510(k) Data Aggregation

    K Number
    K153447

    Validate with FDA (Live)

    Device Name
    MAGNETOM Spectra
    Date Cleared
    2016-03-31

    (125 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGNETOM Spectra is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.

    Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.

    The MAGNETOM systems may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-Safe biopsy needles.

    Device Description

    The subject device, Software syngo MR E11M for the MAGNETOM Spectra, is the latest software version for the Siemens MR MAGNETOM Spectra. It is modified based on the software version syngo MR E11A, which was cleared with K141977 on November 19, 2014. The software functionality and applications are based on syngo MR E11A except in technical instances where adaptations were needed to support the system specific hardware and optimize the sequence/protocols. System tests validating the aforementioned adapations are provided in Appendix 9.

    Listed below are the hardware updates to the MAGNETOM Spectra systems with software syngo MR E11M:

    • · two new coils
      • o Breast 18 (cleared with K133435 on December 12, 2013)
      • o Pediatric 16 (cleared with K140998 on July 11, 2014)
    • a modified MARS(Measurement and Reconstruction System) .
    • an updated MRAWP/MRWP(Syngo Acquisition Workplace/ Syngo . Workplace) based on the new host platform
    • . minor hardware changes to the transmit unit

    The MAGNETOM Spectra with software version syngo MR E11M will be offered ex-factory (new production) as well as in-field upgrades for the currently installed MAGNETOM Spectra systems.

    AI/ML Overview

    The provided text does NOT describe acceptance criteria for a device, nor does it detail a study that proves the device meets specific acceptance criteria in the way envisioned by the request.

    Instead, this document is a 510(k) Premarket Notification from the FDA to Siemens Medical Solutions USA, Inc., concerning the MAGNETOM Spectra, a magnetic resonance diagnostic device. The purpose of a 510(k) submission is to demonstrate that a device is "substantially equivalent" to a legally marketed predicate device, rather than to prove it meets specific performance acceptance criteria through clinical studies.

    Here's an analysis based on the available information, highlighting what is and is not present:

    Missing Information:

    • Acceptance Criteria Table: The document does not provide a table of acceptance criteria for device performance (e.g., sensitivity, specificity, accuracy, signal-to-noise ratio targets).
    • Reported Device Performance against Acceptance Criteria: Since no acceptance criteria are listed for diagnostic performance, there's no data reported against them. The document mentions nonclinical tests for SNR, image uniformity, and heating for new coils, and verification/validation for other software features, but doesn't provide specific numerical results or target values.
    • Sample Size for Test Set and Data Provenance: No test set (clinical or otherwise for diagnostic efficacy) is described, so sample size, country of origin, or retrospective/prospective nature of data are not detailed.
    • Number of Experts and Qualifications for Ground Truth: Not applicable, as there's no diagnostic performance study with ground truth establishment described.
    • Adjudication Method: Not applicable.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not mentioned or conducted.
    • Standalone Performance Study: No standalone diagnostic performance study (algorithm only) is described. The device is an MRI system, not an AI algorithm.
    • Type of Ground Truth Used: Not applicable for diagnostic performance as no such study is presented.
    • Sample Size for Training Set: Not applicable, as this is an MRI system and software update, not an AI model trained on a dataset.
    • How Ground Truth for Training Set was Established: Not applicable.

    **Information that is available: **

    The document focuses on demonstrating substantial equivalence to predicate devices for the MAGNETOM Spectra with software syngo MR E11M.

    • Device Description: The MAGNETOM Spectra with software syngo MR E11M is described as the latest software version for the Siemens MR MAGNETOM Spectra, modified from syngo MR E11A. It includes two new coils (Breast 18 and Pediatric 16), a modified MARS (Measurement and Reconstruction System), an updated MRAWP/MRWP (Syngo Acquisition Workplace/Syngo Workplace) and minor hardware changes.
    • Nonclinical Tests:
      • The coils were tested for SNR (Signal-to-Noise Ratio), image uniformity, and heating.
      • All other software features were verified and validated.
      • The document states these tests "demonstrate that the device performs as intended." However, no specific quantitative results or acceptance thresholds are provided.
    • Clinical Tests: "No clinical tests were conducted to support the subject device and the substantial equivalence argument, however clinical images were provided to support the new coils of the subject device." This explicitly states that no diagnostic clinical study was performed for this 510(k).
    • Ground Truth (for New Coils): While not explicitly stated as "ground truth," the mention of "clinical images provided to support the new coils" implies that the performance of these coils was visually assessed by experts using clinical data, likely within internal validation processes, to ensure image quality was acceptable. However, the details of this assessment are not provided.
    • Predicates:
      • Primary Predicate: MAGNETOM Spectra with syngo MR D12 (K121160)
      • Secondary Predicate: Siemens MAGNETOM Skyra with syngo MR E11A (K141977)
    • Regulatory Basis: The FDA's review concludes that the device is "substantially equivalent" to legally marketed predicate devices, meaning it has the same intended use and technological characteristics as the predicate, or if it has different technological characteristics, it does not raise new questions of safety or effectiveness.

