Found 11 results

510(k) Data Aggregation

    K Number
    K252658

    Validate with FDA (Live)

    Date Cleared
    2025-11-28

    (98 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Diode Laser Systems are intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures. The device's 980nm laser is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology. The device is further indicated for laser assisted lipolysis. The device's 1470nm laser is intended for: Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue-Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux.

    Device Description

    The Medical Diode Laser Systems generate a 980nm wavelength laser to act on a target tissue to achieve resection, hemostasis, ablation, and coagulation of the target tissue.

    The Medical Diode Laser Systems generate a 1470 nm wavelength laser that acts on the water molecules of the target tissue to achieve the function of treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

    The Medical Laser Family has following characteristics:

    • Dual wavelengths
    • Temperature monitoring system
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252564

    Validate with FDA (Live)

    Date Cleared
    2025-11-07

    (85 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The "VeinCure" is indicated for use in the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

    Device Description

    Diode laser is a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source.The diode laser for this unit is GaAlAs (Gallium-aluminum-arsenic) diode bar, and the wavelength is 1470nm. It features impact structure, high efficiency and long lifetime. Generally the beam shall be shamed as the big beam divergence of the laser from the diode. When the coaxiality of laser and fiber meet the requirements, the laser beam can be coupled efficiently into the fiber.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252063

    Validate with FDA (Live)

    Date Cleared
    2025-09-25

    (86 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K242755

    Validate with FDA (Live)

    Date Cleared
    2025-05-19

    (249 days)

    Product Code
    Regulation Number
    890.5500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Diode Laser Systems, include model: VELASⅡ-30A, GBOX-15AB, GBOX-20B, CHEESEⅡ-10B are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation.

    The Medical Diode Laser Systems, model: VELASⅡ-30B is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation.

    The Medical Diode Laser Systems, model: VELASⅡ-30B is also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).

    Device Description

    The "Medical Diode Laser Systems", include model: VELASⅡ-30B, VELASⅡ-30A, GBOX-15AB, GBOX-20B, CHEESEⅡ-10B which consist of three main components: Main device; Foot switch; Accessories.

    Diode laser is a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source.

    The diode laser for this unit is GaAlAs diode bar, and the wavelength is 810nm and/or 980nm+/-10nm (pilot beam: 650nm+/-10nm). It features impact structure, high efficiency and long lifetime. Generally, the beam shall be emitted as the big beam divergence of the laser from the diode. With the unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber.

    MCU (Micro Controller Unit) is the control center of system. MCU controls the drive board and temperature system by changing operating current. The operating power supply is the power source of the whole system. It supplies power for MCU, diode laser module by DC-DC conversion module and temp system by temp-control circuit. MCU controls the drive current of diode laser module by adjusting the DC-DC module and current-control circuit. The MCU has three external input sources: footswitch, remote interlock connector and information input from the touch units, it has one external output unit: display terminal. The diode laser module provides laser power output by fiber with optical fiber coupling system.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the Gigaalaser Medical Diode Laser Systems does not contain a study that proves the device meets specific acceptance criteria related to its clinical efficacy. Instead, it demonstrates substantial equivalence to legally marketed predicate devices based on technological characteristics and safety standards.

    The document focuses on non-clinical performance data (electrical safety, EMC, software V&V, mechanical/acoustic, animal, and clinical studies are noted as "not applicable" for this submission type) to argue that the proposed device is as safe and effective as its predicates.

    Therefore, I cannot directly provide a table of acceptance criteria and reported device performance in the context of a clinical efficacy study, as such a study is explicitly stated as "not applicable" in this 510(k) submission. Similarly, details about sample size, expert ground truth, adjudication methods, MRMC studies, or standalone performance are not present because clinical efficacy testing was not required for this type of submission.

    However, I can interpret the acceptance criteria in the context of a 510(k) submission as the requirements for demonstrating substantial equivalence to predicates, particularly regarding safety and technological characteristics. The "performance" reported is the demonstration that these criteria are met through compliance with recognized standards.

    Here's an interpretation based on the provided document:


    Overview of "Acceptance Criteria" for Substantial Equivalence and Device Performance (as demonstrated in the 510(k))

    Within the context of a 510(k) submission, "acceptance criteria" are primarily established through demonstrating compliance with recognized performance standards and substantial equivalence to legally marketed predicate devices. The "study" proving this involves a comprehensive technical comparison and non-clinical testing.

    The document explicitly states that clinical studies are "not applicable" for this 510(k) submission, meaning the acceptance criteria are focused on non-clinical performance and technological equivalence rather than empirical clinical efficacy data from a human trial.

