predicate device, the Linvatec PLA Bioscrew, in specific mechanical tests.

    | Acceptance Criteria (Implicit) | Reported Device Performance (Mitek Cross Pin) |
    |----------------------------------------------------------------+|---------------------------------------------------------------------------------------------------------------------|
    | No significant strength difference in bending tests compared to predicate (Bioscrew). | "Bench top Bending and Shear tests were done on the PLA pins [...] The results showed that there was no significant strength difference between the two." (Implied comparison to Bioscrew) |
    | No significant strength difference in shear tests compared to predicate (Bioscrew). | "Bench top Bending and Shear tests were done on the PLA pins [...] The results showed that there was no significant strength difference between the two." (Implied comparison to Bioscrew) |
    | Comparable pull-out strength to predicate (Bioscrew). | "strength comparisons to the Bioscrew in pull out tests were done to the Bioscrew. The results showed that there was no significant strength difference between the two." |
    | Comparable strength after 8 weeks (in-vitro) to "zero" day results (indicating stability). | "Eight week In-vitro tests were also done and found to have results that were comparable to 'zero' day results." |

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for the benchtop tests (e.g., number of pins tested for bending, shear, or pull-out). It simply states that "Bench top Bending and Shear tests were done on the PLA pins" and "strength comparisons to the Bioscrew in pull out tests were done to the Bioscrew."

    The data provenance is laboratory-based benchtop testing. There is no information about country of origin, as it pertains to manufactured devices and their testing, not patient data. The tests are non-clinical and therefore cannot be classified as retrospective or prospective in the human study sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. For this type of non-clinical device submission, ground truth is established through standardized engineering and material science testing methods, not by human expert review of clinical data.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 refer to methodologies used in clinical studies for resolving discrepancies among expert readers. Since this submission relies on benchtop testing, such methods are not relevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No MRMC comparative effectiveness study was done. The submission explicitly states: "No clinical tests were done for this product." Therefore, there is no effect size related to human readers improving with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This device is a physical medical implant, not an AI algorithm. Therefore, there is no "standalone" algorithm performance to assess.

    7. The Type of Ground Truth Used

    The ground truth used was mechanical performance data obtained from standardized benchtop testing (Bending, Shear, and Pull-out tests) of the device and a predicate device, as well as in-vitro degradation/stability testing.

    8. The Sample Size for the Training Set

    Not applicable. Clinical training sets for AI algorithms were not used. The manufacturing process of the pins serves as the "training" for the device itself, adhering to design specifications.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set in the AI sense for this device. The "ground truth" for manufacturing would be the established design specifications, material properties, and quality control parameters.