Found 1 results

510(k) Data Aggregation

    K Number
    K220956

    Validate with FDA (Live)

    Device Name
    Libby Echo:Prio
    Manufacturer
    Date Cleared
    2022-07-20

    (110 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Libby™ Echo:Prio is software that is used to process previously acquired DICOM-compliant cardiac ultrasound images, and to make measurements on these images in order to provide automated estimation of several cardiac measurements. The data produced by this software is intended to be used to support qualified cardiologists, sonographers, or other licensed professional healthcare practitioners for clinical decision-making. Libby™ Echo:Prio is indicated for use in adult patients.

    Device Description

    Echo:Prio is an image post-processing analysis software device used for viewing and quantifying cardiovascular ultrasound images. The device is intended to aid diagnostic review and analysis of echocardiographic data, patient record management and reporting. The software provides an interface for a skilled sonographer to perform the necessary markup on the echocardiographic image prior to review by the prescribing physician. The markup includes: the cardiac segments captured, measurements of distance, time, area, quantitative analysis of cardiac function, and a summary report. The software allows the sonographer to enter their markup manually and/or manually correct automatically generated results. It also provides automated markup and analysis, which the sonographer may choose to accept outright, to accept partially and modify, or to reject and ignore. Machine learning based view classification and border segmentation form the basis for this automated analysis. Additionally, the software has features for organizing, displaying, and comparing to reference guidelines the quantitative data from cardiovascular images acquired from ultrasound scanners.

    AI/ML Overview

    The provided text describes the Libby™ Echo:Prio software, its intended use, and performance data from its premarket notification. Here's a breakdown of the acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in a tabular format with defined thresholds. However, it presents performance metrics from the validation study which serve as the evidence of the device's capability. We can infer the implicit "acceptance criteria" from these reported performance metrics, which are presented as achieved targets.

    Metric (Implied Acceptance Criteria)Reported Device Performance
    View Classification Accuracy97%
    View Classification F1 Score> 96.6% (average)
    View Classification Sensitivity (Sn)96.8% (average)
    View Classification Specificity (Sp)98.5% (average)
    Heart Rate (HR) Estimation Bias (Regression Slope)0.98 (95% CI) compared to 12-lead ECG ground truth
    Ejection Fraction (EF) Prediction (Bivariate Linear Regression Slope)0.79 (95% CI: 0.52, 0.98) compared to human expert annotations

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: The document states that performance testing was done "retrospectively on a diverse clinical dataset." However, the exact sample size for this test set is not specified in the provided text.
    • Data Provenance:
      • Country of Origin: Not specified.
      • Retrospective or Prospective: Retrospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: For the Ejection Fraction (EF) prediction, ground truth was established by "four human experts."
    • Qualifications of Experts: The specific qualifications (e.g., years of experience, subspecialty) of these four human experts are not explicitly stated. They are referred to as "human experts" or "clinicians."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The adjudication method for constructing the ground truth from the four human experts for EF prediction is not explicitly stated. It is only mentioned that EF prediction was compared with "annotations by four human experts."

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly described in the provided text. The performance data focuses on the software's standalone accuracy in comparison to a ground truth, rather than measuring the improvement of human readers assisted by the AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation of the algorithm's predictions (view classification, HR, EF) against established ground truth was performed. The data presented ("view classification accuracy of 97%", "HR output estimate is with minimal bias", "prediction of the EF output... had a slope of 0.79") reflects the algorithm's direct performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The types of ground truth used are:

    • 12-lead ECG: For Heart Rate (HR) estimation.
    • Human Expert Annotations/Consensus (implied): For Left Ventricular Ejection Fraction (EF) prediction, derived from the "annotations by four human experts." For view classification, the "accuracy" and "F1 value" imply a comparison to a set ground truth, likely also established by experts.

    8. The sample size for the training set

    The sample size for the training set used to develop the machine learning model is not specified in the provided text.

    9. How the ground truth for the training set was established

    The method for establishing ground truth for the training set is not specified in the provided text. It only vaguely mentions "Machine learning based view classification and border segmentation form the basis for this automated analysis."