Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K163536

    Validate with FDA (Live)

    Date Cleared
    2017-07-26

    (222 days)

    Product Code
    Regulation Number
    866.3361
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MALDI Biotyper CA System is a mass spectrometer systems using matrix-assisted laser desorption/ionization - time of flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens.

    The MALDI Biotyper CA System is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast infections.

    Device Description

    The MBT-CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization - time of flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens. The system uses a different methodology for organism identification based on unique protein patterns of the microorganisms obtained from mass spectrometry. The test organism's spectrum (a pattern of mass peaks) is compared with a reference spectra library (database). Using biostatistical analysis, a probability ranking of the organism identification is generated. The probability ranking is represented as a log(score) between 0.00 and 3.00. Organism identification is reported with high confidence if the log(score) is ≥2.00. An organism identification is reported with low confidence if the log(score) is between 1.70 and <2.00.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the MALDI Biotyper CA (MBT-CA) System, based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for sensitivity, specificity, or overall accuracy. Instead, it presents performance results from various studies (reproducibility, challenge panel, method comparison) and concludes that the device performs acceptably.

    However, based on the reported performance in the Method Comparison study, common metrics for identification systems would be:

    Performance Metric (Interpreted)Acceptance Criteria (Implied / Expected)Reported Device Performance (Overall Isolate Performance from Table 6)
    High Confidence ID Rate (≥ 2.0 log(score))High, ideally >95% for species identification1904 / 1930 = 98.65% (for high resolution species) (1904 + 130) / (1930 + 136) = 98.42% (for high & low resolution species/genus)
    Low Confidence ID Rate (≥ 1.7 to < 2.0 log(score))Low, indicating need for further action23 / 1930 = 1.19% (for high resolution species) 5 / 136 = 3.68% (for low resolution species/genus)
    Incorrect MBT-CA ID (≥ 1.7 log(score)) or No ID (< 1.7 log(score))Very Low, ideally <1%3 / 1930 = 0.16% (Incorrect for high resolution species) 1 / 136 = 0.74% (Incorrect for low resolution species/genus) Equivalent to 4 / 2066 = 0.19% (Total incorrect/no ID for positive cases)
    Overall Identification Rate (High or Low Confidence)High, ideally >95%(1904+23) / 1930 = 99.84% (for high resolution species) (1904+130+23+5) / (1930+136) = 99.81% (for high & low resolution species/genus)
    False Identification Rate0% (critical for diagnostic accuracy)0% reported across several validation studies (Repeatability/Precision, LOD, Sample Stability, Validation of 50 Representative Claimed Species, Nocardia Study). For the overall isolate performance, the "Incorrect MBT-CA ID" for positive cases (3+1=4) indicates a very low rate of incorrect IDs, which are distinct from "negative" cases. The document states "no isolates were falsely identified" in the reproducibility study and similar conclusions in other studies. For the method comparison, it is reported as 0% for negative cases and very low for positive cases.

    Note: The "acceptance criteria" presented above are inferred from the strong performance and conclusions drawn in the document, rather than explicitly stated numerical targets prior to testing.


    2. Sample Size Used for the Test Set and Data Provenance

    • Method Comparison Study (Overall Isolate Performance):

      • Sample Size: 2091 fresh and stored organisms.
      • Data Provenance: Organisms were tested at four (4) US clinical test sites and an in-house laboratory. Isolates were sub-cultured and sent to an interim reference laboratory and then to a sequencing reference laboratory for ground truth determination. This indicates prospective and retrospective data collection with a US origin.
    • Reproducibility Study:

      • Sample Size: 9 unique organisms (REPRO-02 excluded). Each organism tested in duplicate, 5 days, 2 runs/day, 3 sites (9 organisms x 2 replicates x 5 days x 2 runs x 3 sites = 540 measurements). Total MBT-CA IDs for summary = 179/180 per site.
      • Data Provenance: Conducted at three (3) clinical study sites (US, likely, given the FDA submission context). The organisms were "well-characterized," suggesting they might be reference strains or previously identified clinical isolates.
    • Challenge Panel Study:

      • Sample Size: 46 organisms.
      • Data Provenance: Selected from stored organisms from the clinical study, prepared by the interim reference laboratory. Tested at three (3) study sites (US, likely).
    • Biological/Technical Equivalency Studies:

      • Sample Size: 34 species for laser equivalency (4080 spectra). Multiple species for target equivalency (e.g., 1000 measurements for repeatability/precision, 1500 for LOD, 2500 for sample stability prior to matrix, 3000 for post-matrix stability, 50 FDA cleared organisms, 1500 for mass accuracy/edge effects). Nocardia Study: 30 strains covering 6 species, resulting in ~15,000 measurements.
      • Data Provenance: Not explicitly stated for specific origin, but these are technical validation studies performed by the manufacturer, likely controlled lab settings.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Method Comparison Study:

