Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K082493

    Validate with FDA (Live)

    Date Cleared
    2008-10-23

    (55 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MD205MG-1 grayscale display is intended to be used for displaying and viewing of digital image diagnosis including digital mammography by trained physicians.

    To guarantee the display performance as specified, it must only be used for in Conjunction with NEC approved display cards.

    The MD205MG-1 cannot be used for a life-support system.

    This unit is designed for exclusive interconnection with IEC60601-1 certified equipment.

    Device Description

    Medical Display, MD205MG-1 is a 20.1" monochrome LCD monitor that displays image for medical use. It provides 5 mega pixel (2560*2048) resolution with adjustable gamma gray scale for more precise diagnose use in CT, MRI, HIS, and PACS.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical display device, the NEC Medical Display, MD205MG-1. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study with acceptance criteria and device performance metrics in the way a clinical trial or AI algorithm validation study would.

    Therefore, much of the requested information (acceptance criteria table, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth details) is not applicable or not present in this type of regulatory submission.

    Here's what can be extracted based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    This document describes a device (a medical display monitor) and its intended use, rather than a diagnostic algorithm or a device with specific performance metrics like sensitivity or specificity. The "acceptance criteria" here refer to conformity with established standards for medical devices and displays, and "device performance" refers to its technical specifications matching the predicate device or meeting those standards.

    Acceptance Criteria / StandardReported Device Conformance / Specification
    Intended Use"displaying and viewing of digital image diagnosis including digital mammography system by trained physicians."
    Resolution5 mega pixel (2560*2048)
    Gamma grayscaleAdjustable
    Standards ComplianceUL60601-1, EN60601-1-2, PAS-1054, DDN6868-57, ISO13406-2
    CompatibilityMust be used with NEC approved display cards
    SafetyDoes not contact with the patient, does not control any life-sustaining device
    InterconnectionDesigned for exclusive interconnection with IEC60601-1 certified equipment
    Equivalence to PredicateSubstantially equivalent to MDM2010-5NC (K063269) for intended use, despite differences in panel, luminance, and scanning frequency.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not applicable. This is a hardware conformity assessment, not an algorithmic performance study on a test dataset of images/patients.
    • Data Provenance: Not applicable for an "algorithm test set." The device itself (the monitor) underwent testing against compliance standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. There is no "ground truth" for a test set of medical cases/images in this context, as it's a display monitor. The compliance standards themselves form the basis of evaluation.

    4. Adjudication method for the test set

    • Not applicable. No adjudication method for medical cases is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a hardware device (a display monitor), not an AI algorithm. Therefore, no MRMC study, AI assistance, or effect size is relevant or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a hardware device.

    7. The type of ground truth used

    • Not applicable in the context of clinical "ground truth" from pathology, expert consensus, or outcomes data. The "ground truth" for this device's evaluation is its adherence to technical specifications, safety standards (UL60601-1, EN60601-1-2, etc.), and functional equivalence to the predicate device.

    8. The sample size for the training set

    • Not applicable. This is a hardware display device, not an AI algorithm.

    9. How the ground truth for the training set was established

    • Not applicable. This is a hardware display device, not an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1