Search Results
Found 1 results
510(k) Data Aggregation
(210 days)
Use of Dentsleeve Manometric Assemblies is indicated when measurements of gastrointestinal tract pressures are judged to be useful for determining management of adults, children & infants with proven or suspected gastrointestinal motor disorders.
Dentsleeve Manometric Assemblies are designed for the monitoring and analysis of gastrointestinal pressures. They are designed to be used in conjunction with a low compliance manometric perfusion pump. The major structures of the Manometric Assemblies is a multi-channeled, medical grade, silicone rubber extrusion.
The Assemblies are designed with variations in length and channel number (depending on the extrusion and application) to enable precise low compliance pressure measurement from most areas of the GI tract. When attached to a manometric perfusion pump, a constant flow of perfused into the assembly and into the GI tract of the patient provided the perfusion pump is correctly operated. When the pressure changes within the GI tract the pressure is transferred faithfully up the assembly and detected with external pressure transducers mounted on the manometric perfusion pump.
Here's an analysis of the provided text, focusing on the acceptance criteria and study information for the Dentsleeve Manometric Assemblies:
Unfortunately, the provided text does not contain the detailed information required to fill out a complete table of acceptance criteria and reported device performance, nor does it describe a specific study proving the device meets acceptance criteria in the way a typical medical device validation study report would.
The document is a 510(k) submission summary and an FDA clearance letter for a medical device (Dentsleeve Manometric Assemblies). These documents focus on establishing substantial equivalence to a predicate device rather than presenting a detailed, standalone study with specific performance metrics and statistical analyses against pre-defined acceptance criteria.
However, based on the provided text, we can infer some aspects and highlight what is missing:
Inferred Information and Missing Details
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Inferred from product description) | Reported Device Performance (Inferred/Implicit) |
|---|---|
| Biocompatibility: Device materials are safe for patient contact. | "The Manometric assemblies have been tested for biocompatibility..." (Testing results compiled in Attachment A, not provided) |
| Functionality: Device can accurately measure gastrointestinal pressures when connected to a perfusion pump. | "...designed to enable precise low compliance pressure measurement from most areas of the GI tract." |
| Sterilization: Device can be safely sterilized (up to 50 times by autoclave) without degradation. | "...ability to be sterilized up to 50 times by autoclave." (Testing results compiled in Attachment C, not provided) |
| Minimization of discomfort: Device material contributes to patient comfort during intubation. | "The flexibility of silicone rubber aids patient comfort..." |
| Avoidance of pharyngeal water infusion: Suitable for air perfusion in pharyngeal manometry. | "Dentsleeve assemblies are suitable for use with air perfusion which is used only in pharyngeal manometry..." |
| Risk Mitigation: Risks of use are minimal when operated and maintained as described. | "A Risk assessment and Failure Mode Effects and Criticality Analysis, the results of which can be seen in attachment D." (Attachment D not provided) |
Missing Information for a Comprehensive Table:
- Specific quantitative performance metrics: What defines "precise" pressure measurement? What are the tolerance limits for accuracy, drift, or signal-to-noise ratio? How were these measured?
- Detailed results for biocompatibility and sterilization: The document states testing was done, but doesn't provide the actual results or how they were compared against specific acceptance thresholds (e.g., cytotoxicity index, material integrity post-sterilization).
- Clinical effectiveness measures: While the device measures pressure, the document doesn't detail how well these measurements aid in patient management or diagnosis compared to other methods. This is often the focus of performance studies.
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not specified. The document only states "The Manometric assemblies have been tested for biocompatibility, functionality and ability to be sterilized..." without detailing the sample size for these tests (e.g., number of catheters tested, number of sterilization cycles for each, number of in-vitro pressure tests).
- Data Provenance: Not explicitly stated regarding the tests themselves. The company, Dentsleeve Pty Ltd, is located in Parkside, South Australia, so testing was likely conducted in Australia or by a contract lab. The document does not specify if the testing involved human subjects (prospective clinical study) or was entirely bench/lab-based (retrospective on manufactured components). Given the nature of a 510(k) for this type of device, most of the "testing" would likely be bench testing for functionality, biocompatibility, and sterilization, rather than large-scale clinical trials.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable / Not Mentioned: This type of information is typically relevant for studies involving "ground truth" established by human interpretation (e.g., radiology image interpretation). For a device that measures physical parameters like pressure, "ground truth" would be established by validated reference instruments or physical standards, not human experts. For example, pressure accuracy would be assessed against a calibrated pressure manometer.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable / Not Mentioned: Similar to point 3, adjudication methods are used to resolve disagreements among human experts when establishing ground truth. This is not pertinent to the type of bench testing described for the Dentsleeve Manometric Assemblies.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No: This is a manometric assembly, a physical device for measuring pressure, not an AI or imaging-based diagnostic tool. Therefore, an MRMC study is not relevant to this product.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable: This is not an algorithmic or AI device. It's a physical catheter. Standalone performance for this device would refer to its inherent physical properties and measurement capabilities, which are likely what the "functionality" testing aimed to assess.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Inferred: Bench Test Standards / Reference Devices: For functionality, the ground truth would have been established using calibrated pressure measurement equipment and physical standards. For biocompatibility, it would be based on ISO standards for medical device biocompatibility testing. For sterilization, it would involve microbiological testing against sterility assurance levels.
8. The sample size for the training set
- Not Applicable: This device does not have a "training set" in the context of machine learning or AI. It's a physical medical device.
9. How the ground truth for the training set was established
- Not Applicable: As above, there is no training set.
Summary of what the document focuses on:
The provided text from the 510(k) submission and FDA letter primarily focuses on:
- Describing the device and its intended use.
- Highlighting material properties (silicone for flexibility, reduced discomfort, heat tolerance for autoclaving).
- Stating that biocompatibility, functionality, and sterilization testing were done (referencing attachments that are not provided).
- Mentioning a Risk Assessment and FMECA (also referencing an attachment not provided).
- The FDA letter confirms "substantial equivalence" to a predicate device, which is the core of a 510(k) clearance, implying that prior testing demonstrated it meets the same safety and effectiveness standards as previously cleared devices.
To obtain the specific details requested, one would need to access the full 510(k) submission attachments (A, B, C, and D) which are referenced in the provided text.
Ask a specific question about this device
Page 1 of 1