510(k) Data Aggregation

    K Number
    K133708

    Validate with FDA (Live)

    Date Cleared
    2014-02-14

    (72 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MD302C4 color display is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians.

    To guarantee the display performance as specified, it must only be used for in conjunction with NEC approved display controllers.

    MD302C4 cannot be used for a life-support system.

    This device must not be used in digital mammography.

    This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment.

    Device Description

    Medical Display, MD302C4 is a 29.8" Color LCD monitor that displays image for medical use. It provides 4 mega pixel (2560*1600) resolution with adjustable gamma gray scale for more precise diagnose use in CT, MRI, HIS, and PACS.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a diagnostic imaging LCD monitor (MD302C4). The submission aims to demonstrate substantial equivalence to a predicate device (MD242C2) rather than proving the device meets specific acceptance criteria through a clinical study. As such, most of the requested information (sample size, data provenance, expert qualification, adjudication method, MRMC study, standalone performance, training set details) is not applicable to this type of submission.

    Here's the information that can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state acceptance criteria in the typical sense of a clinical study with performance metrics. Instead, it compares the technical specifications and intended use of the new device (MD302C4) to a predicate device (MD242C2) to establish substantial equivalence. The "acceptance criteria" here are implicitly that the new device's performance characteristics are comparable or superior to the predicate device, or fall within acceptable ranges for diagnostic display devices.

    ItemsPredicate Device (MD242C2)Reported Device Performance (MD302C4)
    Panel Size and Type24.1" TFT Color LCD Monitor29.8" TFT Color LCD Monitor
    Pixel Pitch0.270 mm x 0.270mm0.251mm x 0.251mm
    Display Color1,073,741,82416,777,216
    Viewing Angles (°)H:178, V:178H:178, V:178
    Scanning Frequency(H, V)31.5-93.8, 118.4kHz, 50-85 Hz31.5-98.7 kHz, 50-85 Hz
    Native Resolutions1920X12002560X1600
    Brightness180 cd/m² calibrated,350 cd/m² Max.180 cd/m² calibrated,340 cd/m² Max.
    Contrast Ratio1000 : 1 (typical)1000 : 1 (typical)
    DOT Clock202.5 MHz (Max) (Analog)162 MHz (Max) (Digital)269 MHz (Max)
    Input SignalsThree connectors: one DVI port,one Display port, one HDMI port.Three connectors: one DVI port,one Display port, one HDMIport.
    Input TerminalsDVI-D, Display port, HDMI portDVI-D, Display port, HDMI port
    USB (option) / StandardNoNo
    Active Display Size (H x V)Landscape: 518.4mmX324mmPortrait: 324X518.4mmLandscape: 641mmX401mmPortrait: 401X641mm
    Viewable Image Size540 mm (diagonal)756 mm
    Luminance CalibrationSoftwareSoftware
    Default Gamma1.8,2.0,2.2 DICOM part 141.8,2.0,2.2 DICOM part 14
    PowerAC100-240V, 50/60HzAC100-240V, 50/60Hz
    Power Consumption38.4W (Max)87W (Max)
    Power Save Mode<2W<2W
    Dimensions (W x H x D)W:Landscape: 556.8mmPortrait: 362.4 mmH:Landscape: 378 - 528mmPortrait: 572.4-625.2mmD: 227.6 mmW:Landscape: 688mmPortrait: 446.8 mmH:Landscape: 466.4 - 616.4mmPortrait: 707.1-737.0mmD: 301.6 mm
    NET Weight10.2 kg17 kg
    Intended of useDisplaying and viewing of digital images for diagnosis by trained physiciansThis device can not use for a life support system.This device must not be use in digital mammography.This device is designed for exclusive interconnection with IEC60601-1-1 certified equipmentDisplaying and viewing of digital images for diagnosis by trained physiciansThis device can not use for a life support system.This device must not be use in digital mammography.This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment
    Certifications & StandardsCE ITE/Medical Device Directive, UL/cUL (ANSI/AAMI ES 60601-1:2005), FCC Class B, EN60601-1-2, DIN V 6868-57, DICOMCE ITE/Medical Device Directive, UL/cUL (ANSI/AAMI ES 60601-1:2005), FCC Class B, EN60601-1-2, DIN V 6868-57, DICOM

    2. Sample size used for the test set and the data provenance

    Not applicable. This is a 510(k) submission for a display monitor, not an AI/algorithm-based diagnostic device requiring a test set of medical data (images, patient records, etc.). The "test" in this context refers to engineering and performance verification of the display hardware against technical specifications and safety standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth typically refers to clinical diagnoses or pathology results for AI/algorithm validation. For a display monitor, "ground truth" would be the expected performance based on engineering specifications and calibration.

    4. Adjudication method for the test set

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a display monitor, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a display monitor, not an algorithm.

    7. The type of ground truth used

    Not applicable in the clinical sense. The "ground truth" for evaluating this device would be its adherence to established technical specifications, industry standards (like DICOM Part 14), and safety regulations, verified through engineering tests and calibrations.

    8. The sample size for the training set

    Not applicable. This is a hardware device (display monitor), not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.