Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K093919

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-01-21

    (30 days)

    Product Code
    Regulation Number
    882.5950
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MicroPlex Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.

    Device Description

    The Cosmos consist of an implant coil made of platinum alloy. The coils are designed in 3D spherical structure in various loop sizes and lengths. The coil is attached to a V-Trak™ MCS delivery pusher via a polymer filament. The delivery pusher contains radiopaque positioning markers at the distal end. The proximal end is inserted into a hand held battery powered V-Grip™ Detachment Controller. The implant segment detaches upon activation of the Detachment Controller.

    AI/ML Overview

    The MicroPlex Coil System - Cosmos (K093919) is a neurovascular embolization device. The provided document is a 510(k) summary, which indicates that the device's acceptance criteria and performance are typically evaluated through substantial equivalence to a predicate device, rather than a clinical study with detailed statistical endpoints.

    Based on the provided information, here's a breakdown of the acceptance criteria and the "study" (bench testing) that demonstrates the device meets these criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the MicroPlex Coil System - Cosmos are defined by the predicate device's established performance and design specifications. The "reported device performance" in this context refers to the results of bench tests conducted to demonstrate that the new device meets these pre-established criteria. The document states that all listed tests "Met established criteria," implying satisfactory performance against predefined internal specifications or industry standards.

    TestAcceptance Criteria (Implied)Reported Device Performance
    Visual InspectionConformance to visual design specificationsMet established criteria
    Dimensional MeasurementConformance to specified dimensionsMet established criteria
    Simulated UseFunctionality and performance in simulated use conditionsMet established criteria
    Detachment TestReliable and controlled detachment of the coilMet established criteria
    Detachment Zone TensileSufficient tensile strength at the detachment zoneMet established criteria
    Advancement/Retraction ForceAcceptable forces for device manipulationMet established criteria
    Coil to Coupler Weld TensileSufficient strength of the weldMet established criteria
    Spring ConstantConformance to specified mechanical propertiesMet established criteria

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states that the submission is a "510(k) SUMMARY OF SAFETY AND EFFECTIVENESS" and that the "Cosmos Coils are substantially equivalent to the predicate device with regard to intended use, patient population, device design, materials, processes, and operating principal."

    This indicates that the primary "study" proving the device meets acceptance criteria is bench testing to demonstrate performance similar to the predicate device. The document does not specify the sample size used for each of the listed bench tests.

    Regarding data provenance:

    • Country of Origin: The submitting company, MicroVention, Inc., is based in Tustin, California, U.S.A. Therefore, the bench testing was likely conducted in the USA.
    • Retrospective or Prospective: Bench testing, by its nature, is a prospective evaluation where tests are designed and performed to assess the device's characteristics. It is not based on historical patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since this is a 510(k) for a device demonstrated through substantial equivalence and bench testing, there is no mention of experts involved in establishing ground truth for a test set in the clinical sense. Ground truth for bench testing would typically be engineering specifications, established scientific principles, or comparison to the predicate device's known performance. These are not typically established by clinical experts in the same way as, for example, a radiologist establishing ground truth for an imaging AI algorithm.

    4. Adjudication Method for the Test Set

    As there is no clinical test set or human expert review described for establishing ground truth, there is no adjudication method mentioned. The "acceptance criteria" are likely engineering specifications or performance metrics derived from the predicate device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted or mentioned in this 510(k) summary. This type of study is common for evaluating the effectiveness of AI-assisted diagnostic tools, which is not the nature of this device (a physical neurovascular embolization coil).

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This device is a physical medical implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the acceptance criteria for this device is based on:

    • Engineering Specifications: Pre-defined technical requirements for the device's visual, dimensional, and mechanical properties.
    • Performance of the Predicate Device: The MicroPlex Coil System (MCS) - Cosmos (K082461 and K090891) serves as the benchmark against which the new device (Cosmos) is compared for substantial equivalence. The predicate device's established safety and effectiveness dictate the expected performance of the new device.

    There is no mention of pathology, outcomes data, or expert consensus in a clinical setting establishing "ground truth" for these bench tests.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical implant, not a machine learning algorithm. Therefore, there is no "training set" in the context of AI or data-driven model development.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As stated above, there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1