510(k) Data Aggregation

    K Number
    K192607

    Validate with FDA (Live)

    Device Name
    LiteWalk
    Manufacturer
    Date Cleared
    2019-12-19

    (90 days)

    Product Code
    Regulation Number
    890.5650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LiteWalk is intended for home and hospital use for temporary increase in local blood circulation in the lower extremities and temporary relief of minor muscle aches and pains where applied.

    Device Description

    The LiteWalk is a pneumatic sequential foot compression device, designed primarily for temporary improvement in blood flow circulation in the lower limb. The system is based on a pneumatic unit that includes a pump, valves, pressure sensor and tubing, and is fed by an external power supply of 12V DC. The unit is connected through 2 sets of 4-tubes to 2 inflatable sandals. Each sandal has 4 inflatable pressure chambers, located to match the foot anatomy: heel, arch, metatarsals, and toes. The sandals come in a pair, for right and left feet, and in 3 sizes: small, medium, large. The operation is based on sequential inflation of the 4 chambers, from heel to toes.

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter from the FDA for a device called "LiteWalk." It focuses on demonstrating substantial equivalence to a predicate device based on technical characteristics and non-clinical testing. Crucially, the document explicitly states there was "No clinical study was conducted to support this application." Therefore, there is no information in this document regarding acceptance criteria for device performance based on clinical outcomes, nor any study proving the device meets such criteria.

    The information provided primarily relates to comparisons with a predicate device and compliance with general safety standards.

    Here's a breakdown of the specific points requested, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    No clinical performance acceptance criteria or reported device performance for clinical outcomes are mentioned, as no clinical study was conducted. The document focuses on demonstrating substantial equivalence to a predicate device based on technical specifications. A summary table of technical comparison is provided in the document:

    SpecificationLiteWalkPPCIDDifferences discussion
    510(k) numberProposed DeviceK000655NA
    ManufacturerVIASONIX LTD.Medical Dynamics LLCNA
    Product Regulation21CFR890.5650 - Powered inflatable tube massager21CFR890.5650 - Powered inflatable tube massagerIdentical
    Product CodeIRP - Powered inflatable tube massagerIRP - Powered inflatable tube massagerIdentical
    Indications for useIntended for home and hospital use for temporary increase in local blood circulation in the lower extremities and temporary relief of minor muscle aches and pains where appliedIntended for home and hospital use for temporary increase in local blood circulation in the lower extremities and temporary relief of minor muscle aches and pains where appliedIdentical
    ContraindicationsUse of the LiteWalk is contraindicated in patients for whom:• Presumptive evidence of congestive heart failure where increase in blood returned to the heart could be detrimental• Pre-existing deep vein thrombosis (DVT) or blood clots• Existing hemorrhagic disease or coagulation problems• Thrombophlebitis• Pulmonary embolismUse of the PPCID is contraindicated in patients for whom an increase in blood returned to the heart could be detrimental, including those with presumptive evidence of congestive heart failure, pre-existing deep vein thrombosis, thrombophelbitis, existing hemorrhagic disease, coagulation problems, or pulmonary embolismIdentical
    Environment of UseHospital and Home careHospital and Home careIdentical
    Rx OR OTCOTCOTCIdentical
    Mode of compressionSequentialSequential *Not publicly availableSimilar to the predicate
    Use Time [min]Recommended 30 min30 min or more *Not publicly availableSimilar to the predicate
    Sequential cycles time2-3 cycles per minute2-3 cycles per minute *Not publicly availableSimilar to the predicate
    Modes1 mode : Sequential inflation/deflation of 4 chambers from heel to toes1 mode : Sequential inflation/deflation of 4 chambers from heel to toes *Not publicly availableSimilar to the predicate
    SpecificationLiteWalkPPCIDDifferences discussion
    Max Pressure160 mmHg80-200 mmHg *Not publicly availableSimilar to the predicate. Within the predicate pressure range
    Types of GarmentsFoot SandalFoot Sandal *Not publicly availableSimilar to the predicate. Foot Sandal type is supported by the predicate device
    Garment sizes3 sizes: small, medium, large3 sizes: small, medium, large *Not publicly availableSimilar to the predicate
    Patient contact materialsPolyester [PVB0123, Colored Fabric]Not known *Not publicly availableSimilar to the predicate. Both sharing a fabric for the Sandals. LiteWalk Polyester skin contact material comply with ISO 10993
    Patient contactNon conductiveNon conductive *Not publicly availableSimilar to the predicate
    Number of chambers44 *Not publicly availableSimilar to the predicate
    Line Voltage / Frequency100-240 Vac / 50-60Hz 12V , 2.0A115 Vac / 60 Hz, 7W *Not publicly availableSimilar to the predicate
    Power Consumption8 W7W *Not publicly availableSimilar to the predicate
    Safety featurePower button on main unit allows user to stop therapy session at any time.Power button on main unit allows user to stop therapy session at any time. *Not publicly availableSimilar to the predicate
    User InterfaceNoNo *Not publicly availableSimilar to the predicate
    IndicatorGreen ledLed *Not publicly availableSimilar to the predicate
    TechnologyCompressor and valve system which sequentially inflates inflatable chambersCompressor and air distributor system which sequentially inflates inflatable chambers *Not publicly availableBoth sharing the same functionality. The valve allows better control over inflation and allows quick deflation. This slight change doesn't impact product safety and performance.
    StandardsIEC 60601-1IEC60601-1-2IEC60601-1-11ISO 10993-1IEC 60601-1IEC60601-1-2ISO 10993-1*Not publicly availableSimilar to the predicate. The LiteWalk meets all latest standards edition

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable, as no clinical study or test set data for clinical performance was used. The evaluation was based on non-clinical performance (bench tests, verification of specifications, software validation) and comparison to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical study that would require expert-established ground truth was conducted.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical study that would require adjudication was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The LiteWalk is a pneumatic sequential foot compression device, not an AI-assisted diagnostic or imaging device. No MRMC study was mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The LiteWalk is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no clinical study was conducted. The "ground truth" for the substantial equivalence determination relied on demonstrating that the device's technical specifications and safety performance (through non-clinical testing and standard compliance) are comparable to the legally marketed predicate device.

    8. The sample size for the training set

    Not applicable, as no clinical study involving training data was conducted.

    9. How the ground truth for the training set was established

    Not applicable, as no clinical study involving training data was conducted.