Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K142473

    Validate with FDA (Live)

    Date Cleared
    2015-04-13

    (222 days)

    Product Code
    Regulation Number
    880.6375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lubricating Jelly is a medical device intended for medical purposes, to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices when either sterile or non-sterile fields are required.

    Device Description

    Sion's Lubricating Jelly is used to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices when either sterile or non-sterile fields are required. Typical uses of the lubricating jelly include (but not limited) insertion of catheters, endoscopes surgical instruments, gynecological, rectal, oral and rubber or plastic etc. The lubricating jelly is clear, greaseless, odorless, water soluble and non-irritating to the skin, tissue and mucous membranes. Both product types (sterile and non-sterile) are packed in a bacterial barrier tubes or sealed laminated sachets. Sion's Lubricating Jelly is a body contact material and was evaluated for biocompatibility with accordance to FDA's Memorandum - #G95 1, May 1, 1995 and ISO 10993-1:2009.

    AI/ML Overview

    This document describes the 510(k) premarket notification for Sion Biotext Medical's Lubricating Jelly. It is a Class I medical device intended to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a table of tests performed and relevant standards. The reported device performance is a general statement that "All tested devices met the tests' requirements and pre-defined acceptance criteria."

    No.Test NamePer StandardAcceptance CriteriaReported Device Performance
    1.Chemical and Microbiological along Shelf Life:"Specification" for Appearance, Weight, pH, Viscosity. "Specification" for Sterility and GPT per ISO 11737-2, USP <71>. "Specification" for Preservative Effectiveness per USP <51>. "Specification" for Dye Test per ASTM 1929.Met pre-defined acceptance criteria
    1.1AppearanceSpecificationSpecificationMet pre-defined acceptance criteria
    1.2WeightSpecificationSpecificationMet pre-defined acceptance criteria
    1.3pHSpecificationSpecificationMet pre-defined acceptance criteria
    1.4ViscositySpecificationSpecificationMet pre-defined acceptance criteria
    1.5Sterility and GPTISO 11737-2, USP <71>SpecificationMet pre-defined acceptance criteria
    1.6Preservative EffectivenessUSP <51>SpecificationMet pre-defined acceptance criteria
    1.7Dye TestASTM 1929SpecificationMet pre-defined acceptance criteria
    2.Biocompatibility:10993-1 (General)"Specification" for Cytotoxicity per 10993-5. "Specification" for Sensitization per 10993-10. "Specification" for Irritation - Mucosal (Vaginal) per 10993-10. "Specification" for Acute Systemic Toxicity per 10993-11.Met pre-defined acceptance criteria
    2.1Cytotoxicity10993-5SpecificationMet pre-defined acceptance criteria
    2.2Sensitization10993-10SpecificationMet pre-defined acceptance criteria
    2.3Irritation - Mucosal (Vaginal)10993-10SpecificationMet pre-defined acceptance criteria
    2.4Acute Systemic Toxicity10993-11SpecificationMet pre-defined acceptance criteria
    3.Gamma Sterilization ValidationISO 11137-2, AAMI TIR 33"Specification"Met pre-defined acceptance criteria
    4.Sterile Packaging IntegrityISO 11607-1"Specification"Met pre-defined acceptance criteria

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the exact sample sizes for each test mentioned. It only states that "All tested devices met the tests' requirements." The data provenance is implied to be from Sion Biotext Medical Ltd. in Israel, as the company is based there and would conduct its own testing for device clearance. The studies appear to be prospective, designed to evaluate the physical and biological properties of the device against established standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information is generally not applicable to laboratory or performance testing of a lubricating jelly. The "ground truth" for these tests is established by adhering to widely accepted international standards (e.g., ISO, USP, ASTM) and internal specifications, not by expert consensus on clinical interpretation. The tests are objective measurements of physical, chemical, and biological properties.

    4. Adjudication Method for the Test Set

    Not applicable for this type of device and testing. The tests are typically pass/fail based on predetermined specifications or standard requirements.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers (e.g., radiologists) interpret images or data, and their performance with and without AI assistance is evaluated. This is a lubricating jelly, and its performance is assessed through laboratory testing, not human interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies described are standalone performance tests of the device itself, without any human-in-the-loop evaluation of its performance in a diagnostic or interpretive capacity. The "algorithm" in this context would be the physical and chemical formulation of the jelly and the manufacturing process, and its performance is evaluated against predefined specifications.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance is based on established industry standards (e.g., ISO 11737-2, USP <71>, ASTM 1929, ISO 10993 series, ISO 11137-2, AAMI TIR 33, ISO 11607-1) and internal specifications set by Sion Biotext Medical. These standards and specifications define the acceptable ranges for properties like pH, viscosity, sterility, biocompatibility, etc.

    8. The Sample Size for the Training Set

    Not applicable. This device is a medical product (lubricating jelly), not an AI algorithm that requires a "training set" for learning. The development process would involve formulation and optimization, but there isn't a "training set" in the sense used for machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this device. The formulation and characteristics of the lubricating jelly were likely developed based on scientific principles of chemistry and materials science, aiming to meet the performance criteria and safety standards required for medical devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K961502

    Validate with FDA (Live)

    Date Cleared
    1996-07-18

    (90 days)

    Product Code
    Regulation Number
    880.6375
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device.

    Device Description

    Lubricating Jelly is a patient lubricant intended for medical purposes that is used to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device.

    AI/ML Overview

    This document is a 510(k) summary for a medical device (Lubricating Jelly, Low Viscosity). It describes the device, its intended use, and claims substantial equivalence to a predicate device.

    However, this document does not contain any information about acceptance criteria or a study proving device performance against such criteria in the context of diagnostic accuracy, clinical outcomes, or AI algorithm performance. It's a regulatory filing for a low-risk medical device, where the primary demonstration of safety and effectiveness often relies on showing substantial equivalence to an already marketed device, rather than extensive performance studies with acceptance criteria as one would find for diagnostic algorithms.

    Therefore, I cannot extract the requested information (table of acceptance criteria, sample sizes, expert qualifications, etc.) because it is not present in the provided text. The document's purpose is to outline the device's characteristics and its equivalence to a predicate device, not to detail the results of a performance study with specific metrics and acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1