510(k) Data Aggregation

    K Number
    K111381

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-01-13

    (241 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    1.0 The Medi-SIS Syringe Infusion System is indicated for the delivery of enteral fluids through clinically accepted routes of administration.
    2.0 The Medi-SIS Administration Extension Set is intended for single patient use only.
    3.0 The Medi-SIS Syringe Driver can be reused up to 5,000 times

    Device Description

    The Medi-SIS Syringe Infusion System consists of a Syringe Driver (mechanical syringe pump), proprietary administration extension set, and compatible syringe.

    The Medi-SIS Syringe Driver consists of a mechanical syringe pump assembly that is manufactured from plastic housings and end caps, and steel spring (mechanical force from the spring presses against the syringe plunger), and holds a 30mL or 60mL Becton Dickinson syringe (preamendment device) or a 30mL or 60mL Acacia syringe (K092986). The Medi-SIS Syringe Driver must be used in conjunction with a Medi-SIS Administration Extension Set, which controls the time the infusion will occur (by flow restrictor) for the entire volume of the filled syringe. The Medi-SIS Syringe Infusion System infuses fluids in a continuously decreasing rate and the user can calculate the average flow rate based from volume and time infused.

    The Medi-SIS Syringe Driver is reusable and provided non-sterile.

    The Medi-SIS Administration Set consists of various lengths of flexible non-DEHP PVC tubing. on/off clamp, flow restrictor, and male and female connectors. The male and female connectors meet Acacia's internal specifications for enteral applications to ensure cross connections do not occur between IV and Enteral infusions. Additional componentry may be added and configurations modified based on the needs of clinicians.

    The following Administration Extension Sets will be available:

    60mL Syringe Size, 1.0mL per minute delivery rate 60mL Syringe Size, 2.0mL per minute delivery rate 30mL Syringe Size, 0.5mL per minute delivery rate 30mL Syringe Size, 1.0mL per minute delivery rate 30mL Syringe Size. 2.0mL per minute delivery rate

    The approximate residual volume for the syringe/administration set combination is approximately 2.20mL for all configurations. The Medi-SIS Administration Extension Set is provided sterile, single use, and non-pyrogenic.

    AI/ML Overview

    The provided 510(k) summary (K111381) describes performance testing for the Medi-SIS Syringe Infusion System, specifically for the addition of enteral fluid delivery to its intended use.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device Performance
    Enteral fluid delivery within specified timeMet all of the predicate device's specifications.
    Maximum residual volumeApproximate residual volume for syringe/administration set combination is approximately 2.20mL for all configurations (60mL syringe at 1.0mL/min, 60mL syringe at 2.0mL/min, 30mL syringe at 0.5mL/min, 30mL syringe at 1.0mL/min, 30mL syringe at 2.0mL/min).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size (e.g., number of tests, number of devices) used for the non-clinical tests assessing enteral fluid delivery time and residual volume. It also does not specify the provenance of the data (country of origin, retrospective or prospective). It simply states that "non-clinical tests were performed."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. The tests performed were non-clinical, evaluating device performance (fluid delivery and residual volume) against specifications, not requiring expert human interpretation or ground truth establishment in a medical context.

    4. Adjudication Method for the Test Set

    Not applicable. As the tests were non-clinical performance evaluations, no adjudication method (like 2+1 or 3+1 used for expert consensus in medical image analysis) was required.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission focuses on non-clinical performance, not human reader performance with or without AI assistance. The device is an infusion pump, not an AI-powered diagnostic tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, in a sense, a "standalone" performance study was done for the device itself, as the tests focused on the physical device's ability to deliver enteral fluids as specified. However, this isn't in the context of an "algorithm" or AI. The device's performance was evaluated independently of human interaction during the test, purely on its mechanical and fluid delivery capabilities.

    7. Type of Ground Truth Used

    The "ground truth" for the non-clinical tests was based on predicate device specifications and internal Acacia specifications for performance. For example, enteral fluid delivery was verified against the specified time and the residual volume had a specific measurable value.

    8. Sample Size for the Training Set

    Not applicable. This device is an infusion pump, not an AI/machine learning device that requires a training set. The performance evaluation was based on non-clinical engineering tests.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There was no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K972173

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-08-15

    (67 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Medi-SIS Syringe Infusion System was designed for general drug and solution infusion uses with the limitation being the volume of drug to be infused. The routes of administration include those of the similar size syringes, (i.e. intravenous infusions and subcutaneous infusions) ... ....................................................................................................................................
    2. The administration set is intended for single patient use only.
    3. The Medi-SIS driver can be reused up to 5,000 times.
    4. The Medi-SIS system is intended for general infusion use. The Medi-SIS Syringe Infusion System is not limited to any specific type of therapy.
      4.1 The Medi-SIS Syringe Infusion System is not intended for the delivery of blood, blood products or TPN solutions.
    Device Description

    The Medi-SIS Syringe Infusion System consists of a plastic, spring driven infusion pump and various administration sets. Two models of the Medi-SIS pump have been developed to accommodate 20 and 60 ml syringes.

