510(k) Data Aggregation

    K Number
    K172181

    Validate with FDA (Live)

    Device Name
    micro IPL
    Date Cleared
    2017-10-13

    (86 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The micro IPL is an over the counter device intended for the removal of unwanted body and/ or facial hair in adults. The micro IPL is also intended for permanent reduction in unwanted hair reduction is defined as a long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of treatment regimen.

    Device Description

    The micro IPL device is pulsed light hair removal device. Light-based hair removal is based on the theory of selective photothermolysis in which optical energy is used to disable hair growth. The micro IPL device is composed of a hand held applicator and an external power supply. The spot size (treatment area) in the micro IPL device is 3 cm . The device contains a lamp, a skin proximity sensor and a skin pigmentation sensor to detect appropriate skin tones. If the micro IPL is not properly applied (in full contact with the skin) or user skin tone is too dark/tanned. the micro IPL will not trigger a pulse.

    AI/ML Overview

    The provided document is a 510(k) summary for the micro IPL device, an over-the-counter Intense Pulsed Light (IPL) device intended for hair removal and permanent hair reduction.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria & Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative manner (e.g., "device must achieve X% hair reduction"). Instead, it focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and safety. The performance claims are qualitative, primarily related to meeting recognized standards and having similar specifications to already cleared devices.

    Acceptance Criteria (Implied)Reported Device Performance
    Technological Equivalence to Predicate Devices:
    - Wavelength of emitted light (spectrum) is similar.Micro IPL: 475 - 1200nm. Predicate (sensiLight mini): 475 - 1200nm. Predicate (Duo): 480 - 1200nm.
    - Fluence/flux (energy per area) is similar.Micro IPL: Maximum 5 joules/cm². Predicate devices: Maximum 5 joules/cm².
    - Pulse duration is similar.Micro IPL: Same as sensiLight mini predicate, similar to Duo predicate.
    Safety Features are Equivalent:Micro IPL includes skin color sensor, skin proximity sensor, etc., which are stated to be substantially equivalent to those in predicate devices. If not properly applied or skin tone too dark, the device will not trigger a pulse.
    Compliance with Recognized Standards:Device was tested and complies with: IEC 60601-1 (general safety), IEC 60601-1-11 (home healthcare), IEC 60601-2-57 (non-laser light source safety), FCC Part 15B (EMC), IEC 60601-1-2 (EMC), IEC 62471 (photobiological safety).
    Software Validation:Software Validation was conducted according to IEC 62304 and FDA Guidance.
    Biocompatibility:Body contact materials were evaluated for biocompatibility with FDA Memorandum G95-1 and ISO 10993-1:2009.
    Risk Management:Complies with ISO/IEC 14971:2007 (risk management).
    Labeling Claims Supported:"Tests results are supporting all labeling claims."
    Overall substantial equivalence (safety and effectiveness):"The evaluation of our device performances and comparison to predicate devices demonstrate that it is as safe and as effective as the predicate devices."

    2. Sample Size for the Test Set and Data Provenance

    The document explicitly states: "No new clinical performance data is reported in this submission." This indicates that no new clinical test set with human subjects was used for this submission to directly prove the device meets acceptance criteria. The data presented is primarily "bench" (laboratory) testing data and comparison to predicate devices. Therefore, there is no sample size for a clinical test set or data provenance in terms of country of origin or retrospective/prospective nature for this submission.

    3. Number of Experts and Qualifications for Ground Truth

    Since no new clinical performance data was provided, there was no "ground truth" established by experts for a clinical test set in this submission. The submission relies on the established safety and efficacy of the predicate devices and bench testing to demonstrate equivalence.

    4. Adjudication Method

    Not applicable, as no new clinical test set requiring expert adjudication was conducted or reported in this submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done or reported. This device is for direct consumer use for hair removal, not an imaging diagnostic tool that would typically involve multiple readers.

    6. Standalone (Algorithm Only) Performance Study

    No standalone performance study of an algorithm without human-in-the-loop performance was done. The device description focuses on its hardware components and safety features for user interaction (skin proximity/pigmentation sensors), not on an autonomous algorithm's diagnostic or predictive performance.

