Found 2 results

510(k) Data Aggregation

    K Number
    K140177

    Validate with FDA (Live)

    Date Cleared
    2014-02-18

    (25 days)

    Product Code
    Regulation Number
    870.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medos hilite 2800 & 2400 LT Hollow Fiber Oxygenators are indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from human blood. It is designed to operate at a blood flow rate of 0.5 to 2.8 liters per minute for the hilite 2800 and 0.5 to 2.4 liters per minute for the hilite 2400 LT for periods of up to six (6.0) hours.

    Device Description

    The Medos Hilite 2800 & 2400 LT Oxygenators consist of a hollow fiber membrane oxygenator and extracorporeal heat exchanger. The Medos Hilite 2800 hollow fiber membrane consists of a polypropylene gas permeable mat. The Medos Hilite 2400 LT hollow fiber membrane consists of a polymethylpentene plasma tight mat. The unique mat design increases the interaction between blood and gas, creating a highly efficient blood oxygenator. The heat exchanger consists of a polyester non-porous hollow fiber configured heat exchanger as the primary element to affect heat exchange. This element is encased by a polycarbonate housing, which directs the blood around the outside of the fibers while water flows through the inner lumen of the fibers and therefore effects heat exchange while minimizing priming volume.

    The device allows for trapping and removal of air. Oxygenated blood is delivered to the patient through the tubing and appropriate cannula. Blood flow is driven by a roller pump or centrifugal pump connected through the tubing. The Medos Hilite 2800 and 2400 LT Oxygenators may be purchased separately or pre-connected with tubing and other components of an extracorporeal circuit.

    AI/ML Overview

    The provided text is a 510(k) summary for the Medos Hilite 2800 & 2400 LT Oxygenators. It describes the device, its intended use, and argues for its substantial equivalence to a predicate device. However, it does not contain the kind of detailed information about acceptance criteria and a specific study that proves the device meets those criteria, which would be typically found for an AI/ML-based medical device.

    Therefore, it is not possible to fill in all the requested information as the document is for a traditional medical device (blood oxygenator) and not an AI/ML diagnostic tool.

    Here's an attempt to address the points based on the available information, noting where an AI/ML context does not apply:


    Acceptance Criteria and Device Performance:

    The document describes "Performance Testing" which is generally for demonstrating functional requirements and safety. It doesn't present a table of specific numerical acceptance criteria with corresponding reported device performance values in the way you'd expect for, say, sensitivity or specificity of a diagnostic algorithm.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryReported Device PerformanceComments
    Material & Integrity Tests (Due to Change in Potting Material)The document lists tests performed but does not provide specific numerical criteria or results for each. These are presented as tests performed to ensure compliance after a material change.
    Blood Pathway IntegrityPerformedOutcome implied to be satisfactory for substantial equivalence.
    Heat Exchanger Fluid Pathway IntegrityPerformedOutcome implied to be satisfactory for substantial equivalence.
    Chemical AnalysisPerformed on specific materials (Macroplast CR 3502/CR 4100, Hilite 7000, Hilite 7000 LT)Outcome implied to be satisfactory for substantial equivalence.
    Biological Safety - Toxicology
    Evaluation of Biological Safety - toxicologyPerformed on Macroplast CR 3502/CR 4100Outcome implied to be satisfactory for substantial equivalence.
    Cytotoxicity (L 929-Proliferation)Performed on Macroplast CR 3502/CR 4100, EO sterilizedOutcome implied to be satisfactory for substantial equivalence.
    Hemolysis (elution method)Performed on Macroplast CR 3502/CR 4100, EO sterilizedOutcome implied to be satisfactory for substantial equivalence.
    CytotoxicityPerformedOutcome implied to be satisfactory for substantial equivalence.
    SensitizationPerformedOutcome implied to be satisfactory for substantial equivalence.
    Intracutaneous ReactivityPerformedOutcome implied to be satisfactory for substantial equivalence.
    Acute Systemic ToxicityPerformedOutcome implied to be satisfactory for substantial equivalence.
    PyrogenPerformedOutcome implied to be satisfactory for substantial equivalence.
    HemocompatibilityPerformedOutcome implied to be satisfactory for substantial equivalence.