      • The ground truth was established by sequencing (16S rRNA or ITS sequencing and protein gene sequencing). This relies on established molecular biology techniques, not human expert interpretation. While experts run and interpret these sequences, the core ground truth is the genetic information itself. The document does not specify a number or qualification of "experts" in the sense of clinical reviewers for ground truth determination but implies reliance on the robust and objective results of gene sequencing performed by a sequencing reference laboratory.
    • Reproducibility Study:

      • Organisms were "well-characterized." The ground truth was presumably established by prior definitive identification methods, likely including gene sequencing or reputable reference lab methods. No mention of independent experts for this study's ground truth.
    • Challenge Panel Study:

      • Organism identifications were "blinded to test sites," and the panel was prepared by the study interim reference laboratory. The ground truth was established by the reference lab, again likely through gold standard methods like sequencing.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method involving multiple human readers or a specific consensus process for discrepancies in the generated log(scores) or identifications against a human-read ground truth. Instead:

    • The ground truth for the organism identity itself (reference algorithm) was established by molecular sequencing.
    • The device's log(score) provides a quantitative measure of confidence. If the log(score) is too low (<1.70), the result is "no ID." If it's between 1.70 and <2.00, it's "low confidence."
    • For discrepancies in the device's workflow (e.g., if Direct Transfer yields a low score), the protocol specifies progressive steps (extended Direct Transfer, then Extraction procedure) until a high confidence log(score) is achieved or the organism remains unidentified. This is an internal algorithmic "adjudication" within the device's workflow.
    • The final interpretation is stated to be by a "trained microbiologist," who should consider "all relevant information available" (Gram staining, colony morphology, growth characteristics, sample matrix, etc.) in conjunction with the device's results. This implies human oversight and potential adjudication of the device's output in a clinical setting, but not as part of the ground truth establishment for the studies described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is described where human readers' performance with and without AI (MALDI-TOF) assistance is measured to determine an effect size. The studies focus on the standalone performance of the MALDI Biotyper CA System against a molecularly derived ground truth. The device is intended to aid in diagnosis, implying it would be used by humans, but the studies do not quantify the improvement in human performance with the device.


    6. Standalone Performance Study (Algorithm Only)

    Yes, extensive standalone performance studies were done. The "Method Comparison" study, "Reproducibility" study, and "Challenge Panel" study, as well as the various "Equivalency Studies," all evaluate the MALDI Biotyper CA System's performance (algorithm only, with human operation but without human interpretation as the ground truth) against a molecularly derived ground truth or against itself under different conditions. The tables showing "Overall Isolate Performance" (Table 6, 7, 8) are prime examples of standalone performance evaluation.


    7. Type of Ground Truth Used

    • Method Comparison and Challenge Panel Studies: The primary ground truth for organism identification was established by molecular sequencing, specifically 16S rRNA or ITS sequencing and protein gene sequencing.
    • Reproducibility Study: Organisms were "well-characterized," implying a gold standard identification method, most likely molecular sequencing.
    • Equivalency Studies: These compared the performance of different components of the MBT-CA system (e.g., laser types, target plates) against each other or against established performance metrics, rather than identifying unknown organisms from scratch using an external ground truth. The underlying ground truth for the organisms used in these studies would still be molecular identification.

    8. Sample Size for the Training Set

    The document does not explicitly state the sample size used for training the MBT-CA System's reference database. It mentions the "MALDI Biotyper Reference Library for Clinical Applications (MBT-CA)" and states it's an "update" (K163536 vs. K142677). This reference library is what the device compares new spectra against. The training set would be the collection of reference spectra used to build this database. The provided text refers to new organisms added to the library but does not give a total size or the methodology used to build the original or updated library from a training perspective.


    9. How the Ground Truth for the Training Set Was Established

    The document implies that the reference library is built from well-characterized and known organisms. For the specific organisms added to the claimed list (Table 5), they represent further development/validation. The initial development of the database entries would have likely involved:

    • Definitive microbiological identification methods: Classic biochemical tests, microscopy, and, critically, molecular sequencing (e.g., 16S rRNA sequencing), which is mentioned as the ground truth method elsewhere (e.g., Acinetobacter baumannii complex resolution, Method Comparison).
    • The "Matching Algorithm" calculates matches by comparing a new spectrum against each single reference entry of a reference database. This suggests that each entry in the database itself originates from a definitively identified organism whose mass spectrum was collected and included.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1