Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243532

    Validate with FDA (Live)

    Date Cleared
    2025-04-28

    (164 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LumiRex Single-Use Digital Flexible Ureteroscope is a single-use video endoscope designed for use with Contra Healthcare Solutions video processors, urology accessories, and other ancillary equipment used in urology and urological procedures.

    Device Description

    The LumiRex Single-Use Ureteroscope is a flexible video-based medical device designed for insertion through the urethra into the urinary tract. It comprises a long, thin tube equipped with a light source and a camera to provide real-time visualization of the urinary tract and interior of the kidney. The ureteroscope also features a working channel that allows for installation of fluids, the passage of various instruments, such as biopsy forceps, cytology brushes, biopsy brushes, and stone retrieval baskets to facilitate urology and ureteroscopy procedures.

    The ureteroscope can be used wired or wireless through a Wireless Scope Adapter or Wired Scope Adapter. The adapters transmit images from the ureteroscope to the Video Processor Unit where the user can view the image from the ureteroscope, control the LED, and save and export images from the ureteroscope onto USB removable media.

    AI/ML Overview

    The provided FDA 510(k) Clearance Letter for the LumiRex Ureteroscope does not detail a clinical study with acceptance criteria and device performance in the way typically expected for an AI/software as a medical device (SaMD) clearance. The document focuses on hardware aspects of the ureteroscope, such as sterilization, biocompatibility, electrical safety, and basic bench testing (e.g., field of view, depth of field, image quality).

    Specifically, there is no mention of an AI component, an algorithm, or any study involving human readers or ground truth for diagnostic accuracy. The "Software" section primarily refers to verification and validation of the device's operating software (e.g., for image processing, LED control) according to IEC 62304, not an AI or diagnostic algorithm.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and study that proves the device (in the context of an AI/diagnostic algorithm) meets acceptance criteria, an MRMC study, or the establishment of ground truth for a diagnostic task. The document provided does not contain this information.

    Based on the provided text, the LumiRex Ureteroscope is a physical medical device (endoscope) and not an AI/SaMD product that would typically involve the type of diagnostic performance studies you are asking about.

    To answer your prompt with the available information:


    Acceptance Criteria and Device Performance for LumiRex Ureteroscope (as per 510(k) Summary)

    Based on the provided FDA 510(k) Premarket Notification (K243532) for the LumiRex Ureteroscope, the "acceptance criteria" and "study" described are primarily focused on the physical and functional performance of the medical device itself, rather than the diagnostic performance of an AI algorithm. The rigorous detail for AI/diagnostic studies (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies) is not present because this is not an AI-driven diagnostic device.

    The clearance is for a ureteroscope, which is a tool for visualization and intervention. Its performance is evaluated against engineering standards and comparison to a predicate device, not against clinical diagnostic accuracy metrics established by human experts reading images with or without AI assistance.

    Here's an interpretation of the "acceptance criteria" and "reported device performance" based on the document, framed by the categories you requested, but noting where information is absent due to the nature of the device:

