Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121692

    Validate with FDA (Live)

    Date Cleared
    2012-11-05

    (151 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MiniMe 2 Nasal mask is intended to provide a patient interface for application of positive pressure therapy. The mask is to be used as an accessory to CPAP or Bi-level positive pressure systems. It is intended for patients >2 years old and <12 years old. The mask is designed and labeled for different durations of use:

    • Single use, disposable (hospital or institutional settings) .
    • Single patient, multi-use (home setting) .
    • Single patient, short-term use up to 7 days (hospital or institutional settings) .
    Device Description

    The MiniMe 2 Nasal mask is intended to provide a patient interface for application of positive pressure therapy. The mask is to be used as an accessory to CPAP or Bi-level positive pressure systems. It is intended for patients >2 years old and <12 years old. The mask is designed and labeled for different durations of use:

    • Single use, disposable (hospital or institutional settings) .
    • Single patient, multi-use (home setting) .
    • Single patient, short-term use up to 7 days (hospital or institutional settings) .
    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device, the MiniMe 2 Nasal Mask, seeking substantial equivalence to predicate devices. It does not describe an AI/ML device or a study involving human readers or ground truth established by experts/pathology/outcomes data. Therefore, much of the requested information (like sample sizes for test/training sets, data provenance, number and qualifications of experts, adjudication methods, MRMC studies, standalone AI performance, and how ground truth was established for training) is not applicable or cannot be extracted from this document.

    The document primarily focuses on demonstrating that the MiniMe 2 Nasal Mask is substantially equivalent to existing predicate devices (SleepNet MiniMe pediatric nasal mask and ResMed Kidsta Pediatric Mask) through comparative performance testing and material biocompatibility.

    Here's the information that can be extracted based on the provided text, adapted to the questions where applicable:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by demonstrating equivalence to the predicate devices in various performance aspects. The "reported device performance" are the actual measurements for the MiniMe 2 Nasal Mask, which are then compared to the predicates.

    Acceptance Criteria (Implicitly, Equivalence to Predicate)MiniMe 2 Nasal Mask PerformanceSleepNet MiniMe Nasal Mask (K090935) Performance (Predicate 1)ResMed Kidsta (K060105) Performance (Predicate 2)
    Indications for Use: Provide patient interface for positive pressure therapy, accessory to CPAP/Bi-level, for patients >2 yo and <12 yo. Different durations of use: single use/disposable, single patient multi-use (home), single patient short-term use (up to 7 days).IdenticalIdentical (Except for MiniMe 2 adding "Single use, disposable (hospital or institutional settings)" as a shorter duration which is also considered equivalent)Intended to provide non-invasive ventilation with RI or OSA (Broader, but MiniMe 2 aligns with SleepNet MiniMe)
    Patient Population: >2 yo to <12 yoIdenticalIdentical> 7 yo and up (MiniMe 2 aligns with SleepNet MiniMe)
    Environment of Use: Home or hospital/institutionalIdenticalIdenticalIdentical
    Durations of Use: Single use, disposable; Single patient, multi-use (home); Single patient, short-term (up to 7 days).Matched or represented as shorter duration which is acceptable.Single patient, multi-use (home); Single patient, short-term (up to 7 days).Single patient, multi-use (home); Single patient, short-term (up to 7 days).
    Prescriptive: YesYesYesYes
    Cleaning: Soap and waterSoap and waterSoap and waterSoap and water
    Exhalation Port: Incorporates exhalation port with multiple holesYesYesYes
    Non-vented elbow: Configuration used with circuit with integrated exhalation valveYesYesN/A
    Delivered Pressure Range:4-20 cm H2O3-20 cm H2O>4 cm H2O
    CO2 washout profile (ETCO2% at mask at various pressures): Should be comparable to predicates/not excessively high. Testing done per ISO 17510-2 (largest size tested, worst case).4 cm H2O: 5.3 (6%) 5 cm H2O: 5.2 (4%) 10 cm H2O: 5.1 (3%) Occluded: 5.3 (7%)4 cm H2O: 5.3 (6%) 5 cm H2O: 5.2 (4%) 10 cm H2O: 5.1 (3%) Occluded: 5.3 (7%)4 cm H2O: 5.8 (14%) 5 cm H2O: 5.7 (12%) 10 cm H2O: 5.5 (7%) Occluded: 6.8 (34%)
    Dead Space: Small volume.Small 25 ml (Small size) 40 ml (Large size)Small 21 ml (Small size)Small 81 ml (Large size)
    Exhaust - pressure / flow: Comparable flow rates at given pressures.4 cmH2O: 17.1 lpm 12 cmH2O: 33.5 lpm 20 cmH2O: 43.6 lpm4 cmH2O: 19.1 lpm 12 cmH2O: 34.8 lpm 20 cmH2O: 45.8 lpm4 cmH2O: 19.5 lpm 12 cmH2O: 36.2 lpm 20 cmH2O: 48.7 lpm
    Resistance to Flow: Acceptable resistance to flow values at various flow rates.30 lpm: 0.22 cm H2O 50 lpm: 0.72 cm H2O 60 lpm: 1.03 cm H2O 100 lpm: 3.09 cm H2O30 lpm: 0.10 cm H2O 50 lpm: 0.36 cm H2O 60 lpm: 0.54 cm H2O 100 lpm: 1.71 cm H2O30 lpm: 0.12 cm H2O 50 lpm: 0.44 cm H2O 60 lpm: 0.67 cm H2O 100 lpm: 2.10 cm H2O
    Biocompatibility: Materials in patient contact are identical to predicate or tested per ISO 10993.Materials are either identical to predicate SleepNet masks or tested per ISO 10993 for specific contact types (e.g., skin and gas pathway, permanent duration) and passed respective ISO 10993 tests (e.g., Cytotoxicity, Sensitization, Genotoxicity, Implantation, Intracutaneous, Systemic Toxicity for tubing).Defined by the predicate devices.Not explicitly detailed for this predicate.
    Mechanical Testing: Drop TestPerformed, but results not detailed in the summary.Not detailed.Not detailed.
    Cleaning Validation:Performed, but results not detailed in the summary.Not detailed.Not detailed.
    Environmental Testing:Performed, but results not detailed in the summary.Not detailed.Not detailed.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document does not specify a "test set" in terms of subject count. The testing performed is laboratory-based comparative performance testing (e.g., pressure-flow, CO2 washout, dead space, resistance to flow) comparing the device's physical properties to those of predicate devices.
    • Data Provenance: The data is generated from laboratory testing of the physical device, not from human subject data or retrospective/prospective clinical studies. The origin is stated implicitly as SleepNet Corporation facilities (Hampton, NH, USA).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is a physical device performance and biocompatibility study, not one requiring expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No adjudication by experts is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not applicable in the context of expert ground truth. The "ground truth" for the device's performance is established by standardized engineering tests and measurements (e.g., flow meters, pressure sensors for exhaust/resistance, CO2 sensors for washout) and adherence to material standards (ISO 10993). In terms of regulatory "ground truth" for substantial equivalence, it's defined by the performance characteristics of the legally marketed predicate devices.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1