    In summary, the provided document is a regulatory clearance letter and 510(k) summary, not a study report detailing performance against specific acceptance criteria for diagnostic accuracy or efficacy. It relies on non-clinical testing and substantial equivalence to legally marketed predicates for device clearance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121160

    Validate with FDA (Live)

    Device Name
    MAGNETOM SPECTRA
    Date Cleared
    2012-07-16

    (91 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGNETOM Spectra is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.

    Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

    The MAGNETOM Spectra may also be used for imaging during interventional procedures when performed with MR compatible devices such as inroom display and MR-safe biopsy needles.

    Device Description

    MAGNETOM Spectra (3T) is similar to the previously cleared MAGNETOM Skyra and MAGNETOM Verio, utilizing a superconducting magnet design. The open bore, whole body scanners are designed for increased patient comfort. They focus on ergonomics and usability to simplify the MR workflow.

    The MAGNETOM Spectra systems will be available in a fixed configuration.

    AI/ML Overview

    The provided text is a 510(k) summary for the Siemens MAGNETOM Spectra 3T System. It describes the device, its intended use, and argues for substantial equivalence to predicate devices. However, this document does not contain acceptance criteria for specific device performance metrics or the details of a study involving such criteria.

    Instead, it lists general performance and safety parameters that were measured to demonstrate equivalence to predicate devices, and that the device conforms to NEMA, IEC, and ISO standards. It does not provide the specific numerical acceptance criteria or the results of those measurements for the MAGNETOM Spectra, nor does it describe a study specifically designed to prove the device meets predefined acceptance criteria in the way you've outlined.

    Therefore, I cannot populate the table or answer many of your questions as the requested information is not present in the provided text.

    Here is what can be extracted or inferred:

    1. A table of acceptance criteria and the reported device performance:
    This information is not provided in the document. The document lists "Performance: Specification Volume, Signal to Noise, Image Uniformity, Geometric Distortion, Slice Profile. Thickness and Gap, High Contrast Spatial Resolution" as parameters that were measured, but it does not give specific acceptance criteria (e.g., "SNR > X") or the reported performance values for the MAGNETOM Spectra against those criteria. It only states that performance measurements were done and that the MAGNETOM Spectra is equivalent to predicate devices.

    2. Sample size used for the test set and the data provenance:
    This information is not provided in the document. No specific "test set" of patient data or phantoms is described with sample sizes or provenance for demonstrating performance. The evaluation seems to be based on engineering measurements against standards and comparison to predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This information is not provided in the document. There is no mention of a "test set" requiring expert ground truth establishment. The intended use indicates that images are "interpreted by a trained physician," but this pertains to clinical use, not a study evaluating the device's performance against expert consensus.

    4. Adjudication method for the test set:
    This information is not provided in the document. As no test set requiring expert ground truth is described, no adjudication method is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This information is not provided in the document. This is a submission for a Magnetic Resonance Diagnostic Device (an MRI machine), not an AI algorithm. Therefore, an MRMC study comparing human readers with and without AI assistance is irrelevant and not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    This information is not provided in the document. This is an MRI system, not a standalone algorithm.

    7. The type of ground truth used:
    This information is not provided in the document. The document focuses on technical performance metrics as evaluated against industry standards and predicate devices, rather than clinical ground truth (like pathology or outcomes data) from a patient study.

    8. The sample size for the training set:
    This information is not provided in the document. There is no mention of a "training set" as this is not an AI/machine learning device submission.

    9. How the ground truth for the training set was established:
    This information is not provided in the document. As there is no training set mentioned, no establishment of ground truth for it is discussed.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1