510(k) Data Aggregation

    K Number
    K220589

    Validate with FDA (Live)

    Date Cleared
    2022-05-13

    (73 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These inages and / or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.

    Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.

    Device Description

    MAGNETOM Skyra Fit with software syngo MR XA50A includes new and modified hardware and software compared to the predicate device, MAGNETOM Vida with software syngo MR XA31A. A high level summary of the new and modified hardware and software is provided below:

    Hardware
    Modified Hardware

    • Cover: The cover has been modified to bring the system up to the Siemens Healthineers Design incl. all BioMatrix components and interfaces.
    • EPC (Electronic Cabinet and Measurement Control / Electronic Power Cabinet): upgrade of components to upgrade the EPC to the newest electronic cabinet series

    Software
    New Features and Applications

    • TSE MoCo: TSE MoCo is an image-based motion correction in the averagedimension for the TSE pulse sequence type.
    • Automatic fiducial detection: MR Breast Biopsy is improved with an automatic fiducial detection.

    Modified Features and Applications

    • Fast GRE RefScan: A speed-optimized reference scan for GRAPPA and SMS kernel calibration for echo planar imaging pulse sequence types.
      Other Modifications and / or Minor Changes
    • The MAGNETOM Skyra Fit is a new MRI System which is the result of an upgrade from a MAGNETOM Verio.
    AI/ML Overview

    The provided document is a 510(k) summary for the MAGNETOM Skyra Fit device, which is a Magnetic Resonance Diagnostic Device (MRDD). This document primarily focuses on establishing substantial equivalence to a predicate device (MAGNETOM Vida) and does not detail a clinical study with acceptance criteria, human reader performance, or training/test set specifics for an AI algorithm.

    The document states:

    • "No additional clinical tests were conducted to support substantial equivalence for the subject device; however, as stated above, sample clinical images were provided. No clinical publications were referenced." (Page 6, Section 10. Clinical Tests / Publications)
    • The performance tests conducted were "Sample clinical images" and "Image quality assessments by sample clinical images." (Page 6, Section 9. Nonclinical Tests)

    Therefore, based on the provided text, a detailed answer to your request, particularly concerning acceptance criteria for an AI device, human reader studies, and AI training/test set specifics, cannot be fully provided directly from this document. This document describes a new MR system, not an AI software.

    However, I can extract the information that is present and explain where the requested information is not available:

    1. A table of acceptance criteria and the reported device performance

      • Not explicitly provided in terms of numerical acceptance criteria for a clinical study comparing performance against a predicate. The document states: "The results from each set of tests demonstrate that the subject device performs as intended and is thus substantially equivalent to the predicate device to which it has been compared." (Page 6, Section 9. Nonclinical Tests). The performance assessment appears to be qualitative based on "sample clinical images" and "image quality assessments."
      • Device Performance (as described): "The subject device performs as intended and is thus substantially equivalent to the predicate device." (Page 6)
    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

      • Not provided. The document mentions "sample clinical images" but does not specify the sample size, data provenance, or whether they were retrospective or prospective.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

      • Not provided. The document refers to images interpreted by a "trained physician" for diagnosis (Page 5, Section 5. Intended Use), but this is a general statement about the use of an MRDD, not specific to establishing ground truth for a test set in the context of a new feature evaluation or AI study.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

      • Not provided.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • No MRMC study was done. The document explicitly states, "No additional clinical tests were conducted to support substantial equivalence for the subject device..." (Page 6).
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not applicable/Not done for an AI algorithm. This document is for an MR imaging system, not an AI diagnostic algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      • Not explicitly stated/not applicable in the context of an AI study. The "sample clinical images" were likely evaluated for image quality and equivalence to the predicate, rather than for diagnostic ground truth for an AI algorithm.
    8. The sample size for the training set

      • Not applicable/Not provided. This is not a submission for an AI algorithm that would require a distinct training set.
    9. How the ground truth for the training set was established

      • Not applicable/Not provided.

    Summary based on the provided document:

    This 510(k) submission for the MAGNETOM Skyra Fit MRI system focuses on demonstrating substantial equivalence to a previously cleared predicate device (MAGNETOM Vida) by showing that new and modified hardware and software features maintain equivalent safety and performance. This is typically achieved through non-clinical performance testing, including image quality assessments, software verification and validation, and electrical/mechanical safety tests, rather than multi-reader clinical studies or AI algorithm performance evaluations.

    The "sample clinical images" mentioned are likely used to qualitatively demonstrate that the new system's output images are comparable in quality to those of the predicate device, thereby supporting the claim of substantial equivalence for diagnosis when interpreted by a trained physician. The document explicitly states that no additional clinical tests were conducted, meaning there were no new studies involving human readers or comparative effectiveness with or without AI assistance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123510

    Validate with FDA (Live)

    Device Name
    MAGNETOM SKYRA
    Date Cleared
    2013-05-17

    (185 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGNETOM Skyra is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.

    Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

    The MAGNETOM Skyra may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR-safe biopsy needles.

    Device Description

    TimTX TrueShape, an option for the MR system MAGNETOM Skyra, allows parallel transmission of RF pulses, shaping the RF excitation field locally and thus enabling selective excitation.

    syngo ZOOMit (introduced with the software version syngo MR D13C) is the application that utilizes TimTX TrueShape. It allows "zooming into" a part of the human body by selective excitation.

    The option TimTX TrueShape and syngo ZOOMit include two independent RF transmit paths fully integrated with the system, a modified whole-body RF coil, new options for B1 shimming, as well as new inline RF pulse generation. Pulse sequences allowing for different kinds of excitation, or zooming are derived from EPI and SPACE sequences. Protocol optimization for neurological, musculoskeletal, angiographic and oncological imaging is also inherent to the TimTX TrueShape option.

    AI/ML Overview

    The provided document describes a 510(k) submission for a new Magnetic Resonance (MR) system option (TimTX TrueShape) and software version (syngo MR D13C) for the MAGNETOM Skyra. This is a submission for a device upgrade, not a new standalone AI/ML device, and thus does not contain the typical information about acceptance criteria and a study proving the device meets those criteria in the way one would expect for an AI/ML product.

    Instead, the submission focuses on demonstrating substantial equivalence to a previously cleared predicate device (MAGNETOM Skyra with syngo MR D13A). This means the focus is on showing that the new features do not introduce new safety or effectiveness concerns, and that the device retains the core functionalities and performance of the predicate.

    Therefore, many of the requested items (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set details) are not applicable or not provided in this type of regulatory submission. The "acceptance criteria" here relate more to regulatory compliance and demonstrating that the new features do not negatively impact the established safety and performance of the base device.

    However, I can extract the relevant information that is provided and explain what is missing and why:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Safety: Device conforms to applicable FDA recognized and international IEC, ISO, and NEMA standards. Product Risk Management in compliance with ISO 14971:2009. Minimize electrical and mechanical risks. No new issues of safety.- Conformity to applicable FDA recognized and international IEC, ISO, and NEMA standards for performance and safety. - Product Risk Management accomplished through a process in compliance with ISO 14971:2009 to identify and mitigate potential hazards. - Risks controlled via measures in software development, SW testing, and product labeling. - Adherence to recognized industry practices and standards (e.g., IEC 60601-1 series) to minimize electrical and mechanical risk. - Conclusion: "does not introduce any new issues of safety or effectiveness."
    Effectiveness: Same intended use as predicate device. Same basic technical characteristics as predicate (magnetic resonance features and functionalities). Enhanced capabilities without introducing new effectiveness concerns.- Intended Use: "The MAGNETOM Skyra is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities... These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis." (Unchanged from predicate) - Technical Characteristics: "has the same intended use and the same basic technical characteristics as the predecessor device MAGNETOM Skyra with syngo MR D13A, with respect to the magnetic resonance features and functionalities." - Capability: "The differences... give the subject device greater capabilities than the predicate device, but have the enhanced technological characteristics as the predicate device." - Conclusion: "does not introduce any new issues of safety or effectiveness."

    Explanation: In a 510(k) for an MR system update, the "acceptance criteria" are implicitly met by demonstrating compliance with recognized standards and proving that the new features (TimTX TrueShape and syngo MR D13C) do not alter the fundamental safety or effectiveness of the device beyond what was established for the predicate. The "reported device performance" is essentially the statement that these criteria have been met through the mentioned compliance and risk management processes.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable/Not Provided: This document describes a traditional 510(k) for a hardware and software upgrade to an MR system, not a clinical study on an AI/ML algorithm. Therefore, there is no "test set" in the sense of a dataset for evaluating an algorithm's performance on patient data. The evaluation primarily focuses on engineering verification and validation against technical specifications and safety standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable/Not Provided: See explanation for point 2. Ground truth in the context of an AI/ML algorithm's clinical performance is not relevant for this type of submission. The "ground truth" for an MR system's performance is typically established through phantom studies, engineering tests, and adherence to established imaging physics principles.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable/Not Provided: See explanation for point 2 and 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No: No MRMC study was performed or described. This document is for a hardware and software upgrade to an MR scanner, not an AI/ML diagnostic aid that would assist human readers in interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable/Not Provided: This is not an AI/ML algorithm that operates in a "standalone" fashion for diagnosis. It's a component of an imaging system. The performance of the imaging system as a whole is subject to regulatory requirements.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable/Not Provided: For an MR system upgrade, "ground truth" generally refers to physical measurements and known properties of phantoms, or established imaging capabilities. Clinical outcome data or pathology are not used to validate the core performance of the MR system itself in this context.

    8. The sample size for the training set

    • Not Applicable/Not Provided: This document does not describe the development or training of an AI/ML algorithm in the sense of a "training set" of data.

    9. How the ground truth for the training set was established

    • Not Applicable/Not Provided: See explanation for point 8.