510(k) Data Aggregation

    K Number
    K110991

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-08-02

    (116 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Range Spinal System is comprised of the DENALI, DENALI DEFORMITY, and MESA Spinal Systems and the ARI Anterior Vertebral Body Staples, all of which are cleared for the following indications:

    Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

    Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to fusion for the following indications: degenerative disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The Mesa Spinal System is a top-loading, multiple component, spinal fixation system. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

    Materials: The devices are manufactured from titanium and cobalt chrome per ASTM and ISO standards.

    Function: The system functions as an adjunct to provide immobilization and stabilization of spinal segments of the thoracic, lumbar and / or sacral spine.

    AI/ML Overview

    The provided text pertains to a 510(k) summary for the "Mesa Spinal System" and its substantial equivalence to predicate devices. It describes the device, its intended use, and a comparison to other legally marketed systems, primarily focusing on mechanical testing against ASTM F1717 standards.

    However, the document does not contain information about acceptance criteria, a specific study used to prove the device meets those criteria, or details regarding ground truth, expert consensus, or clinical performance data that would typically be found in a study demonstrating effectiveness for a diagnostic or AI-driven device.

    This document is for a spinal fixation system, which is a mechanical implant, not a diagnostic or AI device. Therefore, the questions posed (related to AI performance, expert ground truth, multi-reader studies, etc.) are not applicable to this type of medical device submission.

    The "study" referenced in the document is mechanical testing to ASTM F1717.

    Here's how to address the prompt based on the provided text, while acknowledging the limitations:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied from text)Reported Device Performance (from text)
    Performance equal to or better than predicate and currently marketed systems in ASTM F1717 testing."The Mesa Spinal System was mechanically tested and compared to the predicate systems and other currently marketed systems and performed equal to or better than these systems in ASTM testing to F1717."
    Substantial equivalence in design features, sizing, materials, and intended use as predicate devices."The design features and sizing of the components were also compared and the Range Spinal System was found to be substantially the same as these systems. It is manufactured from the same materials and is indicated for the same intended uses as these systems."
    No significant differences that would adversely affect use compared to currently marketed systems."There are no significant differences between the Range Spinal Systems currently being marketed which would adversely affect the use of the product."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: Not specified in terms of patient data. The "test set" here refers to the components of the Mesa Spinal System subjected to mechanical testing. The number of components tested for ASTM F1717 is not detailed.
    • Data Provenance: Not applicable in the context of patient data for a mechanical fixation device. The testing is performance-based for the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This document describes a mechanical device, not a diagnostic or AI system requiring expert-established ground truth from medical images or clinical outcomes. The "ground truth" for the mechanical testing is defined by the ASTM F1717 standard itself.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are typically used for clinical studies involving human interpretation or uncertain outcomes. This document details mechanical testing against a standard.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a mechanical spinal fixation system, not a diagnostic or AI device. An MRMC study is not relevant to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is a mechanical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" in this context is the ASTM F1717 standard for mechanical testing of spinal implant constructs. The device's performance is compared against the requirements and performance characteristics of legally marketed predicate devices as measured under this standard.

    8. The sample size for the training set

    • Not applicable. This is a mechanical device, not an AI system that requires a "training set."

    9. How the ground truth for the training set was established

    • Not applicable. As a mechanical device, a training set and its associated ground truth are not relevant concepts for this submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K080611

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-03-04

    (365 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Non-cervical spinal fixation devices intended for posterior, non-pedicle fixation for the following indications: degenerative disc disease ( DDD ) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

    Non-cervical pedicle screw fixation for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S 1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum ) with removal of the implants after the attainment of a solid fusion.

    Device Description

    The Mesa Spinal System is a top-loading, multiple component, posterior system which consists of pedicle screws, rods, hooks, and transverse connectors. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

    Materials: The devices are manufactured from Ti6Al-4V ELI alloy per ASTM and ISO standards.

    Function: The system functions as an adjunct to fusion to provide immobilization and stabilization of spinal segments of the thoracic, lumbar and / or sacral spine.

    AI/ML Overview

    This request is about a medical device, the Mesa Spinal System, specifically the 4.0mm Screw component, and not about an AI/ML powered device. Therefore, many of the typical acceptance criteria and study design elements expected for an AI/ML device (like sample size for test/training sets, ground truth methodology, expert adjudication, MRMC studies, standalone performance, etc.) are not applicable here.

    The provided text describes a 510(k) submission for the Mesa Spinal System, which is a spinal fixation device. For such devices, the "acceptance criteria" and "study" typically revolve around demonstrating substantial equivalence to a legally marketed predicate device through engineering and biocompatibility testing, material characterization, and comparison of design features and intended use.