    AI/ML Overview

    The provided documentation describes the I-Flow Medi-SIS Syringe Infusion System, specifically regarding a change in labeling to include Sherwood Monoject Syringes in addition to the previously approved Becton-Dickinson® brand syringes. The submission (K972173) is a 510(k) summary rather than a full study report, and therefore much of the information typically found in a comprehensive clinical or performance study is not present. However, it does provide performance test data related to the device's function.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" in a separate section. However, it implicitly assesses performance against the "labeled delivery time" of the device and states that performance within ±15% of the labeled time is considered acceptable. It also notes that slower delivery rates (even beyond this range) might not affect safety or efficacy for certain drug types.

    Performance MetricAcceptance Criteria (Implied)20ml Sherwood Monoject Syringe (Lot # 724040)20ml Sherwood Monoject Syringe (Lot # 723462)60ml Sherwood Monoject Syringe
    Delivery Time (20ml/15min set)Within ±15% of labeled 15min (implicitly, though deviations noted)19.07 min (27% increase)19.60 min (31% increase)N/A
    Delivery Time (20ml/30min set)Within ±15% of labeled 30min (implicitly, though deviations noted)37.14 min (24% increase)38.60 min (29% increase)N/A
    Delivery Time (60ml/30min set)Within ±15% of labeled 30min (±4.5 min)N/AN/A28.09 min (6% decrease)
    Delivery Time (60ml/60min set)Within ±15% of labeled 60min (±9 min)N/AN/A56.57 min (6% decrease)
    Residual Volume (20ml syringe)"< 1.3 ml" (stated as the average for Monoject)1.19 ml (20ml/15min)0.77 ml (20ml/30min)N/A
    Residual Volume (60ml syringe)"< 1.6 ml" (stated as the average for Monoject)N/AN/A1.55 ml (60ml/30min) & 1.10 ml (60ml/60min)
    Burst PressureMaintain structural integrity up to 60 psiNo effect on structural integrity at 60 psiNo effect on structural integrity at 60 psiNo effect on structural integrity at 60 psi

    Note on Delivery Time Acceptance: For the 20ml syringe, the reported increased delivery times (24% to 31%) exceed the general ±15% specification. However, the document states: "The slower delivery rate has no effect on safety or efficacy of the device when reviewed against its intended use. The types of drugs that are commonly used with this device are generally given by IV push which doesn't have a specified flow rate." This indicates a practical interpretation of acceptance criteria based on clinical context rather than a strict numerical threshold for all situations.

    2. Sample Size Used for the Test Set and Data Provenance

    The data used for testing are from the manufacturer's internal performance tests for the Medi-SIS Syringe Infusion System when used with Sherwood Monoject Syringes. The documentation does not specify the country of origin of the data but implies it was conducted by I-Flow Corporation. The tests appear to be prospective, laboratory-based performance tests evaluating the device.

    • Delivery Time (Sherwood Monoject):
      • 20ml/15min set: n=30 per syringe lot (two lots tested: #724040, #723462, so 60 total)
      • 20ml/30min set: n=30 per syringe lot (two lots tested: #724040, #723462, so 60 total)
      • 60ml/30min set: n=30
      • 60ml/60min set: n=30
    • Temperature Effect (Sherwood Monoject):
      • 20ml/15min set: n=20 (52°F, 90°F), n=30 (72°F) for Lot #724040
      • 20ml/30min set: n=20 (52°F, 90°F), n=30 (72°F) for Lot #724040
      • 60ml/30min set: n=20 (50°F, 90°F), n=30 (72°F)
      • 60ml/60min set: n=20 (50°F, 90°F), n=30 (72°F)
    • Residual Volumes (Sherwood Monoject):
      • 20ml syringe: n=5 for both 20ml/15min and 20ml/30min sets
      • 60ml syringe: n=5 for both 60ml/30min and 60ml/60min sets
    • Burst Pressure: The sample size is not explicitly stated, but it mentions "The 20 ml and 60 ml Sherwood Monoject Syringe have been tested..."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The "ground truth" here is the physical performance of the device (e.g., actual delivery time, residual volume), measured directly by instrumentation in laboratory settings, not by expert interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. This is a technical performance study, not a study requiring expert adjudication of clinical outcomes or images.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This device is an infusion pump, and the evaluation focuses on its mechanical performance, not human reader performance in interpreting results.

    6. Standalone (Algorithm Only) Performance Study

    Yes, the studies presented are essentially "standalone" performance evaluations of the device, albeit in conjunction with different types of syringes. The device itself (the Medi-SIS Syringe Infusion System) is being evaluated for its physical outputs (flow rate/delivery time, residual volume, structural integrity). There is no "human-in-the-loop" component to this performance evaluation.

    7. Type of Ground Truth Used

    The ground truth used is direct measurement of physical parameters in a controlled laboratory environment:

    • Delivery Time: Measured time for complete fluid delivery.
    • Residual Volume: Measured volume of fluid remaining in the syringe.
    • Burst Pressure: Testing to determine the pressure at which structural integrity is compromised.

    8. Sample Size for the Training Set

    Not applicable. This is a performance validation of a mechanical device, not an AI/ML model for which "training sets" are relevant. The device implicitly relies on its mechanical design and calibration, not on data training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.