    1. Table of "Acceptance Criteria" (Substantial Equivalence) and "Reported Device Performance" (Demonstrated Compliance)

    Acceptance Criteria (Demonstration of Equivalence to Predicate)Reported Device Performance (as summarized in the document)
    Similar Indications for UseIdentical or Covered: All proposed device models (VELASⅡ-30A, GBOX-15AB, GBOX-20B, CHEESEⅡ-10B) share identical indications for general topical heating relief as their respective predicates. The VELASⅡ-30B also shares the onychomycosis indication with a co-predicate.
    Similar Technological Characteristics (General)Identical Diode Laser Technology: All models use the same diode laser technology (e.g., 980nm, 810nm, or combinations) as their respective predicates.
    Similar Output Parameters (within clinical needs)Within Predicate Scope: While some proposed device models show slight differences in output power, energy density, power density, spot diameter, operation mode (CW vs. pulse), pulse width, pulse repetition rate, transmission system, and aiming beam compared to their specific predicates, these differences are argued to be "slight" and "do not raise new types of questions regarding the safety and effectiveness" and are "within the scope of what predicate device claims" or "meet clinical needs."
    Compliance with Electrical Safety StandardsVerified Conformity: Demonstrated compliance with IEC 60601-1:2005/AMD1:2012/AMD2:2020.
    Compliance with Electromagnetic Compatibility (EMC) StandardsVerified Conformity: Demonstrated compliance with IEC 60601-1-2:2014+A1:2020.
    Compliance with Laser Safety StandardsVerified Conformity: Demonstrated compliance with IEC 60825-1:2014.
    Compliance with Usability StandardsVerified Conformity: Demonstrated compliance with IEC 60601-1-6:2010/AMD1:2013/AMD2:2020.
    Compliance with Medical Laser Equipment Specific StandardsVerified Performance: Demonstrated compliance with IEC 60601-2-22:2019.
    Software Verification & ValidationConducted and Documented: Software testing conducted as per FDA guidance, with the software deemed a "moderate" level of concern.

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not describe a "test set" in the context of clinical data for efficacy. The "testing" referred to is primarily bench testing and compliance validation with international safety and performance standards. Therefore, concepts like sample size for a patient test set and data provenance (country of origin, retrospective/prospective) are not applicable as no clinical study for efficacy was performed or presented.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. As no clinical efficacy test set was used, there was no need for experts to establish ground truth on patient outcomes/diagnoses. The "ground truth" for the substantial equivalence argument relies on comparison to the published specifications and regulatory clearances of the predicate devices.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set requiring adjudication was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document explicitly states "Clinical Studies: It is not applicable." Therefore, no MRMC study for human reader improvement with AI assistance was conducted or reported.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This device is a laser system hardware, not an AI algorithm. Its performance is inherent to its physical output characteristics, not an algorithm's classification or detection performance.

    7. The Type of Ground Truth Used

    The "ground truth" for this 510(k) submission is primarily:

    • Predicate Device Specifications: The established technical parameters and indications for use of the legally marketed predicate devices (K142078, K150138, K121363, K230047).
    • Industry Standards: The requirements laid out in the cited IEC and other international standards for medical electrical equipment and laser safety.

    8. The Sample Size for the Training Set

    Not applicable. This is a hardware device, not a machine learning model that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250018

    Validate with FDA (Live)

    Date Cleared
    2025-04-01

    (88 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The THEIA808 is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type 1-VI), including tanned skin.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The Diode laser is a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source.

    The diode laser for this unit is GaAlAs diode bar, and the wavelength is 808nm. It features impact structure, high efficiency and long lifetime. Generally the beam shall be emitted as the big beam divergence of the laser from the diode. With the unique fiber-coupling technology, the laser beam can be coupled efficiently into the handpiece.

    When the THEIA808 works, the thermoelectric cooler cools the light outlet, and the main device measures the working temperature of the light outlet through the NTC temperature measuring resistor.

    When the temperature reaches the minimum temperature, the TEC is controlled to stop working. The heat generated by thermoelectric cooler is conducted away by the water flow in the cooling module.

    The energy of the laser is preferentially absorbed by the melanin in the hair, so that the hair follicle produces high heat. Thereby destroying the structure of the hair follicle, making the hair lose the ability to regenerate, and achieving the purpose of hair removal. The light outlet (treatment head) adopts a uniquely designed dynamic cooling device.

    During the hair removal process, it can ensure that the epidermis will not be burned, so that the laser heat is limited to damage the hair follicles, while allowing the heat to diffuse from the epidermis, so that the epidermis can be anesthetized in a short time. The skin will not reach the damage threshold and can be protected from heat damage, thus ensuring a painless, fast and permanent hair removal effect.

    AI/ML Overview

    This document is a 510(k) clearance letter for the Medical Diode Laser Systems (THEIA808). It does not contain any information about acceptance criteria or a study proving the device meets acceptance criteria related to AI or algorithm performance.

