    Intended Use

    The Spectris Solaris MR Injection System is a syringe-based fluid delivery system indicated for delivery of contrast media during MR applications. It is intended to be used for the specific purpose of injecting intravenous MR contrast media and common flushing solutions into the human vascular system for diagnostic studies in magnetic resonance imaging (MRI) procedures. Only trained healthcare professionals are intended to operate this device.

    Device Description

    The Medrad Spectris Solaris MR Injection System maintains the same intended use, similar operational parameters, similar labeling and is essentially used in a manner similar to the predicate device.

    A label will be affixed to the 3.0 Tesla-compatible injector system stating that the injector should be placed at a minimum of 4 ft. from the magnet bore when used with magnetic resonance imaging scanners having a field strength greater than 2.0 Tesla. The user's manual has also been updated to include this warning.

    The Spectris Solaris MR Injector (3T Compatible) is comprised of the same two main components as the predicate device - Injector Head/Stand (HSU) and Display Control Unit (DCU).

    AI/ML Overview

    The provided text is a 510(k) Summary for the Medrad Spectris Solaris MR Injector System. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study to prove acceptance criteria for a novel device. The changes described are primarily related to increasing the system's compatibility with higher magnetic field strengths (from 2.0 T to 3.0 T).

    Therefore, the typical "acceptance criteria" and "study proving the device meets the acceptance criteria" as might be expected for a new AI/software-based medical device with performance metrics (like sensitivity, specificity, or accuracy) are not present in this document. Instead, the "acceptance criteria" are implied by showing the new device maintains similar operational parameters, similar labeling, and the same intended use as the predicate device, with the key difference being the expanded 3T compatibility.

    Here's an attempt to answer your questions based on the available information, noting that many fields will be "Not Applicable" or "Not Provided" given the nature of a 510(k) for a hardware modification:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Functional Equivalence:
    - Maintain existing flow rate range (0.01 - 10 ml/s)- Same (0.01 - 10 ml/s)
    - Maintain fractional volumes capability- Same (Yes)
    - Maintain remote information display (color touch screen)- Same (Color Touch screen)
    - Maintain multi-phase capability (6 phases)- Same (6)
    - Maintain protocol storage capability (32 protocols, 3 steps)- Same (Yes: 3 step procedure, 32 protocols)
    - Maintain hold capability- Same (Yes)
    - Maintain safety stop mechanism- Same (Yes)
    - Maintain syringe system (2 total: 1-115ml, 1-65ml)- Same (2 total: 1-115ml syringe and 1-65ml syringe)
    - Maintain air detection (user observable)- Same (User observable)
    - Maintain syringe sensing- Same (Yes)
    - Maintain volume remaining readout- Same (Graduation on syringe; LCD on Program Panel)
    - Maintain programmed volume (1-63ml and 1-115ml)- Same (Yes, 1 to 63 ml and 1-115 ml)
    - Maintain fill rate (up to 10 ml/sec)- Same (Configurable up to a maximum of 10 ml/sec)
    - Maintain variable flow rate- Same (Variable)
    - Maintain KVO (0.25 ml, time adjustable, every 30 sec default)- Same (Yes: 0.25 ml, time adjustable. Defaults at every 30 sec.)
    - Maintain pressure limit (325 PSI for injector)- Same (325 PSI)
    - Maintain start/stop switch functionality- Same (Yes, Located at injector head or at display unit.)
    - Maintain communication method (Fiber Optic)- Same (Fiber Optic)
    Increased System Compatibility:
    - Operate safely and effectively with 3.0 Tesla MR scanners- Achieved: System Compatibility expanded to 0.2 - 3 T (from 0.2 - 2 T)
    (Implied: Electromagnetic compatibility, minimal interference)(Labeling update for 4 ft. placement from magnet bore)
    Disposable System Equivalence:
    - Syringe system (2 total: 1-115ml, 1-65ml or 2-65ml)- Same (2 total, 1- 115 ml syringe and 1- 65 ml syringe, 2 - 65 ml syringes)
    - Intended use (saline/contrast delivery)- Same (Saline/Contrast delivery)
    - Barrel material/dimensions (PET, 4.5" length, 1.565" OD, 1.415 ID)- Same (PET, 4.5", 1.565", 1.415)
    - Plunger material (Polycarbonate)- Same (Polycarbonate)
    - Barrel flange (Quik-Fit Design)- Same (Quik-Fit Design)
    - Packaging/Sterilization (Polystyrene tray with Tyvek Lid, EtO)- Same (Polystyrene tray with Tyvek Lid, Ethylene Oxide)
    - T-connector material/tubing (Polycarbonate, PVC)- Same (Polycarbonate, PVC)
    - Luer Fitting (Polycarbonate)- Same (Polycarbonate)
    - Tubing length/ID (96", 0.075")- Same (96", 0.075")
    - Maximum Pressure (350 PSI for T-connector)- Same (350 PSI)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Provided. This type of detail is not typically included in a 510(k) summary for a hardware modification of this nature. The "study" here is primarily an engineering assessment of compatibility and equivalence to the predicate device.
    • The comparison matrix explicitly states "Same" for most features, indicating that those parts of the system are unchanged and therefore assumed to perform identically to the previously cleared predicate. For the 3T compatibility, specific testing (e.g., electromagnetic compatibility, performance under higher magnetic fields) would have been performed, but the details (like sample size of tests or data provenance) are not summarized here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable/Not Provided. "Ground truth" in the context of diagnostic accuracy is not relevant here. This is a functional/safety equivalence submission for an injector system. The "experts" involved would be Medrad's own engineers and potentially external testing labs for aspects like EMC, but their qualifications and numbers are not part of the public 510(k) summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable/Not Provided. Adjudication methods are typically for evaluating subjective interpretations, often in diagnostic imaging. This device is an injector, so such methods would not apply.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI or diagnostic imaging device. It's a contrast media injector.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. Not an algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. For this device, "ground truth" would relate to engineering specifications and performance metrics of the injector system (e.g., accuracy of flow rate, pressure limits, structural integrity, electromagnetic compatibility). These are objective measurements against established engineering standards and specifications, not subjective medical diagnoses.

    8. The sample size for the training set

    • Not Applicable. This device does not use a "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    • Not Applicable. This device does not use a "training set."