    Study Details (as they pertain to a traditional medical device submission for substantial equivalence):

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not applicable in the context of an AI/ML test set. The "tests" listed are bench performance and biocompatibility tests on the device materials and components, not a diagnostic test on a data set. No human or animal subject sample size is specified for these tests, nor is there a "test set" in the AI/ML sense.
    • Data Provenance: Not applicable. The document focuses on demonstrating physical and biological equivalence, not data analysis from specific patient populations.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This is not a diagnostic device relying on expert-labeled ground truth for algorithm performance validation. The "ground truth" for the performance tests would be established by validated test methodologies and industry standards for medical device materials and function.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This is not a diagnostic device requiring adjudication of human-labeled classifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an algorithm-based device. The "performance" described is the functional and biological safety performance of the physical oxygenator device itself.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not applicable for AI/ML ground truth. For the physical device, the "ground truth" is implied by adherence to established engineering specifications, material standards, and biocompatibility guidelines. The testing confirms the device's physical and chemical properties and interaction with blood.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K082403

    Validate with FDA (Live)

    Date Cleared
    2009-05-15

    (267 days)

    Product Code
    Regulation Number
    870.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MEDOS HILITE 7000 & 7000 LT Hollow Fiber Oxygenator is indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from human blood. It is designed to operate at a blood flow rate of one (1.0) to seven (7.0) liters per minute for periods of up to six (6.0) hours.

    Device Description

    The MEDOS HILITE 7000 & 7000 LT Oxygenator consists of a hollow fiber membrane oxygenator and extracorporeal heat exchanger. The MEDOS HILITE 7000 hollow fiber membrane consists of a polypropylene gas permeable mat. The MEDOS HILITE 7000 LT hollow fiber membrane consists of a Polymethylpentene plasma tight mat. The unique mat design increases the interaction between blood and gas, creating a highly efficient blood oxygenator. The heat exchanger consists of a polyester non-porous hollow fiber configured heat exchanger as the primary element to effect heat exchange. This element is encased by a polycarbonate housing, which directs the blood around the outside of the fibers while water flows through the inner lumen of the fibers and therefore effects heat exchange while minimizing priming volume. All materials of the heat exchanger are biocompatible and coated with a proprietary coating. The device allows for trapping and removal of air. Oxygenated blood is delivered to the patient through the tubing and appropriate cannula. Blood flow is driven by a roller pump or centrifugal pump connected through the tubing. The MEDOS HILITE 7000 Oxygenator may be purchased separately or pre-connected with tubing and other components of an extracorporeal circuit.

    AI/ML Overview

    The provided text describes the MEDOS HILITE 7000 & 7000 LT Hollow Fiber Oxygenator. However, it does not contain information about acceptance criteria, a study proving the device meets those criteria, or any of the specific details requested regarding sample sizes, ground truth, experts, or AI-related aspects.

    The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting extensive performance studies with specific acceptance criteria and detailed quantitative results.

    Here's a breakdown of what can be extracted and what is missing:

    The device is a blood oxygenator indicated for extracorporeal oxygenation and carbon dioxide removal from human blood at blood flow rates of 1.0 to 7.0 liters per minute for up to 6.0 hours.

    Missing Information:

    • Acceptance Criteria Table: The document does not specify any quantitative acceptance criteria or report specific device performance metrics against such criteria (e.g., oxygen transfer rate, CO2 removal rate, pressure drop, hemolysis rates, etc., with specific thresholds).
    • Sample Size for Test Set and Data Provenance: This information is not provided. The summary mentions "extensive safety, performance, and validations" and "final testing for the systems," but no details on sample sizes or the nature (retrospective/prospective, country of origin) of the data from these tests are given.
    • Number of Experts and Qualifications: Not applicable, as no expert-derived ground truth or review process is detailed concerning a test set.
    • Adjudication Method: Not applicable for the same reason as above.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: This is entirely irrelevant to this type of medical device (a physical blood oxygenator) and is not mentioned. There is no AI component involved.
    • Standalone Performance (Algorithm Only): Not applicable, as this is a physical medical device, not an algorithm.
    • Type of Ground Truth Used: Not explicitly stated in terms of a "ground truth" as might be used in an AI or diagnostic imaging study. The "test data" likely involved laboratory measurements and possibly animal or human studies to compare to the predicate device's performance.
    • Sample Size for Training Set: Not applicable or mentioned, as there is no AI algorithm being trained.
    • How Ground Truth for Training Set was Established: Not applicable.

    What the document does state regarding testing:

    • "The MEDOS HILITE 7000 & 7000 LT Oxygenator has been subjected to extensive safety, performance, and validations prior to release."
    • "Final testing for the systems includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications."
    • "Safety tests have further been performed to ensure the device complies with applicable industry and safety standards."
    • "A review of literature pertaining to the safety and effectiveness has been conducted."
    • The conclusion is based on "testing and comparison to the predicate device," asserting equivalence in safety and effectiveness.

    In summary, this document is a regulatory submission for a physical medical device (blood oxygenator) focusing on substantial equivalence to a predicate device, not on detailed performance studies with specific statistical acceptance criteria as would be found for a diagnostic or AI-driven device. Therefore, most of the requested specific information is not contained within this text.