    7. Type of Ground Truth Used

    As stated under point 2, no new clinical ground truth (e.g., expert consensus, pathology, or outcomes data) was generated for this specific submission. The "ground truth" for the device's safety and effectiveness relies on the demonstrated substantial equivalence to already cleared predicate devices, whose efficacy and safety would have been established through prior studies. The compliance with various international standards serves as a form of "ground truth" for certain performance aspects (e.g., electrical safety, EMC).

    8. Sample Size for the Training Set

    No training set (for a machine learning algorithm) or clinical training data for human subjects is mentioned or implied, as no new clinical studies or AI/ML algorithm development are detailed.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set or associated ground truth establishment is described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113096

    Validate with FDA (Live)

    Device Name
    MICRO I
    Date Cleared
    2012-04-13

    (177 days)

    Product Code
    Regulation Number
    868.1840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Micro I spirometer is intended to measure the maximal volume and flow of air that can be moved out of a patient's lungs. The system is intended for use with pediatic and addlt patients over the age of 3 years in hospitals, physician offices, laboratories and occupational health testing environments.

    Device Description

    The Micro I spirometer is a hand held portable microprocessor based device designed to measure expiratory flows and volumes. To perform a spirometry test the user first inserts a mouthpiece into the mouthpiece holder of the spirometer, which aligns it with the volume transducer. The unit is then turned on and displays instructions and prompts according to the device type. When prompted to do so by the operator, the patient inhales as deeply as possible, seals his/her lips around the mouthpiece and exhales as hard and as fast as possible until no more air can be exhaled. The device converts the airflow to an electrical signal, and the onboard software calculates the required values.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the Micro I spirometer, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/ParameterAcceptance Criteria (Predicate: MicroLab K031102 & Micro Diary K100928/ATS/ERS)Reported Device Performance (Micro I K113096)
    Indications for UseMeasure maximal volume and flow of air for pulmonary function testing. (Microlab: prescription use, pediatric (4-17), adult (18-99). Micro Diary: measures FEV1, FVC, FEV6, PEF).Measures maximal volume and flow of air moved out of patient's lungs. For pediatric and adult patients over 3 years in hospitals, physician offices, laboratories, and occupational health.
    Target PopulationPediatric (4-17 years), Adult (18-99 years) for Microlab. 4 years and older for Micro Diary.Over the age of 3 years.
    Performance Specifications (Measurements)MicroLab: VC, FEV1, FEV3, FEV6, FVC, PEF, FEV1/FVC, etc. Micro Diary: FEV1, FVC, FEV6, PEF.FEV1, FVC, FEV6, FEV1/FVC, FEV1/FEV6, PEF, FEF25, FEF75, FEF25-75.
    Transducer Material (Breathing Path Contacting)MicroLab/Micro Diary: Turbine swirl plate (Polycarbonate/Makrolon 2607), Turbine van (Polyester type 427), Turbine tube (Clear Acrylic/Makrolon 2607), Pivot (Nivapoint stainless steel).Identical to K100928 (Micro Diary).
    Transducer TypeMicroLab: Type 36-TDX 1048 CareFusion Bi-Directional Digital Volume. Micro Diary: Type 36-TDX 1050 CareFusion Uni-Directional Digital Volume.Type 36-TDX 1051 CareFusion Uni-Directional Digital Volume.
    Resolution0.01 liter.Identical to K100928 & K031102.
    Accuracy (FEV1/FVC/PEF)+/- 3% to ATS Recommendations (MicroLab). To the requirements of the ATS/ERS Taskforce: Standardization of Spirometry 2005 (Micro Diary).Identical to K100928 (i.e., conforms to ATS/ERS 2005).
    Volume Range0.1 - 8 Litres (MicroLab). 0 - 8 l as per ATS/ERS (Micro Diary).Identical to K100928 (0 - 8 l as per ATS/ERS).
    Flow Range0.2 - 15 Litres/Second (MicroLab & Micro Diary).0 - 14 l as per ATS/ERS.
    SterilizationTurbine can be sterilized or cleaned.Identical to K100928 & K031102.
    Environmental SpecificationsOperating: 0 to +40 °C, 30% to 90% RH (MicroLab & Micro Diary).Operating: 10 to +35 °C, 20% to 80% RH.