    The document specifically states:

    • "No clinical study is included in this submission." This means there are no clinical trials proving efficacy or performance based on patient outcomes related to hair removal or reduction using this specific device.
    • The clearance is based on "Substantial Equivalence" to a predicate device (K162659 Diode Laser Hair Removal System). This means the FDA determined the new device is as safe and effective as a legally marketed predicate device, primarily through non-clinical testing and comparison of specifications (e.g., wavelength, power, cooling methods).
    • The non-clinical tests conducted verified compliance with electrical safety, EMC, laser safety, and biocompatibility standards. These are not performance metrics for hair removal efficacy or AI assistance.

    Therefore, I cannot provide the requested information regarding acceptance criteria and study details for AI or algorithm performance because the provided text does not contain any such details. The device itself is a laser system, not an AI or algorithmic diagnostic/assistive device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241791

    Validate with FDA (Live)

    Date Cleared
    2025-03-19

    (271 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    (1) The Medical Diode Laser Systems, model: GBOX-6H, VELAS II-6H is indicated for use in surgical applications requiring the vaporization, excision, ablation, cutting and hemostasis, or coaqulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The output wavelength of model GBOX-6H and VELAS II -6H is 1940nm.

    (2) The Medical Diode Laser Systems, model: GBOX-10D, GBOX-12D is indicated for use in surgical applications requiring the vaporization, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endloscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The output wavelength of model GBOX-10D and GBOX-12D is 1470nm.

    (3) The Medical Diode Laser Systems, model: GBOX-15D, VELAS II -15D is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The GBOX-15D, VELAS II - 15D is further indicated for laser assisted lipolysis. The output wavelength of model GBOX-15D and VELAS II -15D is 1470nm.

    (4) The Medical Diode Laser Systems, model: VELAS Pro-30B15D is indicated for use in surgical applications requiring the vaporization, incision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, Endovascular coagulation and endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. Medical Diode Laser Systems, model: VELAS Pro-30B15D is further indicated for laser assisted lipolysis. The output wavelength of model VELAS Pro-30B15D is 980nm or 1470nm.

    Device Description

    The "Medical Diode Laser Systems", include model: GBOX-6H, VELAS II -6H, GBOX-10D, GBOX-12D, GBOX-15D, VELAS II-15D, VELAS Pro-30B15D which consist of three main components: Main device; Foot switch; Accessories. Diode laser is a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source. The diode laser for this unit is GaAlAs diode bar, and the wavelength is 1940nm for GBOX-6H, VELAS II -6H; 1470nm for GBOX-10D, GBOX-12D, GBOX-15D, VELAS III-15D; 980nm or 1470nm for VELAS Pro-30B15D. It features impact structure, high efficiency and long lifetime. Generally, the beam shall be emitted as the big beam divergence of the laser from the diode. With the unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber. MCU (Micro Controller Unit) is the control center of system. MCU controls the drive board and temperature system by changing operating current. The operating power supply is the power source of the whole system. It supplies power for MCU, diode laser module by DC-DC conversion module and temp system by temp-control circuit. MCU controls the drive current of diode laser module by adjusting the DC-DC module and current-control circuit. The MCU has three external input sources: footswitch, remote interlock connector and information input from the touch units, it has one external output unit: display terminal. The diode laser module provides laser power output by fiber with optical fiber coupling system.

    AI/ML Overview

    The provided document is a 510(k) summary for Medical Diode Laser Systems. It focuses on establishing substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study demonstrating the device meets those criteria in terms of clinical performance. The document primarily discusses the similarity of the proposed device to predicate devices. Therefore, much of the requested information cannot be extracted directly from this type of regulatory submission.

    Here's a breakdown of what can be inferred or explicitly stated:

    1. A table of acceptance criteria and the reported device performance

    The document does not present explicit acceptance criteria for clinical performance (e.g., specific efficacy rates or safety profiles) for the proposed device, nor does it report its performance against such criteria. Instead, it compares the technological characteristics of the proposed device to predicate devices. The "acceptance criteria" here are implicitly meeting the performance characteristics of the predicate devices to prove substantial equivalence.

    Acceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance (as compared to Predicate)
    Indications for Use (GBOX-6H, VELAS II-6H): Surgical applications requiring vaporization, incision, excision, ablation, cutting, hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for various medical specialists.Substantially Equivalent (Same indications as K222701, covered by K193656)
    Indications for Use (GBOX-10D, GBOX-12D): Surgical applications requiring vaporization, incision, excision, ablation, cutting, hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for various medical specialists.Substantially Equivalent (Same indications as K222701, covered by K160952)
    Indications for Use (GBOX-15D, VELAS II-15D): Endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux; laser assisted lipolysis.Substantially Equivalent (Same indications as K160952, covered by K222701)
    Indications for Use (VELAS Pro-30B15D): Surgical applications requiring vaporization, incision, excision, ablation, cutting, hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for various medical specialists; endovascular coagulation and endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux; laser assisted lipolysis.Identical to K240747
    Laser Type: Diode laserIdentical (Diode laser)
    Wavelength (GBOX-6H, VELAS II-6H): 1940nmIdentical (1940nm)
    Wavelength (GBOX-10D, GBOX-12D, GBOX-15D, VELAS II-15D): 1470nmIdentical (1470nm)
    Wavelength (VELAS Pro-30B15D): 980nm or 1470nmIdentical (980nm or 1470nm)
    Output Power (GBOX-6H, VELAS II-6H): 1-6WSimilar (Predicate K222701 at 5W, Predicate K193656 up to 7.5W)
    Output Power (GBOX-10D, GBOX-12D, GBOX-15D): GBOX-10D: 1-10W; GBOX-12D: 1-12W; GBOX-15D: 1-15WSimilar (Within scope of Predicate K222701: Max. 12W and Predicate K160952: Max. 15W)
    Output Power (VELAS Pro-30B15D): 980nm: 30W; 1470nm: 15WIdentical (Same as K240747)
    Pulse Width (GBOX-6H, VELAS II-6H): 10ms-10sSimilar (Predicate K222701: 10ms-30s; Predicate K193656: 10ms-10s)
    Pulse Width (GBOX-10D, GBOX-12D, GBOX-15D, VELAS II-15D): 10ms-10sSimilar (Predicate K222701: 10ms-30s; Predicate K160952: 10-990ms)
    Pulse Width (VELAS Pro-30B15D): 10ms-1sIdentical (Same as K240747)
    Operation Mode: CW, single pulse, repeat pulseIdentical (CW, pulsed, single pulse; Continuous wave, single pulse, pulsed)
    Operation Interface: Color LCD touch screenIdentical (Color touch screen; Touch screen control; LCD touch screen)
    Repetition Rate (GBOX-6H, VELAS II-6H): 0.05Hz-50HzSimilar (Predicate K222701: 0.02-50Hz; Predicate K193656: CW or up to 100Hz)
    Repetition Rate (GBOX-10D, GBOX-12D, GBOX-15D, VELAS II-15D): 0.05Hz-50HzSimilar (Predicate K222701: 0.02-50Hz; Predicate K160952: CW or up to 100Hz)
    Aiming Beam: Specifications varying by modelSubstantially Equivalent
    Laser Class: 4Identical
    Electrical Safety and EMC: Conformity to IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22, IEC 60601-1-6Verified.
    Software Verification and Validation: Conformity to FDA Guidance, "moderate" level of concern.Documentation provided.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document explicitly states: "Animal Study - It is not applicable." and "Clinical Studies - It is not applicable." This indicates that no clinical "test set" (i.e., patient data for direct performance evaluation) was used for this 510(k) submission. The data provenance is therefore not relevant in this context.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Since no clinical studies or test sets with patient data were conducted for this submission, there were no experts used to establish ground truth for such a test set. The submission relies on establishing equivalence to legally marketed predicate devices, implying their established safety and efficacy.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a laser system, not an AI-powered diagnostic tool, and no MRMC studies were mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a hardware medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no clinical studies requiring a ground truth for performance evaluation were conducted. The "ground truth" in this context is the safety and efficacy profiles of the predicate devices as legally marketed.

    8. The sample size for the training set

    Not applicable, as this is a hardware medical device, not an AI/ML algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable, as this is a hardware medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K234004

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-20

    (429 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the "CHARISMA" is:
    -Dermatological procedures requiring coagulation of soft tissue and skin resurfacing procedures. further indicated for treatment of benign pigmented lesions, such as, but not limited to lentigines (age spots), solar lentigos (sun spots), melasma, dyschromia, and for treatment of facial wrinkles and fine lines.
    -Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue
    -Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux
    -Further indicated for laser assisted lipolysis.

    The intended use of the "REGAL" is:
    -Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue
    -Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux
    -Further indicated for laser assisted lipolysis.

    Device Description

    Diode laser is a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source. The diode laser for this unit is GaAlAs diode bar, and the wavelength is 1470nm. It features impact structure, high efficiency and long lifetime. Generally, the beam shall be emitted as the big beam divergence of the laser from the diode. With the GIGAA'S unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber. MCU (Micro Controller Unit) is the control center of system. MCU controls the drive board and temperature system by changing operating current. The operating power supply is the power source of the whole system, it supplies power for MCU, diode laser module by DC-DC conversion module and temp-control circuit. MCU controls the drive current of diode laser module by adjusting the DC-DC module and current-control circuit. The MCU has three external input sources: footswitch, remote interlock connector and information input from the touch units, it has one external output unit: display terminal. The diode laser module provides laser power output by fiber with optical fiber coupling system.

