Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K103189

    Validate with FDA (Live)

    Date Cleared
    2010-12-20

    (53 days)

    Product Code
    Regulation Number
    888.3353
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medacta Total Hip Prosthesis System – Quadra C Femoral Stems are intended for cemented use in total or partial hip arthroplasty and in primary or revision surgery. Hip replacement is indicated in the following cases: Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia; Avascular necrosis of the femoral head; Acute traumatic fracture of the femoral head or neck; Failure of previous hip surgery, joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement. Size 0 implants should not be implanted in patients with a mass of 65 kg or greater.

    Device Description

    This modification to the original Medacta Total Hip Prosthesis System -Quadra C femoral stems is a line extension to include the AMIStem C femoral stems and Quadra C Short Neck (SN) femoral stems. AMIStem C femoral stems are based on the design of the Quadra C femoral stems in the original submission. The AMIStem C femoral stems differ from the Quadra C femoral stems by a decreased length of 15% and a reduced shoulder. The Quadra C short neck femoral stems are the same design femoral stems as the Quadra C femoral stems in the original submission but the length of the neck that is 5 mm shorter. Like the Quadra C femoral stems, the AMIStem C femoral stems and the Quadra C SN femoral stems are highly polished, collarless femoral components manufactured from forged stainless steel according to ISO 5832-9. The AMIStem C femoral stems and the Quadra C SN femoral stems are straight quadratic stems with a triple taper shape. The proximal portion of the stem has a standard 12/14 taper for mechanical attachment to cleared Medacta International metallic or ceramic femoral heads. Like the Quadra C femoral stems, two different caput-collum-diaphyseal angles (CCD) are available for the AMIStem C femoral stems and the Quadra C Short neck femoral stems: 135° for the standard offset and 127° for the lateralized offset. The Quadra C short neck femoral stems come in eight sizes with a standard offset and seven sizes with a lateral offset. The AMIStem C femoral stems come in eight sizes with a standard offset and seven sizes with a lateral offset. There are no short neck versions for the AMIStem C femoral stems. The AMIStem C femoral stems and the Quadra C short neck femoral stems provide additional femoral stem options to the surgeon for use with the Medacta Total Hip Prosthesis System.

    AI/ML Overview

    This document describes the Medacta Total Hip Prosthesis System - AMIStem C and Quadra C SN femoral stems. As a 510(k) summary, the focus is on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with detailed performance metrics against specific acceptance criteria.

    The information provided does not include:

    • A specific table of acceptance criteria and reported device performance.
    • Details on sample sizes used for test sets, training sets, or data provenance.
    • Information on the number or qualifications of experts for ground truth, adjudication methods, MRMC studies, or standalone algorithm performance.
    • The type of ground truth used or how it was established for training sets.

    Instead, the document states that performance testing was conducted according to various standards and FDA guidance documents, with predefined acceptance criteria based on these standards. The conclusion is that all acceptance criteria were met, verifying the performance and supporting substantial equivalence.

    Here's a summary of the available information and a structured table for acceptance criteria, acknowledging the lack of specific numerical performance data within this summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (as implied)Reported Device Performance
    Material CompositionConformance to ISO 5832-9 for forged stainless steel"made of the same materials as the Quadra C femoral stems"
    General DesignMaintenance of general design principles of predicate device"same general design as the Quadra C femoral stems"
    CompatibilitiesCompatibility with existing Medacta Total Hip Prosthesis System components"same compatibilities as the Quadra C femoral stems"
    Mechanical EnduranceMeeting endurance standards for neck and stem regions"endurance testing of the neck and stem regions...met all acceptance criteria"
    Range of MotionUnspecified range of motion standards"range of motion analysis...met all acceptance criteria"
    Risk AnalysisNo new unacceptable risks identified from modifications"risk analysis to identify any new risks...design verification was conducted"
    Overall PerformanceMeeting all pre-defined acceptance criteria based on standards and FDA guidance"design verification activities met all acceptance criteria and verified the performance"

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. This document describes bench testing for a medical device rather than a clinical study involving human data. The testing was conducted on "worst case based on engineering analysis."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. Ground truth, in the context of this device, refers to meeting established engineering and material standards for the implant, not to expert interpretation of medical images or patient outcomes.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods are typically relevant for clinical studies or studies involving human judgment (e.g., image interpretation), which is not the nature of this submission. The "adjudication" here would be the verification against predefined engineering acceptance criteria.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No. This is a 510(k) submission for a hip prosthesis, focusing on substantial equivalence through bench testing, not a clinical effectiveness study involving human readers or AI.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, in a sense. The "performance testing" described is for the device itself (the femoral stem), without human interaction in its functional evaluation. The testing verifies the physical and mechanical properties of the implant.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance testing is adherence to established engineering standards and FDA guidance documents for hip prostheses, covering aspects like material properties, mechanical strength, endurance, and design specifications. This ensures the device is mechanically sound and safe for its intended use.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device; therefore, there is no "training set" in that context. The "training" for such a device would involve its design and manufacturing processes, which are guided by established engineering principles and standards.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the AI/ML sense. The design and manufacturing of the device are based on long-standing medical device engineering principles, material science, biomechanics, and ISO standards (e.g., ISO 5832-9 mentioned for stainless steel). The "ground truth" here is derived from scientific and engineering knowledge accumulated over time regarding safe and effective hip prostheses.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1