    Summary of Device Testing (Evidence of Meeting Acceptance Criteria):

    The document states that the following practices were followed and monitored for the development of the Micro I Spirometer, forming the basis of its substantial equivalence claim:

    • Design and Development QP 0301: Device was developed in accordance with CareFusion's internal design and development quality procedure.
    • IEC 62366 (Usability): Modifications were developed according to this usability standard.
    • ISO 14971 (Risk Analysis): A Failure Modes and Effects Analysis (FMEA) was used to assess the impact of modifications.
    • Biocompatibility (ISO 10993-1): Tested for patient-contacting materials at Nelson test laboratory in the USA.
    • Safety and EMC Testing (IEC 60601-1 and IEC 60601-2): Device underwent electrical safety and electromagnetic compatibility testing.
    • Packaging Test: Performed.
    • Environmental Testing (Climatic Chamber): Performed to evaluate performance under various environmental conditions.
    • Testing according to ATS/ERS standardization of lung function testing: This is the most direct evidence for the spirometry performance characteristics (accuracy, volume/flow ranges, etc.). The device's accuracy and volume/flow ranges are stated to be "Identical to K100928" or "as per ATS/ERS," directly referring to these standards as the acceptance criteria.

    2. Sample Size for Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set. It mentions "Testing according ATS/ERS standardization of lung function testing," which implies that the device was validated against the performance requirements outlined in these standards. However, the exact number of test subjects or a specific test dataset is not provided.

    The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective. The testing was carried out by CareFusion internally and via a third-party lab (Nelson Test Laboratory for biocompatibility, located in the USA).

    3. Number of Experts and Qualifications for Ground Truth

    The document does not mention the use of experts to establish ground truth for a test set in the traditional sense of clinical studies with expert reviewers. The ground truth for spirometer performance is typically established by comparing the device's measurements against established, precisely calibrated reference systems or standards (like those from ATS/ERS).

    4. Adjudication Method

    Not applicable. Since the 'ground truth' for this type of device is based on technical standards and calibrated measurements rather than expert consensus on diagnostic images or clinical outcomes, an adjudication method like 2+1 or 3+1 is not relevant here.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No. This type of study (MRMC) is typically performed for AI-powered diagnostic devices where human readers are interpreting data (like medical images) with and without AI assistance to measure improvement in diagnostic performance. The Micro I is a measurement device (spirometer), not an AI-assisted diagnostic tool.

    6. Standalone (Algorithm Only) Performance Study

    Yes, implicitly. The "Testing according ATS/ERS standardization of lung function testing" would be a standalone performance evaluation of the device's measurement algorithms and hardware against established standards, without human interpretation as part of the primary measurement. The spirometer outputs numerical values (e.g., FEV1, FVC), and its accuracy is assessed against a gold standard for these measurements.

    7. Type of Ground Truth Used

    The ground truth used for performance evaluation is technical standards and calibrated measurements, specifically those established by the American Thoracic Society (ATS) / European Respiratory Society (ERS) standardization of lung function testing (2005). These standards define the acceptable accuracy, volume range, and flow range for spirometers.

    8. Sample Size for the Training Set

    Not applicable. The Micro I spirometer is described as a microprocessor-based device converting airflow to an electrical signal, and
    calculating values using "onboard software." It does not appear to be an AI/machine learning-based device that would typically have a "training set" in the sense of supervised learning. Its algorithms are likely deterministic and based on physical principles of flow and volume measurement.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no mention or indication of a training set in the context of machine learning. The device's operation is based on established scientific principles and engineering design, with performance validated against international standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K973978

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-07-29

    (282 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is restricted to sale or use by or on the order of a physician licensed in the state in which he or she is practicing. It is to be used solely for the symptomatic relief of chronic intractable pain.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for a medical device called "MICRO-II MICROCURRENT TENS." This type of document is a regulatory approval, not a scientific study report. It states that the device is "substantially equivalent" to legally marketed predicate devices.

    Therefore, this document does not contain the information requested regarding:

    • Acceptance criteria and reported device performance.
    • Details of a study proving the device meets acceptance criteria.
    • Sample sizes, data provenance, number/qualifications of experts, adjudication methods.
    • Multi-reader multi-case (MRMC) comparative effectiveness study results.
    • Standalone algorithm performance.
    • Type of ground truth used or how it was established for training or testing.
    • Training set sample size.

    The document's purpose is to grant market clearance based on substantial equivalence, implying that the device performs similarly to existing, approved devices, rather than presenting novel performance data from a new study against explicit acceptance criteria.