    AI/ML Overview

    The document provided is a 510(k) premarket notification for two medical diode laser systems, CHARISMA and REGAL. It aims to demonstrate substantial equivalence to legally marketed predicate devices, not to prove the device meets specific acceptance criteria through a clinical study. Therefore, much of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth cannot be extracted from this document in the typical sense of a clinical performance study for an AI/software device.

    However, I can provide information on the non-clinical tests performed to demonstrate safety and performance equivalence, as these serve as the "acceptance criteria" for this type of submission.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this 510(k) submission, the "acceptance criteria" are compliance with established electrical safety, EMC, and laser safety standards, as well as satisfactory software verification and validation. The "reported device performance" is the demonstration of compliance through non-clinical testing.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Standard & Version)Reported Device Performance (Compliance Status)
    Electrical SafetyIEC60601-1: 2005/(R)2012Conformity verified
    Electromagnetic Compatibility (EMC)IEC60601-1-2: 2014-02Conformity verified
    Laser SafetyIEC 60825-1: 2007-03Conformity verified
    Laser Equipment PerformanceIEC 60601-2-22: 2012-10Performance verified
    Software Verification and ValidationFDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Moderate Level of Concern)Testing conducted and documentation provided, demonstrating compliance

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable. The testing described is non-clinical, involving the physical device and its software being tested against engineering and safety standards, rather than a "test set" of patient data or samples.
    • Data Provenance: Not applicable. The testing involves the device itself and its components, not patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is a non-clinical submission. Ground truth, in the context of clinical studies for AI/software, refers to definitive diagnoses or measurements. Here, the "ground truth" is compliance with international safety and performance standards. The "experts" involved would be the engineers and quality assurance personnel conducting and verifying the tests, but their specific number and qualifications are not detailed in this document.

    4. Adjudication Method for the Test Set

    Not applicable. There is no "adjudication method" in the sense of resolving discrepancies in expert opinions on patient data. Compliance with standards is typically determined by objective measurements and verification protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states under "VIII Clinical Testing" that it "is not applicable." This type of study is usually performed for diagnostic AI devices where human performance with and without AI assistance is being compared.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone performance study in the context of AI/software algorithm only was not done. The device is a physical medical laser system, and the non-clinical tests assess the safety and performance of the hardware and integrated software, not a standalone algorithm.

    7. The Type of Ground Truth Used

    As this is a non-clinical submission, the "ground truth" is compliance with recognized international safety and performance standards (IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22) and FDA guidance for software. These standards define the acceptable limits and behaviors for medical electrical equipment, laser products, and software.

    8. The Sample Size for the Training Set

    Not applicable. This document describes a medical device's safety and performance testing for regulatory clearance, not the development or training of an AI algorithm. Therefore, there is no "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set mentioned or implied in this regulatory submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230047

    Validate with FDA (Live)

    Date Cleared
    2024-03-04

    (423 days)

    Product Code
    Regulation Number
    878.4810
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medical Diode Laser Systems are intended for intra- and extra-oral surgery including incision, excision, hemostasis, coagulation and vaporization of soft tissue including marginal and epithelial lining of free gingiva and are indicated for: frenectomy; frenotomy; biopsy; operculectomy; implant recovery; gingivoplasty; gingivoplasty; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery;debridement of diseased epithelial lining; incisions and draining of abscesses; tissue retraction for impressions;papillectorny; vestibuloplasty ; excision of lesions; exposure of unerupted partially erupted teeth; removal of hyperplastic tissues; treatment of aphthous ulcers; leukoplakia; laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket; sulcular debridement (removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss and tooth inability); pulpotomy; pulpotomy as adjunct to root canal therapy; fibroma removal; gingival incision and excision; treatment of canker sores; herpetic ulcers of the oral mucosa; laser soft tissue curettage; reduction of gingival hypertrophy.

    Whitening: Medical Diode Laser Systems are indicated for light activation materials for teeth whitening and for laser-assisted whitening/bleaching of teeth.

    Low Level Laser Therapy: Medical Diode Laser Systems are intended to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffins pain, or muscle spasm, and for the temporary increase in local blood circulation and/or temporary relaxation of muscles.

    Device Description

    The Medical Diode Laser System is a compact, air-cooled unit. It is a complete self-contained instrument, which includes a high efficiency power supply, a microprocessor controlled, adjustable light output with automatic power stabilization, (fan cooled), as well as a switch panel and LCD screen display panel designed to be user friendly. The system includes high power lasers, safety features and an SMA fiber output connector. The system includes a Lithium battery, it can work 1hour when not external power input at maximum power output in pulse or CW but intermittent mode. The diodes are made from GaAlAs semiconductor material for high output and superior reliability. The diode lasers are enclosed in a rugged, factory-aligned, replaceable, environmentally protective module. High-capacity fans eliminate the need for water-cooling, and assure low maintenance and reliable laser operation. The diodes convert electric energy into coherent laser radiation with a wavelength of 810nm or 980nm+/-10nm (pilot beam: 650nm+/-10nm).