    Here's an analysis based on the provided text, addressing the relevant points where possible:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Biomechanical Performance (ASTM F1717)"The Mesa Spinal System was biomechanically tested and compared to the predicate systems and other currently marketed systems and performed equal to or better than these systems in ASTM testing to F1717."
    Material Composition (ASTM & ISO Standards)"The devices are manufactured from Ti6Al-4V ELI alloy per ASTM and ISO standards." "It is manufactured from the same FDA recognized materials..."
    Design Features and Sizing"...the design features and sizing of the components were also compared and the Mesa System found to be substantially the same as these systems."
    Intended Use"is indicated for the same intended uses as these systems." (The specific indications are listed in the 'Intended Use' and 'Indications for Use' sections of the document).
    No Adverse Effects"There are no significant differences between the Mesa Spinal Systems currently being marketed which would adversely affect the use of the product."
    Substantial Equivalence"It is substantially equivalent to these other devices in design, function, material and intended use." (This is the overarching acceptance criterion for a 510(k) submission).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: This is not applicable in the context of this 510(k) submission. "Test set" typically refers to data used to evaluate an AI/ML model's performance on unseen data. For a spinal fixation device, performance is evaluated through material testing and biomechanical testing on representative samples of the device components, not a "test set" of patient data. The text indicates that the device was "biomechanically tested," which implies a certain number of physical samples were subjected to various tests, but the exact count is not provided.
    • Data Provenance: Not applicable in the AI/ML sense. The "data" here would be the results of laboratory biomechanical and material tests. The document does not specify the location where these tests were performed or if they were retrospective or prospective, as these terms are generally used for clinical studies or data collection for AI.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. "Ground truth" and "experts" in this context are not relevant for demonstrating substantial equivalence for a spinal fixation device. Performance is determined by objective engineering and material standards (e.g., ASTM F1717).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are typically for resolving discrepancies in expert labeling or diagnoses, which is not part of a spinal fixation device's 510(k) submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is for AI-assisted image interpretation or diagnostic tools. The Mesa Spinal System is a physical implant, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This concept pertains to AI algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable. "Ground truth" as an expert-derived label or outcome is not used here. Instead, the "truth" or standard for performance is defined by established engineering and material standards (e.g., ASTM F1717 for biomechanical testing) and comparison to the predicate device's characteristics.

    8. The sample size for the training set

    • Not applicable. There is no "training set" for physical medical devices in this context.

    9. How the ground truth for the training set was established

    • Not applicable. As there's no training set, there's no ground truth to establish for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K052398

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-12-07

    (98 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mesa Spinal System is a non-cervical spinal fixation device intended for posterior, non-pedicle fixation for the following indications: degenerative disc disease ( DDD ) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

    The Mesa Spinal System is also intended for non-cervical pedicle screw fixation for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe 'pondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by utogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum ) with removal of the implants after the attainment of a solid fusion.

    Device Description

    The Mesa Spine System is a top-loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods, hooks, and transverse connectors. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

    Materials: The devices are manufactured from Ti6Al-4V ELI alloy per ASTM and ISO standards.

    Function: The system functions as an adjunct to fusion to provide immobilization of spinal segments of the thoracic, lumbar and / or sacral spine.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Mesa Spinal System, which is a medical device. This type of document is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed device. It describes the device's intended use, materials, and a comparison to predicate devices, but it does not contain information about acceptance criteria or a study proving device performance against such criteria in the context of clinical or diagnostic efficacy.

    Instead, the document primarily focuses on biomechanical testing to demonstrate equivalence to predicate devices, which is typical for spinal fixation systems. Here's a breakdown of why the requested information cannot be fully provided from the given text:

    1. Acceptance criteria and reported device performance (Table 1): The document states: "The Mesa Spine System was biomechanically tested and compared to the predicate systems and other currently marketed systems and performed equal to or better than these systems in ASTM testing to F1717." This indicates a type of performance standard (ASTM F1717) was used, and the device met or exceeded it. However, specific numerical acceptance criteria (e.g., "must withstand X N of force," "fatigue life must exceed Y cycles") and the exact reported performance values are not detailed in this summary. The table below reflects the limited information available.

    2. Sample size, data provenance, number of experts, adjudication method, MRMC studies, standalone performance, type of ground truth, training set size, and ground truth establishment for a clinical/diagnostic study: These questions are entirely not applicable to the type of information presented in this 510(k) summary. The document describes a biomechanical study, not a clinical study involving human patients, image analysis, or diagnostic accuracy where such details would be relevant.

    Therefore, most of the requested fields cannot be filled based on the provided text.

    Here's the closest possible answer based only on the provided text, with explanations where information is absent:


    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Biomechanical Performance: Equal to or better than predicate devices in ASTM F1717 testing.Performed equal to or better than predicate systems and other currently marketed systems in ASTM testing to F1717.

    Explanation: The document explicitly states that the device was biomechanically tested against the ASTM F1717 standard and compared to predicate devices. The acceptance criterion was implicitly defined as performing "equal to or better than" these established systems. Specific numerical thresholds for this ASTM standard and the exact results are not provided in this 510(k) summary.


    Details of the Study

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample Size: Not specified. (This likely refers to the number of devices or components tested in the biomechanical study).
      • Data Provenance: Not applicable, as this was a biomechanical study, not a clinical data collection from patients or specific countries.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not Applicable. This was a biomechanical test of physical properties, not a study requiring expert clinical "ground truth" establishment.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not Applicable. (Again, this pertains to clinical/diagnostic studies, not biomechanical tests).
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. An MRMC study was not done. This device is a spinal fixation system, not a diagnostic AI tool, so this type of study is irrelevant.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not Applicable. This is a physical medical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Not Applicable. The "ground truth" for a biomechanical study would typically be the physical properties and performance standards defined by the ASTM F1717 standard and the performance of the predicate devices.
    7. The sample size for the training set:

      • Not Applicable. This device is hardware; there is no "training set" in the machine learning sense.
    8. How the ground truth for the training set was established:

      • Not Applicable. As there is no training set for this type of device and study.