Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K042923

    Validate with FDA (Live)

    Date Cleared
    2004-12-17

    (56 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MD-6W Wireless Critical Data Viewer is intended to display patient vital signs in such a way that the data is continuously in a medical specialist's field of view. A typical application would be to view the vital signs of a patient undergoing surgery.

    Device Description

    The MD-6W Wireless Critical Data Viewer contains a mobile viewer (consisting of an LCD display unit, non-prescription safety glasses, and a control box with signal converter and lithium-ion rechargeable battery) and a base station that connects via cable to a patient monitor. The LCD display unit is mounted on the non-prescription safety glasses and is connected via a four-foot cable to the control box. The control box receives signal via wireless telemetry from the base station. The battery is recharged using a separate battery charger.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for MD-6W Wireless Critical Data Viewer

    This document describes the acceptance criteria and the study that proves the MD-6W Wireless Critical Data Viewer meets these criteria, based on the provided 510(k) summary.

    Note: The provided document is a 510(k) summary, which typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive de novo performance studies with detailed statistical analyses for novel devices.

    1. Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Technological CharacteristicsIndication, display format, field of view, focal range similar to predicate device (MD-6 Critical Data Viewer).The MD-6W Wireless Critical Data Viewer is similar to the predicate device in terms of indication, display format, field of view, and focal range.
    Performance TestingAcceptability demonstrated through standard optical, electrical, firmware, and mechanical tests.The MD-6W Wireless Critical Data Viewer has been subjected to standard optical, electrical, firmware, and mechanical tests to demonstrate the acceptability of the device.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided 510(k) summary does not explicitly detail a specific "test set" or its sample size in the way one might expect for a de novo performance study involving human subjects or image analysis. The performance testing appears to be centered on engineering and technical validation rather than clinical efficacy measured in a patient cohort.

    • Sample Size for Test Set: Not specified. The testing described (optical, electrical, firmware, mechanical) would typically involve testing of device units/components rather than a patient or clinical data set.
    • Data Provenance: Not applicable in the context of a clinical test set. The tests are likely internal engineering and quality assurance tests.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable or provided in the context of the described performance testing. The "ground truth" for optical, electrical, firmware, and mechanical tests would be defined by engineering specifications and industry standards, not expert consensus from medical professionals.

    4. Adjudication Method for the Test Set

    This information is not applicable or provided. Since the testing focuses on technical performance (optical, electrical, mechanical), there would not be an adjudication method like 2+1 or 3+1 used for expert consensus in a clinical evaluation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned or performed. The 510(k) summary focuses on substantial equivalence to a predicate device and technical performance testing, not on the impact of the device on human reader performance (e.g., medical specialists). Therefore, no effect size of human readers improving with AI vs. without AI assistance can be determined from this document. The device is a display tool, not an AI-powered diagnostic aid.

    6. Standalone Performance Study

    Yes, a standalone performance assessment was conducted, but it was purely technical. The summary states, "The MD-6W Wireless Critical Data Viewer has been subjected to standard optical, electrical, firmware, and mechanical tests to demonstrate the acceptability of the device." This implies an evaluation of the device's technical specifications and functionality independent of human interaction beyond basic usability. It's not a standalone clinical performance study, but a standalone engineering performance evaluation.

    7. Type of Ground Truth Used

    The "ground truth" for the performance testing cited (optical, electrical, firmware, and mechanical) would be based on:

    • Engineering Specifications: The design and performance parameters defined by MicroOptical Corporation.
    • Industry Standards: Relevant electrical safety, optical performance, and mechanical durability standards.
    • Predicate Device Performance: Implicitly, the performance of the predicate device (MD-6 Critical Data Viewer) sets a benchmark for substantial equivalence.

    It does not involve expert consensus, pathology, or outcomes data in the context of clinical "ground truth."

    8. Sample Size for the Training Set

    Not applicable. This device is a display system, not a machine learning or AI-driven diagnostic tool that requires a "training set" of data. Its function is to display vital signs, and its performance is evaluated based on its ability to do so reliably and according to its specifications.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for this device, a method for establishing its ground truth is not relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1