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device (Medical Diode Laser Systems) and explicitly states that no clinical testing was performed. Therefore, the request for "acceptance criteria and the study that proves the device meets the acceptance criteria" cannot be fully answered as it pertains to clinical performance and AI integration.

    The provided document focuses on demonstrating substantial equivalence to predicate devices primarily through non-clinical testing (electrical safety, EMC, laser safety standards, and software verification). There is no mention of AI/ML, human-in-the-loop performance studies, or ground truth establishment based on expert consensus, pathology, or outcomes data, which are typical requirements for AI/ML device clearances.

    Based on the provided text, here's what can be inferred/extracted and what cannot be answered:

    1. Table of acceptance criteria and the reported device performance:

    Since no clinical or AI performance study was presented, the "acceptance criteria" here refer to conformance with established safety and performance standards for laser medical devices.

    Acceptance Criteria (from Non-Clinical Tests)Reported Device Performance (Compliance)
    Conformity to IEC 60601-1 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance)Verified
    Conformity to IEC 60601-1-2 (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance Collateral Standard: Electromagnetic compatibility)Verified
    Conformity to IEC 60825-1 (Safety of laser products - Part 1: Equipment classification and requirements)Verified
    Performance according to IEC 60601-2-22 (Medical electrical equipment Part 2: Particular Requirements for basic safety and essential performance of surgical, cosmetic, therapeutic, and diagnostic laser equipment).Verified
    Software Verification and Validation (for "moderate" level of concern) as per FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."Conducted and documentation provided.
    Device Specifications (Comparison to predicate, serving as a form of "performance" for equivalence)(All stated as "Identical" or "Substantial Equivalence" to predicate/reference devices)
    Product Code (NVK, GEX, ILY)Identical (to predicate K210367)
    Regulation Number (21 CFR 878.4810)Identical
    Classification (Class II)Identical
    Indications for UseSubstantial Equivalence (Proposed device's intended use is same as predicate; reference device does not have light activation for bleaching materials for teeth whitening, but this difference does not affect safety and effectiveness based on the submission.)
    Application (Dental Laser)Identical
    Laser Classification (Class IV)Substantial Equivalence (Predicate states 976nm and 450nm: Class IV, 650nm: Class II; Reference states Class IV). The document implies this is an acceptable difference for substantial equivalence.
    Type of Laser (Diode Laser)Identical
    Wavelength (810 ± 10nm; 980 ± 10nm)Substantial Equivalence (Predicate: 976nm, 650nm, 450nm; Reference: 810nm, 980nm). Stated that modification (different wavelengths) does not adversely affect safety and effectiveness.
    Average Power (Range specific to models)Similar to predicate/reference device. Example: Proposed DEN7A/DEN7B: 0.1W-7W vs. Predicate D-Laser Blue 976nm: 0.2W-4W (CW), 7W (peak). Reference: 0.1-2W, 20W (peak). The ranges for the proposed device generally align with or explain differences within the scope of the predicate devices.
    Operation Mode (Continuous Wave; Pulse)Identical to the predicate.
    Pulse width (e.g., 1ms-1000ms for DEN models)Similar (Predicate: 5 µsec. to 0.9 sec.; Reference: Variable).
    Pulse repetition rate (e.g., 0.5Hz-500Hz for DEN models)Similar (Predicate: 1 Hz—20 kHz; Reference: 50Hz).
    Aiming Beam (650±10 nm, 5mw (max))Substantial Equivalence (Predicate: 650±20 nm, Pmax<5 mW; Reference: 5mW laser diode, 650nm, Class 1).
    Delivery system (Fiber optic cable, handpiece and fiber tips)Substantial Equivalence (Proposed: 200µm, 400µm, 600µm and 800µm fibers; Predicate: 200µm, 300µm, 400µm; Reference: 400µm). Differences in fiber sizes are considered substantially equivalent.
    Operation interface (Color touch screen graphical user interface)Substantial Equivalence (Predicate: Color touch screen graphical user interface; Reference: Electroluminescent glass display with capacitive touch interface). Considered substantially equivalent.
    Activation Means (Foot Switch, with electronic access key)Substantial Equivalence (Predicate: Handpiece finger switch, Wireless foot switch; Reference: Wireless Foot Switch, with electronic access key). Differences in activation methods are considered substantially equivalent.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not applicable/Not provided for AI/ML performance testing. The document refers to non-clinical tests on the device itself.
    • Data Provenance: Not applicable for AI/ML performance testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable as no AI/ML clinical performance study requiring expert ground truth was conducted. For device testing, compliance with standards is the "ground truth" and is assessed by qualified test engineers/labs.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable as no AI/ML clinical performance study requiring adjudication was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states "Clinical Testing: It is not applicable."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, a standalone AI algorithm performance study was not done. The device is a physical laser system, not an AI algorithm for diagnostic or therapeutic assistance. The "Software Verification and Validation" mentioned pertains to the embedded software controlling the laser, not a separate AI model.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the non-clinical tests, the "ground truth" is compliance with international and national recognized standards (e.g., IEC 60601-1, IEC 60825-1) and internal design specifications, verified through laboratory testing. There is no clinical or AI-related ground truth as defined in the prompt.