    1. Table of Acceptance Criteria & Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/Test DescriptionReported Device Performance/Conclusion
    Sterilization & Shelf-LifeSterilization validated to ISO 11135:2014.Sterile Barrier evaluated to ISO 11607-2:2019 & ASTM F1980.Shelf life validated.Ethylene Oxide Sterilization evaluated successfully.Sterile Barrier evaluated successfully.The shelf life of the LumiRex Single-Use Bronchoscope (Note: document inconsistently states Bronchoscope, likely a typo for Ureteroscope) has validated.
    ReprocessingReusable components reprocessing evaluated per:** Guidance for Industry & FDA Staff,** AAMI TIR12:2020,** AAMI/ANSI/ISO 11737-1:2018,** ANSI/AAMI ST79:2017,** ANSI/AAMI/ISO 17665-1:2006,** AAMI ST98:2022.Reprocessing evaluation performed in accordance with listed standards.
    BiocompatibilityEvaluation per ISO 10993-1, including Cytotoxicity, Sensitization, Systemic Toxicity, Pyrogenicity, Irritation.Biocompatibility evaluation conducted in accordance with referenced FDA guidance and ISO 10993-1.
    SoftwareVerification & validation per IEC 62304, ANSI/AAMI/IEC ES60601-1, FDA guidance "General Principles of Software Validation," & "FDA Guidance for the Content of Premarket Submissions for Device Software Functions."Software verification and validation testing conducted in compliance with listed standards and guidance. (Note: This refers to operational software, not AI/diagnostic algorithms).
    CybersecurityConformance to "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" (September 2023).Cybersecurity evaluated in conformance to the listed guidance.
    Electrical Safety & EMCCompliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18 for the Ureteroscope, Adapters, and Video Processor.Electrical Safety and Electromagnetic Compatibility testing performed, complying with listed standards.
    Bench Testing - ConnectorsISO 80369-7:2021 Small-bore Connectors.Conformance to ISO 80369-7:2021.
    Bench Testing - EndoscopesISO 8600-1:2015 General Requirements.ISO 8600-3:2019 Field of View & Direction of View.ISO 8600-4:2014 Max Width of Insertion Portion.LumiRex Bronchoscope (typo for Ureteroscope) met requirements defined in the standard. Comparative testing showed a wider field of view than the predicate, with the same depth of view.
    Bench Testing - Photobiol.IEC 62471:2006 Photobiological Safety.Compliance with IEC 62471:2006.
    Bench Testing - Image QualitySpatial Resolution MTF (ISO 12233), Signal to Noise Ratio (ISO 15739), Dynamic Range, Depth of Field, Geometric Distortion, Image Intensity Uniformity, Color performance.Comparative testing performed; results met requirements and demonstrated substantially equivalent performance to the predicate device. Specific numerical performance values are not provided in this summary.
    Bench Testing - System FunctionalLeaking, Irrigation, Bending, Articulating Bending Angle, Endurance of Bending Section, Instrument & Accessory Compatibility, Bending Section Radius.System Functional Testing performed. LumiRex Bronchoscope (typo for Ureteroscope) met the requirements defined in the standard.
    Design Verification (Wireless)Evaluation for Reprocessing, Software, Electrical Safety, EMC, System Functional Testing for wired/wireless scope adapters (FDA Guidance, IEEE/ANSI C63.27-2021).Demonstrated that the device design met requirements in referenced standards or matched predicate performance. Wireless functionality validated.

    Since the device is a ureteroscope and not an AI/SaMD, the subsequent points of your request are not applicable or details are not provided in this type of 510(k) summary.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Applicable in the context of an AI diagnostic test. The "test set" here refers to physical product units subjected to bench testing (e.g., specific endoscopes for image quality testing, or multiple cycles for endurance testing). No specific "sample size" for patient data or cases is mentioned because this is not an AI diagnostic device.
    • Data Provenance: Not applicable for diagnostic data as described. The testing is laboratory-based and engineering-focused.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not Applicable. "Ground truth" in this context (pathology, clinical outcome) is not established or used for the clearance of an endoscope's basic imaging and physical functionality. Experts might be involved in evaluating usability or design, but not for "ground truth" to validate a diagnostic algorithm.

    4. Adjudication Method for the Test Set:

    • Not Applicable. No diagnostic test set requiring adjudication (e.g., 2+1, 3+1) is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and Effect Size:

    • No. Such a study is typically performed for AI-assisted diagnostic tools. This 510(k) does not describe any MRMC study.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:

    • No. There is no standalone diagnostic algorithm described for this device.

    7. The Type of Ground Truth Used:

    • Not Applicable. No diagnostic ground truth (e.g., pathology, outcomes data) is used for the clearance of this device as described. Performance is measured against engineering standards and comparison to a predicate device's physical and functional attributes.

    8. The Sample Size for the Training Set:

    • Not Applicable. There is no AI/machine learning component described that would require a "training set" of diagnostic data.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable. As there is no training set for an AI algorithm, no ground truth establishment method is mentioned.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1