Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K091084

    Validate with FDA (Live)

    Date Cleared
    2009-04-29

    (14 days)

    Product Code
    Regulation Number
    884.5470
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to be inserted into the vagina to absorb menstrual fluid.
    Cotton menstrual tampons are unscented menstrual tampons that are inserted into the vagina and used to absorb menstrual fluid.

    Device Description

    The device is a conventional unscented menstrual tampon consisting of an absorbent pledget, a withdrawal cord, and an applicator. These tampons will be provided as three absorbencies: regular, super and super plus. Each tampon is wrapped in an individual wrapper and packaged in sealed multi-unit containers for retail sale.

    AI/ML Overview

    The provided documents describe the regulatory submission for "COTTON MENSTRUAL TAMPONS" (also branded as MAXIM MENSTRUAL TAMPONS and ORGANYC MENSTRUAL TAMPONS) seeking substantial equivalence to existing predicate devices. The evaluation focuses on the technical characteristics and functionality of the tampons, rather than complex AI or medical imaging. As such, many of the requested elements for AI/medical device studies (e.g., sample size for test sets, expert qualifications, MRMC studies, training set details) are not applicable to this submission.

    Here is an analysis based on the available information:

    1. Table of acceptance criteria and the reported device performance

    The acceptance criteria are implicitly defined by the characteristics of the predicate devices. The study aims to demonstrate that the proposed device, COTTON MENSTRUAL TAMPONS, has similar technological features and equivalent performance, particularly in absorbency and dimensions, to the predicate devices.

    Characteristic / FeatureAcceptance Criteria (Predicate: TAMPAX® Tampons / NATRACARE TAMPON)Reported Device Performance (COTTON MENSTRUAL TAMPONS)
    Intended useInserted into the vagina to absorb menstrual fluidInserted into the vagina to absorb menstrual fluid
    Indication for useAbsorb menstrual or other vaginal dischargeAbsorb menstrual or other vaginal discharge
    DesignCylindrical shape; Applicator with silky and rounded tipCylindrical shape; Applicator with silky and rounded lip
    Dimensions (Regular)Tampon: 47 mm (L), 12.5-13 mm (D) Applicator: 120-125 mm (L), 14 mm (D)Tampon: 47.5 mm ± 5% (L), 12 mm (D) Applicator: 120 mm ± 5% (L), 14 mm (D)
    Absorbency (Regular)6-9 g6-9 g
    Dimensions (Super)Tampon: 47 mm (L), 14 mm (D) Applicator: 120-125 mm (L), 16 mm (D)Tampon: 47.5 mm ± 5% (L), 14 mm (D) Applicator: 120 mm ± 5% (L), 16 mm (D)
    Absorbency (Super)9-12 g9-12 g
    Dimensions (Super Plus)Tampon: 47 mm (L), 14 mm (D) Applicator: 120 mm (L), 16 mm (D) (Tampax only)Tampon: 47.5 mm ± 5% (L), 14 mm (D) Applicator: 120 mm ± 5% (L), 16 mm (D)
    Absorbency (Super Plus)12-15 g (Tampax only)12-15 g
    MaterialsPredicate 1 (Tampax): Viscose Absorbent Core, Polypropylene / Polystyrene Nonwoven Predicate 2 (Natracare): 100% Organic Cotton Absorbent Core, 100% Organic Cotton Nonwoven FiberProposed Device: 100% Organic Cotton Absorbent Core, 100% Organic Cotton Nonwoven Fiber (Identical to Natracare)
    Additives/Finishing AgentsNo-containNo-contain
    Biological SpecificationsNon sterile, single-use onlyNon sterile, single-use only

    The reported device performance aligns with or is within acceptable deviations from the predicate devices' characteristics, particularly matching the Natracare Tampon in terms of material composition and overall absorbency ranges.

    2. Sample size used for the test set and the data provenance

    The document indicates "Functional laboratory testing performed in foreseeable operating conditions." This suggests physical testing of the tampons. However, details regarding the specific sample size for these tests (e.g., number of tampons tested for dimensions, absorbency) are not provided in the summary. The data provenance is from the manufacturer, COTTON HIGH TECH SL, located in Spain. The study appears to be part of a prospective evaluation for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable to this type of device and study. The "ground truth" (e.g., absorbency, dimensions) is established through standardized laboratory testing methods, not expert consensus in the clinical sense.

    4. Adjudication method for the test set

    Not applicable. Laboratory test results are typically objective measurements, not subject to adjudication in the way clinical interpretations are.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device, and therefore, an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (tampon), not an algorithm or software.

    7. The type of ground truth used

    The ground truth used for performance evaluation is based on objective laboratory measurements of physical characteristics (dimensions, absorbency) and chemical/biological testing of materials, adhering to relevant standards for menstrual tampons. The "ground truth" for comparison is the established characteristics and performance of the legally marketed predicate devices.

    8. The sample size for the training set

    Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1