    8. The sample size for the training set:

    • Not applicable, as this is not an AI/ML device submission for which a training set would be described.

    9. How the ground truth for the training set was established:

    • Not applicable, as this is not an AI/ML device submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211977

    Validate with FDA (Live)

    Date Cleared
    2021-10-26

    (123 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The "CHERYLAS-15N and CHERYLAS-20N" are indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

    Device Description

    Diode laser is a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source.The diode laser for this unit is GaAlAs diode bar, and the wavelength is 1470mm. It features impact structure, high efficiency and long lifetime. Generally the beam shall be shamed as the big beam divergence of the laser from the diode. When the coaxiality of laser and fiber meet the requirements, the laser beam can be coupled efficiently into the fiber.

    The block diagram shows how the diode laser system works. MCU (Micro Controller Unit) is the control center of system. MCU controls the drive board and temperature system by changing operating current. The operating power supply is the power source of the whole system, it supplies power for MCU, diode laser module by DC-DC conversion module and temp system by temp-control circuit. MCU controls the drive current of diode laser module by adjusting the DC-DC module and current-control circuit. The MCU has three external input sources: footswitch, remote interlock and information input from the touch units, it has one external output unit: display terminal. The diode laser module provides laser power output by fiber with optical fiber coupling system. The system can test and calibrate power by the calibration and feedback unit.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device (Medical Diode Laser Systems) and primarily focuses on proving substantial equivalence to a predicate device based on technical characteristics and adherence to safety standards. It explicitly states "Clinical data: Not applicable." Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving a device meets those criteria, specifically concerning performance metrics like sensitivity, specificity, or the improvement of human readers with AI assistance.

    The information provided pertains to the device's technical specifications, intended use, and comparison to a predicate device, focusing on safety and technological equivalence rather than AI/algorithm performance.

    To answer your request, the document would need to include details about a clinical study or a robust non-clinical performance study with specific metrics, a test set, ground truth establishment, and statistical analysis, which is absent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160549

    Validate with FDA (Live)

    Date Cleared
    2017-04-07

    (403 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The "VELASI -15D" is indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

    Device Description

    Medical diode laser systems are a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source. The medical diode laser systems for this unit is GaAlAs diode bar, and the wavelength is 1470nm of VELAS II-15D. It features impact structure, high efficiency and long lifetime. The beam divergence angle is too large lead to poor beam quality. In order to better meet the needs of Laser Diode practical applications, Laser Diode carried beam shaping is very necessary. With the GIGAA'S unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber. To start the laser unit, turn the main switch ON and turn the key switch clockwise to "I" position. The power indicator will turn green immediately with the system fans working. At the same time, the LCD screen lights up. After the parameters setting is finished, press the "Ready" button and the system will remind you to wear the protective eyewear (protective wavelength is from 800nm to 1700mm). At this time when you press down the footswitch, the laser will emit.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device, specifically a Medical Diode Laser System (VELAS II -15D). It focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than defining specific performance acceptance criteria for the novel device itself.

    Therefore, the document does not explicitly state specific acceptance criteria in terms of numerical performance thresholds (e.g., accuracy, sensitivity, specificity) for the device's clinical efficacy in treating reflux of saphenous veins. Instead, the "acceptance criteria" are implied by the demonstration of substantial equivalence to the predicate device, meaning the new device should perform as safely and effectively as the predicate.

    The "study that proves the device meets the acceptance criteria" refers to the non-clinical testing performed to demonstrate adherence to relevant safety and performance standards, and a comparison presented in the "Summary of Substantial Equivalence" table.

    Here's an analysis based on the information provided, though it will not contain specific performance metrics for clinical efficacy as none are present.


    1. Table of Acceptance Criteria and the Reported Device Performance

    As specific clinical performance acceptance criteria (e.g., success rate of vein closure, reduction in vein diameter) are not provided in this regulatory submission for the novel device, the "acceptance criteria" are inferred from the comparison to the predicate device, primarily focusing on technical specifications and safety standards.

    Acceptance Criterion (Inferred from Substantial Equivalence)Reported Device Performance (VELAS II -15D)
    Indications for Use"The 'VELAS II -15D' is indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities."
    Laser TypeDiode laser
    Wavelength$1470nm \pm 10nm$
    Output power1-15W
    Operation modeCW, single pulse, repeat pulse
    Pulse width10ms-2.5s
    Pulse repetition rate0.2Hz-50Hz
    Application / Light delivery systemRecommended disposable sterile fiber (K124003, MED-Fibers, Inc.), bare fiber, 3m, 600μm fiber core diameter, NA ≥0.22, SMA905 connector, single use.
    Aiming BeamDiode laser of 635/532nm, power max. < 5mW, adjustable brightness.
    Laser Class4
    Operation interfaceColor LCD touch screen
    Power Supply100-240VAC, 50/60Hz, 350VA
    Safety classificationClass I Type B
    Waterproof levelIPX1 (device) / IPX8 (footswitch)
    Standards ComplianceIEC60601-1, IEC60601-1-2, IEC60601-2-22, IEC60825-1, FDA Guidance for Software.
    Non-sterileFiber is sterilized by EO.
    Microprocessor ControlYes
    Substantial Equivalence Conclusion"The proposed device of VELAS II -15D has the same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performance safety as the predicate device."

    Comparison to Predicate (K082225):

    • Indications for Use: Same.
    • Laser Type: Same.
    • Components: Similar (Proposed device does not equip with fiber and handpiece initially, but recommends a specific disposable fiber set; predicate includes these components directly). The submission argues this does not affect safety and effectiveness.
    • Wavelength: Similar ($1470nm \pm 10nm$ vs $1470nm \pm 30nm$). Proposed device has better accuracy.
    • Output power: Same (1-15W).
    • Operation mode: Same (CW, single pulse, repeat pulse).
    • Pulse width: Similar (10ms-2.5s vs 0.01s-60s). Proposed device's range is contained within the predicate, deemed sufficient for intended use.
    • Pulse repetition rate: Similar (0.2Hz-50Hz vs 0.01Hz-50Hz). Deemed not to affect safety and effectiveness.
    • Application / Light delivery system: Similar. Proposed device is more detailed, fiber core diameter range is contained within predicate's general description, deemed sufficient for intended use.
    • Aiming Beam: Similar. Proposed device's power range contains predicate's, deemed sufficient for intended use.
    • Laser Class: Same (4).
    • Operation interface: Same (Color LCD touch screen).
    • Power Supply: Similar (350VA vs 450VA). Rated power difference not affecting safety and effectiveness.
    • Safety classification: Same (Class I Type B).
    • Dimensions & Weight: Similar, proposed device is bigger and heavier. Not noted to affect safety or effectiveness.
    • Waterproof level: Same (IPX1 device, IPX8 footswitch).
    • Standard Compliance: Same listed standards.
    • Non-sterile: Same (Fiber sterilized by EO).
    • Microprocessor Control: Same (Yes).

    2. Sample size used for the test set and the data provenance

    The document specifies non-clinical tests (compliance with standards) and software verification. It does not describe a clinical study or a test set in the sense of patient data. Therefore, information regarding "sample size used for the test set" and "data provenance" (country of origin, retrospective/prospective) is not applicable or not provided in this regulatory submission, which focuses on substantial equivalence based on technical specifications and international standards compliance.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    As no clinical test set/study is described, the concept of establishing ground truth by experts is not applicable or not provided in this document. The "ground truth" for the device's technical specifications and safety is established by adherence to international standards and internal quality assurance procedures, as overseen by the manufacturer's technical staff and verified through testing.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set or study requiring adjudication is described.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a surgical laser system, not an AI-assisted diagnostic tool or an imaging device involving "human readers." No MRMC study was mentioned or required for this type of submission.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a laser surgical instrument, meaning it inherently has a "human-in-the-loop" (the surgeon) performing the procedure. No standalone algorithm performance is relevant or described.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the technical performance aspects, the "ground truth" relies on:

    • International standards compliance: Demonstrated by testing against IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.
    • Software verification: According to FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
    • Predicate device's established safety and effectiveness: The core of the 510(k) submission is to argue that the new device is "substantially equivalent" to an already approved predicate device, implying its "ground truth" for safety and effectiveness is comparable to that of the predicate, which has presumably demonstrated its clinical efficacy through other means (e.g., initial clinical trials for its approval, or long-standing safe use).

    No specific clinical "outcomes data" or "expert consensus" related to a new discrete clinical study for this specific device is provided in the document.


    8. The sample size for the training set

    Not applicable. This submission concerns a physical medical device (laser system), not an AI/machine learning model that requires a training set.


    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI/machine learning model described.